MEDRELIEF SE SERIES, MODELS SE-50, SE-100, SE-200, SE-300
K030998 · Healthonics, Inc. · IPF · Oct 10, 2003 · Physical Medicine
Device Facts
Record ID
K030998
Device Name
MEDRELIEF SE SERIES, MODELS SE-50, SE-100, SE-200, SE-300
Applicant
Healthonics, Inc.
Product Code
IPF · Physical Medicine
Decision Date
Oct 10, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MedRelief SE-50 and SE-100. and the MedRelief SE-200 and SE-300 in subthreshold and interferential modes, are intended for the following indications: - Relief of chronic intractable pain. . - Adjunctive treatment of post-surgical or post-traumatic acute pain. . The MedRelief SE-200 and SE-300 in muscle stimulation mode are intended for the following indications: - Relaxation of muscle spasms. ● - Prevention or retardation of disuse atrophy. . - Muscle re-education. . - . Increasing local blood circulation. - Maintaining or increasing range of motion . - . Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis (SE-300 only). For the SE-200 and SE-300 in muscle stimulation mode, the lowest three intensity settings may not provide effective muscle contractions to achieve the indications for use, especially when applied to large muscles or muscle groups. These settings are provided so that the intensity may be ramped up slowly for improved user comfort.
Device Story
MedRelief SE Series (SE-50, SE-100, SE-200, SE-300) provides electrotherapeutic stimulation via self-adhesive skin electrodes; battery-powered (9V); single-channel. Inputs: user-selected mode via rotary/slide switches. Operation: analog circuitry generates 4150 Hz carrier wave pulses (30 or 120 microseconds); modulated by envelopes (0.25-130 Hz) for Structured Energy (subthreshold), muscle stimulation, or interferential stimulation. Output: electrical pulses to body. Used in clinical/home settings; operated by patient or clinician. Benefits: pain relief, muscle spasm reduction, atrophy prevention, improved circulation/ROM. No microprocessor or software.
Clinical Evidence
Bench testing only. Performance testing conducted per FDA guidance for Powered Muscle Stimulators (June 9, 1999). Electrical safety testing performed to UL-2601. Electromagnetic compatibility evaluated against EN 60601-1-2 (1993). Conformance to ANSI/AAMI NS4-1985 verified via performance testing.
Technological Characteristics
Analog electronic circuitry; 9V battery-powered; single-channel. Output: 4150 Hz carrier wave; 30 or 120 microsecond pulse duration; 0.25-130 Hz modulation. Compliance: UL-2601, ANSI/AAMI NS4-1985. No microprocessor; no software. Form factor: compact, lightweight, rotary/slide switch controls.
Indications for Use
Indicated for patients requiring relief of chronic intractable pain, adjunctive treatment of post-surgical or post-traumatic acute pain, relaxation of muscle spasms, prevention/retardation of disuse atrophy, muscle re-education, increased local blood circulation, maintenance/increase of range of motion, and immediate post-surgical calf muscle stimulation to prevent venous thrombosis.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
OCT 1 0 2003
# 510(k) SUMMARY.
### MedRelief SE ("Structured Energy") Series (consisting of models SE-50, SE-100, SE-200 and SE-300)
Common/Classification Names: Microcurrent TENS Device, 21 CFR 882.5890 Interferential stimulator. 21 CFR 890.5850 Powered muscle stimulator, 21 CFR 890.5850
Healthonics, Inc. 903 Main Street South New Ellenton, SC 29809 Establishment Registration #1067052
> Contact: James W. Kronberg Prepared: March 27, 2003
#### A. LEGALLY MARKETED PREDICATE DEVICES.
Devices in the MedRelief SE Series™ are substantially equivalent: (1) in powered muscle stimulation functions to the Microstim J-1304D, made by Nisha Communications Industries, Light Industrial Area, Jodhpur 342003 India, which was cleared for marketing by the FDA on July 13, 1998 as K980040; (2) in interferential stimulation functions to the IF-II Model 7200, made by Medical Devices, Inc., 833 Third Street Southwest, St. Paul, MN 55112-3483, which was cleared for marketing by the FDA on November 6, 1992 as K923914: and. (3) in microcurrent TENS stimulation functions to the original MedRelief device, made by Healthonics, Inc., 903 Main Street South, New Ellenton, SC 29803, which was cleared for marketing by the FDA on Feb. 21. 2001 as K003507. and to the Home Microcurrent HMC, also made by Medical Devices, Inc., which was cleared for marketing by the FDA on March 14, 1994 as K935132.
{1}------------------------------------------------
### B. DEVICE DESCRIPTION.
The MedRelief SE Series™ provides a selection of both high-level and subthreshold electrotherapeutic modes in a compact, lightweight and userfriendly package. "Subthreshold" as used here means electrical stimulation not strong enough to cause nerve depolarization or muscle contraction, but still active at a cellular level, as in microcurrent TENS. Healthonics' proprietary version of subthreshold stimulation has been named Structured Energy™
Devices in the MedRelief SE Series™ provide combinations of these stimulation modes as follows:
| Model<br>number | Includes: muscle<br>stimulation | interferential<br>stimulation | Structured<br>EnergyTM | Controls |
|-----------------|---------------------------------|-------------------------------|------------------------|----------|
| SE-300 | X | X | X | Rotary |
| SE-200 | X* | X* | X | Rotary |
| SE-100 | -- | -- | X | Rotary |
| SE-50 | -- | -- | X | Recessed |
* - Output provided over reduced intensity range.
