GE 3.0T GENERAL PURPOSE FLEX COIL

K030953 · Ge Medical Systems, Inc. · MOS · Apr 10, 2003 · Radiology

Device Facts

Record IDK030953
Device NameGE 3.0T GENERAL PURPOSE FLEX COIL
ApplicantGe Medical Systems, Inc.
Product CodeMOS · Radiology
Decision DateApr 10, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

The GE 3.0T General Purpose Flex Coil is designed to facilitate MR imaging of the Brachial Plexus, Unilateral Hip and Thigh. The GE 3.0T General Purpose Flex Coil will facilitate dynamic MR imaging of the joints of lower extremity (Knee and Ankle). This coil consists of two 5" x 6.5" surface coils coupled as a Helmholtz pair in a flexible package that can be wrapped around the anatomy of interest. This flexibility allows for the coil to be applied to a wider range of patients when compared to a similar coil in a rigid package. The coil is an option to the GE Signa® 3.0T whole body MR System.

Device Story

The GE 3.0T General Purpose Flex Coil is a flexible surface coil accessory for the GE Signa 3.0T whole body MR system. It consists of two 5" x 6.5" surface coils configured as a Helmholtz pair. The device is wrapped around the patient's anatomy to capture magnetic resonance signals during imaging procedures. It is operated by trained MRI technologists or radiologists in a clinical setting. The coil facilitates high-resolution imaging of the brachial plexus, hip, thigh, knee, and ankle. By providing a flexible form factor, it accommodates a wider range of patient sizes compared to rigid coils. The output is raw MR signal data processed by the host MR system to generate diagnostic images, which are interpreted by radiologists to inform clinical decision-making.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Flexible surface coil; Helmholtz pair configuration; two 5" x 6.5" coils; retuned for 3.0T magnetic resonance imaging; includes two passive blocking networks for safety; compatible with GE Signa 3.0T MR system.

Indications for Use

Indicated for MR imaging of the Brachial Plexus, Unilateral Hip, and Thigh, and dynamic MR imaging of lower extremity joints (Knee and Ankle).

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Kó 3.0953 Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The border has a decorative, swirling design around the letters. GE Medical Systems Page 1 of 2 General Electric Company P.O. Box 414, Milwaukee, WI 53201 APR 1 0 2003 ## 510(k) Summary This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92(c). | Submitter: | GE Medical Systems<br>PO Box 414<br>Milwaukee, WI 53201 | |-----------------|---------------------------------------------------------| | Contact Person: | Larry A. Kroger Ph.D.<br>Manager, Regulatory Programs | | Telephone: | 262- 544-3894 | | Fax: | 262- 548-4768 | | Date Prepared: | March 24.2003 | #### Device Name: GE 3.0T General Purpose Flex Coil Magnetic Resonance Coil, 21 CFR 892.1000, 90-MOS #### Marketed Device: The GE 3.0T General Purpose Flex Coil is substantially equivalent to the currently marketed GE Medical Systems 1.5T GP Flex coil M1085GP (K923264) #### Device Description: The GE 3.0T General Purpose Flex Coil is a modification of the existing 1.5T GP Flex Coil (K923264), to increase the field strength from 1.5T to 3.0T. In addition, two passive blocking networks are added to improve reliability and patient safety. #### Indications for Use: The GE 3.0T General Purpose Flex Coil is designed to facilitate MR imaging of the Brachial Plexus, Unilateral Hip and Thigh. The GE 3.0T General Purpose Flex Coil will facilitate dynamic MR imaging of the joints of lower extremity (Knee and Ankle). This coil consists of two 5" x 6.5" surface coils coupled as a Helmholtz pair in a flexible package that can be wrapped around the anatomy of interest. This flexibility allows for the coil to be applied to a wider range of patients when compared to a similar coil in a rigid package. The coil is an option to the GE Signa® 3.0T whole body MR System. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The logo is black and white. General Electric Company P.O. Box 414, Milwaukee, WI 53201 ## Comparison with Predicate Device: The GE 3.0T General Purpose Flex Coil is a modification of the existing cleared 1.5T General Purpose Flex Coil (K923264), with the main difference being the coil has been retuned for operation in a 3.0T magnetic field instead of a 1.5T field strength. In addition, two passive blocking networks are added to improve reliability, and patient safety. ## Summary of Studies: Testing was performed to demonstrate that the design modifications to the 1.5T general Purpose Flex coil meet predetermined acceptance criteria. ## Conclusion: It is the opinion of GE that the 3.0T General Purpose Flex Coil is substantially equivalent to the GE 1.5T general Purpose Flex coil (K923264). Usage of the GE 3.0T General Purpose Flex Coil does not result in any new potential hazards. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure with three flowing lines above it, symbolizing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the central image. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # APR 1 0 2003 Larry Kroger, Ph.D. Senior Regulatory Programs Manager GE Medical Systems P.O. Box 414 MILWAUKEE WI 53201 Re: K030953 Trade/Device Name: GE 3.0T General Purpose Flex Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 MOS Dated: March 24, 2003 Received: March 27, 2003 Dear Dr. Kroger: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050, {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive. Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure Page 2 {4}------------------------------------------------ 510(k) Number (if known): _ 大 o 3 Device Name: GE 3.0T General Purpose Flex Coil Indications For Use: The GE 3.0T General Purpose Flex Coil is designed to facilitate MR imaging of the Brachial Plexus, Unilateral Hip and Thigh. The GE 3.0T General Purpose Flex Coil will facilitate dynamic MR imaging of the joints of lower extremity (Knee and Ankle). This coil consists of two 5" x 6.5" surface coils coupled as a Helmholtz pair in a flexible package that can be wrapped around the anatomy of interest. This flexibility allows for the coil to be applied to a wider range of patients when compared to a similar coil in a rigid package. The coil is an option to the GE Signa® 3.0T whole body MR System. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | <div style="text-align:left;"></div> | |----------------------|--------------------------------------| | (Per 21 CFR 801.109) | | OR Over-The-Counter Use(Division Sign-Off) Division of Reproductive, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...