← Product Code [IYN](/productcode/IYN) · K030934

# GE LOGIQ 9 MODEL 2375600 (K030934)

_Ge Medical Systems, Inc. · IYN · Apr 17, 2003 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K030934

## Device Facts

- **Applicant:** Ge Medical Systems, Inc.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Apr 17, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transvaginal; and Intraoperative (abdominal, thoracic, vascular and neurosurgical).

## Device Story

GE LOGIQ 9 is a mobile, general-purpose diagnostic ultrasound system; provides digital acquisition, processing, and display of ultrasound images. System includes console, keyboard, specialized controls, and CRT/LCD display. Operates via various transducers to perform B-mode, M-mode, PW/CW Doppler, Color Doppler, Power Doppler, and harmonic imaging. Used by qualified physicians in clinical settings for diagnostic imaging and fluid flow analysis. Output allows clinicians to visualize internal anatomy and hemodynamics, aiding in diagnosis and clinical decision-making. Benefits include non-invasive assessment of various organ systems and vascular structures.

## Clinical Evidence

No clinical data required. Substantial equivalence supported by non-clinical testing, including acoustic output measurements, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, and mechanical safety testing in accordance with applicable standards.

## Technological Characteristics

Mobile console (64x90x140-160 cm); digital acquisition/processing; CRT and LCD display. Supports multiple transducer types (e.g., 3.5C, 3.5Cs, M7C, 8C, E8C, 7L, M8L, 10L, i12L, M12L, 3S, 4S, 7S, 10S, 2D, 6D). Modes: B, M, PW/CW Doppler, Color/Power Doppler, Harmonic, Coded Pulse. Conforms to 21 CFR 820, ISO 9001:2000, and ISO 13485. Safety standards: acoustic output, thermal, electrical, mechanical.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- GE LOGIQ 9 Diagnostic Ultrasound System ([K011188](/device/K011188.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K030934

APR 1 7 2003

# Attachment B:

Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).

Image /page/0/Picture/5 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular shape. The logo is black and white.

GE Medical Systems

General Electric Company P.O. Box 414, Milwaukee, WI 53201

#### Section a):

| 1. Submitter:       | GE Medical Systems<br>PO Box 414<br>Milwaukee, WI 53201                                                                                                                              |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:     | Allen Schuh,<br>Manager, Safety and Regulatory Engineering<br>Telephone: 414-647-4385; Fax: 414-647-4090                                                                             |
| Date Prepared:      | March 21, 2003                                                                                                                                                                       |
| 2. Device Name:     | GE LOGIQ 9 Diagnostic Ultrasound System, BT03<br>Ultrasonic Pulsed Echo Imaging System, 21 CFR 892.1560, 90-IYO<br>Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550, 90-ΙΥΝ |
| 3. Marketed Device: | GE LOGIQ 9 Diagnostic Ultrasound System K011188 (90-IYO/IΥΝ)                                                                                                                         |

A device currently in commercial distribution.

4. Device Description: The GE LOGIQ 9 is a full featured general purpose diagnostic ultrasound system. It consists of a mobile console approximately 64 cm wide, 90 cm deep and 140-160 cm (adjustable) high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls and a color video CRT and LCD display. This modification will provide users with additional probe options, improved user interface and image enhancement.

5. Indications for Use: The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transvaginal; and Intraoperative (abdominal, thoracic, vascular and neurosurgical).

6. Comparison with Predicate Device: The GE LOGIQ 9 BT03 is of a comparable type and substantially equivalent to the current GE LOGIQ 9. It has the same technological characteristics, key safety and effectiveness features, physical design, construction, and has the same intended uses and basic operating modes as the predicate device.

#### Section b):

1. Non-clinical Tests: The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been found to conform with applicable medical device safety standards.

#### 2. Clinical Tests: None required.

3. Conclusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. The design and development process of the manufacturer conforms with 21 CFR 820, ISO 9001:2000 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore, it is the opinion of GE Medical Systems that the GE LOGIQ 9 BT03 Diagnostic Utrasound is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 7 2003

Mr. Allen Schuh Manager, GE Ultrasound Safety and Regulation Engineering GE Medical Systems P.O. Box 414 MILWAUKEE WI 53201

Re: K030934 Trade Name: GE LOGIQ 9 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 12 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: March 21, 2003 Received: March 25, 2003

Dear Mr. Schuh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with. the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the GE LOGIQ 9, as described in your premarket notification:

Transducer Model Number

3.5C
3.5Cs
M7C
8C

{2}------------------------------------------------

E8C 7L M8L 10L i12L M12L ਤੇ ਫ 48 78 102 2D 6D

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed 

{3}------------------------------------------------

predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

If you have any questions regarding the content of this letter, please contact Rodrigo C Perez at (301) 594-1212.