Models SE-200™ and SE-300™ provide all three stimulation modes, while Models SE-50™ and SE-100™ are miniaturized, single-mode Structured Eneray™-only versions. All devices are completely powered by 9-volt batteries, without provisions for line power or AC adapters. All devices are single-channel only.
All SE Series ™ devices produce trains of brief electrical pulses which are applied to the body through self-adhesive electrodes placed on the skin. Pulses recur at a nominal 4150 Hz and may be either 30 microseconds or 120 microseconds long. These pulses form a carrier wave which may then be modulated by one of three different envelopes, yielding respectively Structured Energy™, muscle stimulation, or premodulated interferential stimulation. Envelope frequencies range from 0.25 Hz to 130 Hz, depending on the mode being used.
All functions for a particular mode are preset, and accessed through a single rotary switch which also turns the device on and off. In the SE-200™ and SE-300™, a second rotary switch provides intensity control. In the SE-100™, all control functions are integrated into a single rotary switch, while in the SE-50™ this switch is replaced by two recessed, miniature slide switches.
While the MedRelief SE Series™ incorporates some ancillary digital components such as frequency dividers, its primary functioning is analog, with all
## 000063
{2}------------------------------------------------
digital functions hard-wired. Since there is no microprocessor, no software is needed.
#### C. INTENDED USE.
Since the MedRelief SE Series™ is a family of devices with differing functions, only those indications which the FDA has traditionally approved for the corresponding device types are sought.
For the SE-50™ and SE-100™, which include only subthreshold stimulation substantially equivalent to subthreshold TENS devices previously cleared by the FDA, the indications sought are:
- * Relief of chronic intractable pain.
- * Adjunctive treatment of post-surgical or post-traumatic acute pain.
For the SE-200™ and SE-300™, which include powered muscle stimulation as well as interferential and subthreshold stimulation, the following additional indications are sought:
- * Relaxation of muscle spasms.
- * Prevention or retardation of disuse atrophy
- * Muscle re-education
- * Increasing local blood circulation
- * Maintaining or increasing range of motion
- * Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis.
#### D. SUBSTANTIAL EQUIVALENCE SUMMARY.
The MedRelief SE Series™ is a family of medical devices, having the same indications for use, target populations, and technological characteristics as the legally marketed predicate devices. This premarket notification has described the characteristics of the MedRelief SE Series™ in sufficient detail to assure substantial equivalence.
#### E. TECHNOLOGICAL CHARACTERISTICS.
Both new and predicate devices employ electronic circuitry to produce electrical output waveforms at closely comparable voltage and current levels and with closely comparable waveforms and timing. The MedRelief SE Series™ is primarily analog, with some hard-wired digital functions but no microprocessor and thus no software.
{3}------------------------------------------------
#### F. TESTING.
Healthonics has carried out design and performance testing to address the following issues:
(1) Performance testing for the purpose of addressing the points in the FDA guidance document, "Guidance Document for Powered Muscle Stimulator 510(k)'s," dated June 9, 1999.
(2) Electrical safety. The SE Series™ was designed to comply with UL-2601 and will be tested to this standard by an independent test laboratory before marketing.
(3) Electromagnetic compatibility. The SE Series™ was designed to comply with EN 60601-1-2, May 1993. However, the FDA guidance document exempts powered muscle stimulators from compatibility requirements unless specific claims of compatibility are made. Since Healthonics makes no such claims, and since no other output mode exceeds muscle stimulation in voltage, current or power. Healthonics believes that the SE Series™ is also exempt from this requirement.
(4) Conformance to NS4-1985. The SE Series™ was designed to comply with applicable requirements of ANSI/AAMI NS4-1985, as supported by performance testing.
#### G. CONCLUSIONS.
In summary, this pre-market submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food, Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 0 2003
Mr. James W. Kronberg Chief Engineer Healthonics, Inc. 903 Main Street South New Ellenton, SC 29809
Re: K030998
Trade/Device Name: MedRelief SE ("Structured Energy") Series™, Models SE-50, SE-100, SE-200 and SE-300 Regulation Number: 21 CFR 890.5850 and 21 CFR 882.5890 Regulation Name: Powered muscle stimulator and Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: II Product Code: IPF, GZJ, LIH Dated: August 4, 2003 Received: August 5, 2003
Dear Mr. Kronberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. James W. Kronberg
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
for Mark A. Millman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/21 description: The image is a black and white abstract design. A large black shape occupies the upper left corner, contrasting sharply with the white background. The black shape has an irregular edge, and there are a few small black spots scattered in the lower left area.
### STATEMENT OF INDICATIONS FOR USE.
510(k) number (if known) K030998
Device Name: MedRelief SE ("Structured Energy") Series, consisting of model numbers SE-50, SE-100, SE-200 and SE-300.
Indications for Use:
The MedRelief SE-50 and SE-100. and the MedRelief SE-200 and SE-300 in subthreshold and interferential modes, are intended for the following indications:
- Relief of chronic intractable pain. .
- Adjunctive treatment of post-surgical or post-traumatic acute pain. .
The MedRelief SE-200 and SE-300 in muscle stimulation mode are intended for the following indications:
- Relaxation of muscle spasms. ●
- Prevention or retardation of disuse atrophy. .
- Muscle re-education. .
- . Increasing local blood circulation.
- Maintaining or increasing range of motion .
- . Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis (SE-300 only).
For the SE-200 and SE-300 in muscle stimulation mode, the lowest three intensity settings may not provide effective muscle contractions to achieve the indications for use, especially when applied to large muscles or muscle groups. These settings are provided so that the intensity may be ramped up slowly for improved user comfort.
(Please do not write below this line - continue on another page if needed.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) 510(k) (gmbs: --Over The Counter Use
000069
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.