SYVA EMIT II PLUS AMPHETAMINES ASSAY

K031004 · Dade Behring, Inc. · DKZ · Jun 6, 2003 · Clinical Toxicology

Device Facts

Record IDK031004
Device NameSYVA EMIT II PLUS AMPHETAMINES ASSAY
ApplicantDade Behring, Inc.
Product CodeDKZ · Clinical Toxicology
Decision DateJun 6, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3100
Device ClassClass 2

Indications for Use

The Emit® II Plus Amphetamines Assay is a homogeneous enzyme immunoassay with a 300 ng/mL cutoff, 500 ng/mL cutoff or a 1000 ng/mL cutoff (SAMHSA initial test cutoff level). The assay is intended for use in the qualitative and semiquantitative analyses of amphetamines in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers. The Emit® II Plus Amphetamines Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

Device Story

Emit® II Plus Amphetamines Assay is a homogeneous enzyme immunoassay for detecting amphetamines in human urine. Input: urine samples; processed via chemistry analyzers using monoclonal antibody technology. Output: qualitative or semiquantitative concentration values. Used in clinical laboratories; operated by trained technicians. Results serve as preliminary screening; require confirmation by GC/MS or other chemical methods. Clinical decision-making relies on professional judgment alongside these results. Benefits include rapid, automated screening for drug-of-abuse testing.

Clinical Evidence

Bench testing only. Performance evaluated via method comparison studies using 124 urine specimens per cutoff level (300, 500, 1000 ng/mL) on the Syva®-30R Biochemical System. Compared against GC/MS reference method. Percent agreement reported as 97% (300 ng/mL), 96% (500 ng/mL), and 86-97% (1000 ng/mL) depending on the reference/comparative method used. Discrepant results were analyzed semiquantitatively to confirm values relative to cutoffs.

Technological Characteristics

Homogeneous enzyme immunoassay; monoclonal antibody-based. Designed for use on automated chemistry analyzers. Qualitative and semiquantitative detection of amphetamines in urine. Cutoff levels: 300, 500, 1000 ng/mL.

Indications for Use

Indicated for the qualitative and semiquantitative analysis of amphetamines in human urine to provide preliminary analytical test results. Intended for use with chemistry analyzers in clinical settings. Requires confirmation by alternative chemical methods (e.g., GC/MS).

