K030985 · Vital Signs, Inc. · BZD · Oct 15, 2003 · Anesthesiology
Device Facts
Record ID
K030985
Device Name
BREAS PV 10I CPAP SYSTEM, MODEL PV 10I
Applicant
Vital Signs, Inc.
Product Code
BZD · Anesthesiology
Decision Date
Oct 15, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Breas PV 10i CPAP System is intended to deliver Continuous Positive Airway Pressure (CPAP) therapy for the treatment of adult Obstructive Sleep Apnea (OSA).
Device Story
Breas PV10i CPAP System provides CPAP therapy for OSA treatment. Device inputs include patient breathing patterns via airflow sensor and output pressure via pressure transducer. Microprocessor-controlled blower (brushless DC motor) adjusts pressure based on firmware-based flow signal processing. Operates in constant CPAP mode or self-adjusting mode (CPAP 'i'). Used in clinical settings (hospitals, sleep labs) or home environments; prescribed by physicians. Clinicians set parameters (start/limit pressures) via device panel, optional Remote Control Module, or PC-based software. Pressure transducer enables automatic altitude compensation. Output affects clinical decision-making by maintaining airway patency; benefits patients by reducing apnea-hypopnea events and improving comfort through self-adjusting pressure.
Clinical Evidence
Two clinical studies conducted. Study 1: 12-patient randomized crossover comparing PV10i, ResMed AutoSet T, and constant CPAP. No significant differences in AHI (6.8 vs 4.0 vs 1.8); PV10i applied lower mean pressure (6.7 cmH2O) than AutoSet T (8.4 cmH2O, p<0.05). Patients preferred self-adjusting mode (p<0.01). Study 2: 100-patient randomized evaluation comparing four self-adjusting devices. PV10i showed efficacy in reducing AHI from 55.6 to 6.3 (titration) and 7.5 (6-month follow-up), comparable to predicates.
Technological Characteristics
Microprocessor-controlled blower with brushless DC motor. Pressure range 4-20 cmH2O. Includes pressure and airflow sensors. Dimensions: 6.3 x 4.5 x 9.3 inches; weight: 3.7 lbs. Power: 100-240 VAC or 12-24 VDC. Connectivity: PC-based software for parameter setting/compliance tracking. Non-sterile. Firmware-based flow signal processing for self-adjusting pressure.
Indications for Use
Indicated for adult patients with Obstructive Sleep Apnea (OSA) requiring Continuous Positive Airway Pressure (CPAP) therapy. Not intended for life support, life-sustaining applications, or transport of critical care patients.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/3 description: The image shows a close-up, angled view of a handwritten or printed text string. The text appears to be a combination of letters and numbers, with the sequence "K03098" clearly visible. The characters are bold and dark, contrasting with the lighter background. The angle and close proximity suggest the image is focused on highlighting this specific code or identifier.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
| Submitter: | OCT 1 5 2003<br>Vital Signs, Inc.<br>20 Campus Road<br>Totowa, NJ 07512-1200 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Anthony P. Martino<br>V. P. Quality Assurance & Regulatory Affairs<br>Phone: (973) 790-1330, ext. 356<br>Fax: (973) 790-4150 |
| Summary Date: | July 16, 2003 |
| Name of Device: | Breas PV10i CPAP System |
| Common Names: | CPAP System<br>Self-adjusting CPAP System<br>Auto-adjusting CPAP System |
| Classification Name: | Non-continuous ventilator (21 CFR 868.5905) |
| Product Code: | 73 BZD |
Predicate Devices:
Breas PV100 CPAP System (K001553)
ResMed Sullivan Autoset® T (K970516, K970771, and K984428)
.
{1}------------------------------------------------
## Device Description:
In the treatment of sleep disorders, CPAP is well established and common practice as a means to prevent upper airways from collapsing and therefore avoid breathing problems associated with airway collapse and obstruction. There are a number of devices legally marketed in the United States for this application, including self-adjusting CPAP systems such as the predicate ResMed Sullivan Autoset T. The self-adjusting systems accommodate pressure requirements that may vary (e.g., as patients change bodies positions or enter different sleep stages).