Sincerely yours,

David St. Lynn

for

V Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{4}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

## GE LOGIQ 9 Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical<br>Application       | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                    |       |
|-------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|--------------------|-------|
|                               | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Code<br>d<br>Pulse | Other |
| Anatomy/Region of Interest    |                   |   |               |               |                  |                    |                  |                   |                     |                    |       |
| Ophthalmic                    | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Fetal / Obstetrics            | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Abdominal[1]                  | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Pediatric                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Small Organ[2]                | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Neonatal Cephalic             | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Adult Cephalic                | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Cardiac[3]                    | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Peripheral Vascular           | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Musculo-skeletal Conventional | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Musculo-skeletal Superficial  | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Other[4]                      | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Exam Type, Means of Access    |                   |   |               |               |                  |                    |                  |                   |                     |                    |       |
| Transesophageal               |                   |   |               |               |                  |                    |                  |                   |                     |                    |       |
| Transrectal                   | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Transvaginal                  | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                    |       |
| Transuretheral                |                   |   |               |               |                  |                    |                  |                   |                     |                    |       |
| Intraoperative[5]             | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Intraoperative Neurological   | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P                  |       |
| Intravascular                 |                   |   |               |               |                  |                    |                  |                   |                     |                    |       |
| Laparoscopic                  |                   |   |               |               |                  |                    |                  |                   |                     |                    |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes renal, GYN/Pelvic

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] Other use includes Urology/Prostate

[5] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV).

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel A. Asam

(Division Sign-Off) Division of Reproductive, Al and Radiological Dev 510(k) Number _

{5}------------------------------------------------

### GE LOGIQ 9 with 3.5C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                              | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|--------------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical<br>Application<br><i>Anatomy/Region of Interest</i> | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                                   |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                           | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              | P     |
| Abdominal[1]                                                 | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              | P     |
| Pediatric                                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Small Organ (specify)                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Neonatal Cephalic                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                               |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac                                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                          | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              | P     |
| Musculo-skeletal Conventional                                |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other[4]                                                     | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              | P     |
| Exam Type, Means of Access                                   |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                               |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                                |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraarterial                                                |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN;

[4] Other use includes Urology;

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Bowman

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device 510(k) Number _

{6}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

### GE LOGIQ 9 with 3.5Cs Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                       | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|-------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical<br>Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                    | N                 | N | N             |               | N                | N                  | N                | N                 | N                   | N              |       |
| Abdominal[1]                                          | N                 | N | N             |               | N                | N                  | N                | N                 | N                   | N              |       |
| Pediatric                                             |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Small Organ (specify)                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Neonatal Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac                                               |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                   | N                 | N | N             |               | N                | N                  | N                | N                 | N                   | N              |       |
| Musculo-skeletal Conventional                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other[4]                                              | N                 | N | N             |               | N                | N                  | N                | N                 | N                   | N              |       |
| Exam Type, Means of Access                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| anarosconic                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN;

[4] Other use includes Urology;

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David R. Ingram

(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number __

{7}------------------------------------------------

## GE LOGIQ 9 with M7C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical<br>Application       | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|-------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Anatomy/Region of Interest    | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics            | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Abdominal                     | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Pediatric                     | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Small Organ[2]                | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Neonatal Cephalic             |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular           | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Musculo-skeletal Conventional |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Superficial  |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Other (specify)               |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Exam Type, Means of Access    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal               |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                   |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                  |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative Neurological   |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                 |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                  |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid.

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Lyman

(Division Sign-Off) Division of Reproductive, Abdon and Radiological Devices 510(k) Number_

{8}------------------------------------------------

## GE LOGIQ 9 with 8C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                        | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|--------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical<br>Application<br>Anatomy/ Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                             |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Abdominal[1]                                           | E                 | E | E             |               | E                | E                  | E                | E                 | E                   | E              |
| Pediatric                                              | E                 | E | E             |               | E                | E                  | E                | E                 | E                   | E              |
| Small Organ (specify)                                  | E                 | E | E             |               | E                | E                  | E                | E                 | E                   | E              |
| Neonatal Cephalic                                      | E                 | E | E             |               | E                | E                  | E                | E                 | E                   | E              |
| Adult Cephalic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                                |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Conventional                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Superficial                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Other[4]                                               |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Exam Type, Means of Access                             |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)                               |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative Neurological                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN/Pelvic;

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Lyman

(Division Sign-Off) Division of Reproductive, Abdomi and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

{9}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

## GE LOGIQ 9 with E8C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                        | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|--------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical<br>Application<br>Anatomy/ Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                             |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                     | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Abdominal[1]                                           | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Pediatric                                              |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Small Organ (specify)                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Neonatal Cephalic                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                                |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Conventional                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Superficial                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Other[4]                                               | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Exam Type, Means of Access                             |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                            | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Transvaginal                                           | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Transuretheral                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)                               |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative Neurological                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN/Pelvic;