Regulatory Classification

Identification

An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K031004 Dade Behring Inc. 510(k) Premarket Notification Syva® Emit® II Plus Amphetamines Assay # 510(k) Summary of Safety and Effectiveness Information Syva® Emit® II Plus Amphetamines Assay March 28, 2003 #### MANUFACTURER AND CONTACT INFORMATION I. Manufacturer: Syva Company - Dade Behring Inc. 20400 Mariani Ave. Cupertino, CA 95014 Contact Information: Dade Behring Inc. 500 GBC Drive Newark, DE 19702 Attn: Radames Riesgo Phone: 305.480.7558 FAX: 305.552.5288 #### II. DEVICE NAME AND CLASSIFICATION NAME Trade or Proprietary Name: Syva® Emit® II Plus Amphetamines Assay Common or Usual Name: Enzyme Immunoassay, Amphetamine/Methamphetamine Amphetamine test system (21 CFR §862.3100) Methamphetamine test system (21 CFR §862.3610) Class II #### IDENTIFICATION OF THE LEGALLY MARKETED DEVICE III. Syva® Emit® II Plus Monoclonal Amphetamine/Methamphetamine Assay (K993982). #### DEVICE DESCRIPTION IV. The Emit® II Plus Amphetamines Assay is a homogeneous enzyme immunoassay for the qualitative and semiquantitative analysis of amphetamines in human urine. #### V. DEVICE INTENDED USE The Emit® II Plus Amphetamines Assay is a homogeneous enzyme immunoassay with a 300 ng/mL cutoff, 500 ng/mL cutoff or a 1000 ng/mL cutoff (SAMHSA initial test cutoff level). The assay is intended for use in the qualitative and semiquantitative analyses of amphetamines in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers. ﺗﻘ {1}------------------------------------------------ The Emit® II Plus Amphetamines Assay provides only a preliminary analytical test result. A more specific alternative chemical must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used. #### VI. SUBSTANTIAL EQIVALENCE The Emit® II Plus Amphetamines Assay is substantially equivalent in intended use and methodology to the Emit® II Plus Monoclonal Amphetamine/Methamphetamine Assay currently in the market. The proposed device, like the predicate device is a monoclonal antibody assay intended for the qualitative and semiquantitative analysis of amphetamines in human urine. ### VII. DEVICE PERFORMANCE CHARACTERISTICS #### 300 ng/mL Cutoff A. ### Method Comparison - Qualitative The results of one hundred and twenty-four (124) urine specimens obtained with the Syva® Emit® II Plus Amphetamines Assay on the Syva®-30R Biochemical System (K912024) were compared to the results obtained with the reference method, GC/MS. Both methods used a cutoff level of 300 ng/mL for total amphetamines. Of the sixty-six (66) samples found to be positive by GC/MS (≥ 300 ng/mL total amphetamines), sixty-two (62) were also found positive by Syva Emit II Plus Amphetamines Assay for a percent agreement of 97% (120/124). Fifty-eight (58) samples were found to be negative by GC/MS (<300 ng/mL total amphetamines) and sixty-two (62) were found to be negative by the proposed device. Four (4) discrepant samples were found by the proposed device. Semiquantitative analysis of the discrepant samples using the proposed device showed values for total amphetamines within -25% of the cutoff. | Qualitative Results 300 ng/mL Cutoff | | | | | |--------------------------------------|--|--|--|--| |--------------------------------------|--|--|--|--| | | | Reference Method (GC/MS) | | |-------------------------------------------|----------|--------------------------|----------| | | | Positive | Negative | | Syva® Emit® II Plus<br>Amphetamines Assay | Positive | 62 | 0 | | | Negative | 4 | 58 | Percent agreement: 97% ﺔ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﺘﻲ ﺗﻘﺮﻳﺒﺎ {2}------------------------------------------------ ### 500 ng/mL Cutoff B. Method Comparison - Qualitative The results of one hundred and twenty-four (124) urine specimens obtained with the Syva® Emit® II Plus Amphetamines Assay on the Syva®-30R Biochemical System (K912024) were compared to the results obtained with the reference method, GC/MS. The proposed method used a cutoff level of 500 ng/mL for total amphetamines. While the results of GC/MS were interpreted using the proposed SAMHSA confirmatory guidelines (amphetamine 250 ng/mL or methamphetamine 250 ng/mL and minimum amphetamine 100 ng/mL). Fifty-nine (59) samples were found to be positive by GC/MS (proposed SAMHSA guidelines) and sixty-two (62) were found positive (≥500 ng/mL) by Syva® Emit® II Plus Amphetamines Assay for a percent agreement of 96% (119/124). Of the sixty-five (65) samples found to be negative by GC/MS, sixty-two (62) were also found to be negative by ﺎ ﺗﻘﺮ the proposed device. Five (5) discrepant samples were found. Four (4) of the five discrepant samples were identified as positive by the proposed device and were also found to have more than 500 ng/mL of total amphetamines when they were tested semiquantitatively. However, these four samples were considered negative by GC/MS regardless of the total amphetamines results because they did not meet the acceptance criterion of at least 100 ng/mL of amphetamine of the proposed SAMHSA guidelines for confirmatory test. The other discrepant sample was identified as negative by