The Breas PV10i CPAP is similar to the predicate Breas PV100 CPAP System, but with the addition of self-adjusting pressure features. Both Breas devices are microprocessor controlled based on firmware (embedded software). Both designs incorporate pressure sensors that monitor output pressure and provide feedback to the microprocessors. The PV10i also incorporates an airflow sensor and flow signal processing means in firmware to provide self-adjusting capability.
The therapy delivered by the Breas PV10i CPAP System can be either:
- 1) Self-adjusting in response to variations in patient breathing patterns (CPAP "i" mode), or
- 2) Set to a constant pressure level (constant CPAP mode).
The primary hardware component is a blower that generates airflow. The blower assembly consists of a brushless DC motor that drives a fan, entraining ambient air through a filter and pressurizing it to provide the prescribed airflow with CPAP for the patient. The microprocessor controls the motor and hence the speed of the fan.
The PV10i pressure range is 4 - 20 cmH20. Starting pressure and self-adjusting pressure limits are settable by clinical personnel in accordance with the patient's prescription.
The CPAP airflow is delivered via a single lumen outlet tube that may be connected to various non-invasive patient interfaces, such as nasal masks. To minimize CO2 rebreathing, masks or other interfaces permitting a leak flow of at least 12 liters/minute at the minimum output pressure setting of 4 cm of H20 are recommended for use with the PV10i.
The pressure transducer incorporated in the PV10i continuously monitors output pressure to the patient and reference ambient pressure. This enables the device to automatically compensate for altitude changes.
The PV10i has an auto-switching power supply that facilitates use in conjunction with international travel (100 - 240 VAC). It can also be used with an external 12 - 24 VDC power source when AC mains line voltage is not available. This DC power source
{2}------------------------------------------------
capability is not intended as a battery power back-up system for any critical treatment applications. An accessory Remote Control Module may be connected that enables clinicians to operate the PV10i without using the control and setting panels on the device itself. A PC-based software program may also be supplied that provides an optional means for clinicians to set patient parameters (e.g., start and limit pressures) and track patient compliance.
The PV10i and its accessories are not sterile. The outer dimensions of the PV10i housing are 6.3 × 4.5 × 9.3 inches, and the device weighs 3.7 pounds.
## Intended Use:
The Breas PV10i CPAP System is intended to deliver Continuous Positive Airway Pressure (CPAP) therapy for the treatment of adult Obstructive Sleep Apnea (OSA).
## Comparison of Use and Technological Characteristics:
The PV10i may be used in clinical settings (e.g., hospitals, sleep laboratories, sub-acute care institutions) and home environments and must be prescribed by a physician. It is not intended for life support or life sustaining applications or for transport of critical care patients.
As compared with the cited predicate devices, the Breas PV10i CPAP System has:
- Same intended uses
Same environments of use
Similar design (microprocessor-controlled blower as air source)
Same technology (software based flow signal processing means to provide self-adjusting pressure capability)
Same materials (with particular reference to the air flow pathway)
The differences that do exist are minimal and involve primarily user preference features. These features are described in labeling for the device that includes an Operator Manual.
## Summary of Performance Testing:
- 1. Non-clinical testing was conducted to verify that the Breas PV10i CPAP System is capable of meeting its stated performance specifications and that all Risk Analysis issues have been appropriately addressed. The device passed all tests.
{3}------------------------------------------------
- 2. Testing was conducted to demonstrate compliance with applicable requirements in the November 1993 draft "Reviewer Guidance for Premarket Notification Submissions" published by the FDA's Division of Cardiovascular, Respiratory, and Neurological Devices. The testing included but was not limited to:
- Electrical Safety testing per IEC 601 ●
- Electromagnetic Compatibility testing (EMC testing) .