[4] Other use includes Urology/Prostate;

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Syverson

(Division Sign-Off) Division of Reproductive, Abdo and Radiological Devices 510(k) Number _

{10}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form

### GE LOGIQ 9 with 7L Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                              | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|--------------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical<br>Application<br><i>Anatomy/Region of Interest</i> | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                                   |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                           | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Abdominal                                                    | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Pediatric                                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Small Organ(2)                                               | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Neonatal Cephalic                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                               |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Musculo-skeletal Conventional                                | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Musculo-skeletal Superficial                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Other[4]                                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Exam Type, Means of Access                                   |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                              |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                               |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative Neurological                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                                |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid.

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Syverson

(Division Sign-Off) Division of Reproductive, Ab and Radiological Devices 510(k) Number _

{11}------------------------------------------------

## GE LOGIQ 9 with M8L Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                       | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|-------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical<br>Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                    | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |
| Abdominal                                             | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |
| Pediatric                                             |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Small Organ[2]                                        | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |
| Neonatal Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                               |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular                                   | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |
| Musculo-skeletal Conventional                         | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |
| Musculo-skeletal Superficial                          | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |
| Other[4]                                              |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Exam Type, Means of Access                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)                              | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |
| Intraoperative Neurological                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid.

[5] Intraoperative includes abdominal, thoracic, and vascular.

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Leppem

(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number _

{12}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

## GE LOGIQ 9 with 10L Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                       | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|-------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical<br>Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                    | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Abdominal                                             | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Pediatric                                             | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Small Organ[2]                                        | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Neonatal Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                               |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular                                   | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Musculo-skeletal Conventional                         | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Musculo-skeletal Superficial                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Other (specify)                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Exam Type, Means of Access                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative[5]                                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Intraoperative Neurological                           | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Intravascular                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Lanarosconic                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid.

[5] Intraoperative includes abdominal, thoracic, and vascular.

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Lynn

(Division Sign-Off) Division of Reproductive. ang Radiological Devices 510(k) Number _

{13}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

## GE LOGIQ 9 with i12L Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/ Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
|-----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Ophthalmic                                          |   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                  |   |   |               |               |                  |                    |                  |                   |                     |                |
| Abdominal[1]                                        | P | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Pediatric                                           | P | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Small Organ (specify)                               | P | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Neonatal Cephalic                                   |   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                      |   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac[3]                                          | P | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Peripheral Vascular                                 | P | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Musculo-skeletal Conventional                       | P | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Musculo-skeletal Superficial                        | P | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Other (specify)                                     |   |   |               |               |                  |                    |                  |                   |                     |                |
| Exam Type, Means of Access                          |   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                     |   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                         |   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                        |   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                      |   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative[5]                                   | P | P | P             |               | P                | P                  | P                | P                 | P                   | P              |
| Intraoperative Neurological                         |   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                       |   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                        |   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal is via Intraoperative;

[3] Cardiac is Adult and Pediatric via Intraoperative;

[5] Intraoperative includes abdominal, thoracic, and vascular.

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(Please Do NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David Lee Lynn

(Division Sign-Off) Division of Reproductive, and Radiological Device 510(k) Number

{14}------------------------------------------------

## GE LOGIQ 9 with M12L Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                       | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |  |
|-------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|--|
| Clinical<br>Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |  |
| Ophthalmic                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Fetal / Obstetrics                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Abdominal                                             |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Pediatric                                             | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |  |
| Small Organ[2]                                        | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |  |
| Neonatal Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Adult Cephalic                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Cardiac                                               |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Peripheral Vascular                                   | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |  |
| Musculo-skeletal Conventional                         | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |  |
| Musculo-skeletal Superficial                          | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |  |
| Other (specify)                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Exam Type, Means of Access                            |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Transesophageal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Transrectal                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Transvaginal                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Transuretheral                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Intraoperative [5] (specify)                          | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |  |
| Intraoperative Neurological                           |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Intravascular                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |  |
| Laparosconic                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid.

[5] Intraoperative includes abdominal, thoracic, and vascular.

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel A. Lyons

(Division Sign-Off) Division of Reproductive, Ab and Radiological Devices 510(k) Number

{15}------------------------------------------------

### GE LOGIQ 9 with 3S Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical<br>Application<br>Anatomy/Region of Interest | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|-------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
|                                                       | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                    | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Abdominal[1]                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |
| Pediatric                                             | P                 | P | P             | P             | P                | P                  |…

---

**Source:** [https://fda.innolitics.com/device/K030934](https://fda.innolitics.com/device/K030934)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