proposed device and was found to have total amphetamines within -25% of the cutoff level when analyzed semiquantitatively. # Qualitative Results 500 ng/mL Cutoff | | Reference Method (GC/MS) | | | |-------------------------------------------|--------------------------|----------|----| | | Positive | Negative | | | Syva® Emit® II Plus<br>Amphetamines Assay | Positive | 58 | 4 | | | Negative | 1 | 61 | Percent agreement: 96% ### ﺰ 1000 ng/mL Cutoff Method Comparison - Qualitative The results of one hundred and twenty four (124) urine specimens obtained with the Syva® Emit® II Plus Amphetamines Assay on the Syva®-30R Biochemical System (K912024) were compared to GC/MS as reference methodology. The proposed method used a cutoff level of 1000 ng/mL while the reference method used a cutoff level of 500 ng/mL according to SAMHSA confirmatory guideline (amphetamine 500 ng/mL or methamphetamine 500 ng/mL plus minimum amphetamine 200 ng/mL). In addition, the Syva® Emit® II Plus Amphetamines Assay was compared to the Syva® Emit® II Plus Monoclonal Amphetamine/Methamphetamine Assay (K993982) by evaluating the same number of urine {3}------------------------------------------------ specimens on the Syva®-30R Biochemical System. For this comparative study, both devices used a cutoff level of 1000 ng/mL for total amphetamines. Sixty-five (65) samples were found to be positive by GC/MS (SAMHSA confirmatory guidelines) and sixty-two (62) were found positive (≥1000 ng/mL) by Syva® Emit® II Plus Amphetamines Assay for a percent agreement of 86% (107/124). Fifty-nine (59) samples were found to be negative by GC/MS and sixty-two (62) were found to be negative by the proposed device. Seventeen (17) discrepant samples were found. Seven (7) of the discrepant samples that were found positive by the proposed device showed semiquantitative results over 1000 ng/mL, however, they were considered negative by GC/MC because they did not meet the minimum 200 ng/mL amphetamine criterion for SAMHSA confirmatory test. Ten (10) discrepant samples were found to be negative by the proposed test and were found to have total amphetamines values within the -25% and the cutoff. ﻧﺘﻘ | Qualitative Results 1000 ng/mL Cutoff | | | | | |---------------------------------------|--|--|--|--| |---------------------------------------|--|--|--|--| | | | Reference Method (GC/MS) | | |-------------------------------------------|----------|--------------------------|----------| | | | Positive | Negative | | Syva® Emit® II Plus<br>Amphetamines Assay | Positive | 55 | 7 | | | Negative | 10 | 52 | Percent agreement: 86% Sixty (60) samples were found to be positive by the comparative method and sixty-two (62) were found positive (≥1000 ng/mL) by Syva® Emit® II Plus Amphetamines Assay for a percent agreement of 97% (120/124). Sixty-four (64) samples were found to be negative by the comparative method and sixty-two (62) were found to be negative by the proposed device. Four discrepant samples were found. Three (3) of the discrepant samples that were found to be positive by the proposed device, were confirmed positive by GC/MS results. One (1) discrepant samples found to be negative by the proposed device was also confirmed as negative by GC/MS results. # Qualitative Results 1000 ng/mL Cutoff | | | Comparative Method | | |-------------------------------------------|----------|--------------------|----------| | | | Positive | Negative | | Syva® Emit® II Plus<br>Amphetamines Assay | Positive | 59 | 3 | | | Negative | 1 | 61 | Percent agreement: 97% {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular arrangement of the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" surrounding a stylized eagle symbol. The eagle symbol is composed of three curved lines that form the shape of an eagle's head and wings. JUN - 6 2003 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Rebecca S. Ayash Director. Regulatory Affairs & Compliance Dade Behring, Inc. 500 GBC Drive Newark, DE 19702 Re: k031004 Trade/Device Name: Emit® II Plus Amphetamines Assay Regulation Number: 21 CFR 862. 3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: DKZ Dated: March 28, 2003 Received: March 31, 2003 Dear Ms Ayash: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {5}------------------------------------------------ ### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ 510(k) Number (if known): _K03 1004 Device Name: Emit® II Plus Amphetamines Assay ### Indications for Use: The Emit® II Plus Amphetamines Assay is a homogeneous enzyme immunoassay with a 300 ng/mL cutoff, 500 ng/mL cutoff or a 1000 ng/mL cutoff (SAMHSA initial test cutoff level). The assay is intended for use in the qualitative and semiquantitative analyses of amphetamines in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers. The Emit® II Plus Amphetamines Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used. Sean Cooper Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K031004 ### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) X ﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ OR Over-The-Counter-Use (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...