- Mechanical Safety testing .
- Environmental testing ●
- Functional testing ●
The PV10i device passed all tests.
- 3. All device software (the embedded software and the optional PC software) was documented and tested in accordance with the FDA's May 29, 1998 "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices". The PV10i software passed all tests.
- 4. Two clinical studies were performed as part of the process of validating the software algorithm for the PV10i CPAP System's self-adjusting pressure features. They involved direct comparisons with predicate devices.
- In a three-way randomized crossover study involving 12 patients (of whom 11 A) completed all treatments) with symptomatic obstructive sleep apnea, the PV10i in self-adjusting mode was compared with the predicate ResMed AutoSet®T (K980721) and with conventional constant CPAP. Efficacy evaluation was based on analysis of sleep disordered breathing and sleep parameters assessed by nocturnal polysomnography. Further study outcomes were quality of life assessment and subjective preference of each device with respect to comfort.
No significant differences were noted in sleep or ventilatory variables between the devices (mean apnea-hypopnea index (AHI) 6.8, 4.0 and 1.8 with PV10i, Autoset T and constant CPAP respectively). The PV10i applied lower mean pressure (6.7) as compared with the Autoset T (8.4) (p<0.05), and both selftitrating devices applied significantly lower mean pressure as compared with constant CPAP (9.4) (p<0.01). One patient could not be treated effectively with either self-adjusting pressure device due to increased arousability from This patient was effectively treated with manually titrated CPAP. sleep. Patients preferred self-adjusting CPAP pressure over constant pressure CPAP treatment (10 vs. 1, p<0.01).
- In a randomized treatment study evaluation involving initially 100 patients B) with apnea-hypopnea syndrome, four self-adjusting CPAP devices from
{4}------------------------------------------------
different manufacturers and conventional constant CPAP were compared. The study included comparison of results for a titration night and for a period of home use, with follow-up after three and six months. The self-adjusting CPAP devices included in this study were the Breas PV10i, the ResMed AutoSet®T (K980721), the Mallinckrodt GoodKnight® 418P (K993584), and the Weinmann SomnoSmart®.
Conclusions were based on apnea-hypopnea indices (AHI), Epworth sleepiness scores, and patient compliance. The PV10i patient group had a mean AHI of 55.6 initially, of 6.3 after titration and of 7.5 after 6 months. The comparable values were 56.5, 5.5 and 3.6 respectively for the predicate AutoSet T device and 48.5, 4.3 and 3.4 for the predicate Goodnight 418P. All four of the selfadjusting CPAP devices included in the study were judged to be efficient (efficacious) for home treatment.
#### Conclusions:
The Breas PV 10i CPAP System meets its stated performance specifications and criteria outlined in the Reviewer Guidance publications referenced above. Non-clinical test results and clinical studies demonstrate performance substantially equivalent to predicate devices. We conclude that the device is capable of operating safely in its intended environments and will be effective in fulfilling its intended use.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three swooping lines representing its body and wings. The eagle's head is facing left. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# OCT 1 5 2003
Mr. Anthony P. Martino Vice President of Quality Assurance & Regulatory Affairs Vital Signs, Incorporated 20 Campus Road Totowa, New Jersey 07512-1200
Re: K030985
Trade/Device Name: Breas PV10i CPAP System Regulation Number: 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: July 23, 2003 Received: July 24, 2003
Dear Mr. Martino:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket, approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2 - Mr. Martino
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Ques
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number:
K030982
Device Name:
Breas PV 10i CPAP System -
Indications for Use:
The Breas PV 10i CPAP System is intended to deliver Continuous Positive Airway Pressure (CPAP) therapy for the treatment of adult Obstructive Sleep Apnea (OSA).
signature
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital, Division of Ancolnoontal Devices KO 309 B
510(k) Number:_
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
X Prescription Use (Per 21 CFR 801.109)
Over-the-Counter Use
(Optional Format 1-2-96)
Page 6
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.