COLLAGEN TOPICAL WOUND DRESSING
K030921 · Collagen Matrix, Inc. · KGN · May 15, 2003 · SU
Device Facts
| Record ID | K030921 |
| Device Name | COLLAGEN TOPICAL WOUND DRESSING |
| Applicant | Collagen Matrix, Inc. |
| Product Code | KGN · SU |
| Decision Date | May 15, 2003 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic |
Intended Use
Collagen Topical Wound Dressing is indicated for the management of moderately to heavily exudating wounds and to control minor bleeding. Collagen Topical Wound Dressing may be used for the management of exudating wounds such as: Pressure ulcers . Venous stasis ulcers . Diabetic ulcers . Acute wounds, for example trauma and surgical wounds . Partial-thickness burns .
Device Story
Collagen Topical Wound Dressing is a sterile, single-use, white to off-white, absorbent, microfibrillar particulate collagen matrix. Applied topically to wounds to manage exudate and control minor bleeding. Used in clinical settings for various wound types including pressure, venous, diabetic, acute, and burn wounds. Device functions as a physical matrix to support wound management.
Clinical Evidence
Bench testing only. Biocompatibility evaluated per ISO 10993-1 standards. No clinical data presented.
Technological Characteristics
Microfibrillar particulate collagen matrix. Sterile, single-use. Biocompatibility tested per ISO 10993-1.
Indications for Use
Indicated for management of moderately to heavily exudating wounds and control of minor bleeding in patients with pressure ulcers, venous stasis ulcers, diabetic ulcers, acute wounds (trauma/surgical), and partial-thickness burns.
Predicate Devices
- Medifil® Kollagen Particles (K910944)
- Collatek® Powder (K012990)
- HeliDerm™ Collagen Wound Dressing (K990086)
- hyCurc® Advanced Collagen Wound Care (K955506)
- FIBRACOL™ Collagen-Alginate Dressing (K925548)
Related Devices
- K040211 — MODIFICATION TO: COLLAGEN TOPICAL WOUND DRESSING · Collagen Matrix, Inc. · Feb 27, 2004
- K092805 — COLLAGEN SPONGE · Innocoll Pharmaceuticals · Feb 16, 2010
- K213341 — Fibrillar Collagen Wound Dressing · Collagen Matrix, Inc. · May 31, 2022
- K040558 — MODIFICATION TO: COLLAGEN TOPICAL WOUND DRESSING · Collagen Matrix, Inc. · Mar 17, 2004
- K070269 — MODIFICATION TO COLLAWOUND DRESSING · Collamatrix Co., Inc. · Mar 2, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
# MAY 1 5 2003
KOJO9ZI
## 510(k) Summary
| Applicant Name and Address: | Collagen Matrix, Inc.<br>509 Commerce Street<br>Franklin Lakes, New Jersey 07417 |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Peggy Hansen, RAC<br>Director, Clinical, Regulatory, and Quality Assurance<br>Tel: (201) 405-1477<br>Fax: (201) 405-1355 |
| Date of Summary: | March 21, 2003 |
| Device Common Name: | Collagen Topical Wound Dressing |
| Device Trade Name: | To be determined |
| Device Classification Name: | Bandage, Liquid<br>Class I<br>880.5090<br>KMF |
| Predicate Device(s): | Medifil® Kollagen Particles, K910944<br>Collatek® Powder, K012990<br>HeliDerm™ Collagen Wound Dressing, K990086<br>hyCurc® Advanced Collagen Wound Care, K955506<br>FIBRACOL™ Collagen-Alginate Dressing, K925548 |
### Description of the Device
The Collagen Topical Wound Dressing is a white to off-white, absorbent, microfibrillar particulate collagen matrix intended for topical use. The product is supplied sterile and for single use only.
#### Indications for Use
Collagen Topical Wound Dressing is indicated for the management of moderately to heavily exudating wounds and to control minor bleeding.
Collagen Topical Wound Dressing may be used for the management of cxudating wounds such as:
- Pressure ulccrs
- . V cnous stasis ulcers
- Diabetic ulcers .
{1}------------------------------------------------
- Acute wounds, for cxample trauma and surgical wounds ●
- Partial-thickness burns
#### Summary/Comparison of Technical Characteristics
Collagen Topical Wound Dressing and its predicates have similar technological characteristics. In particular, the Collagen Topical Wound Dressing and its predicates are similar with respect to intended use, material, form, size, etc.
#### Safety
Collagen Topical Wound Dressing has been evaluated by a number of tests to assess its safety/biocompatibility. The device passed all applicable ISO 10993-1 testing for the biological evaluation of medical devices.
#### Conclusion
The results of the in vitro product characterization studies and biocompatibility studies show that the Collagen Topical Wound Dressing is safe and substantially equivalent to its predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.
Public Health Service
MAY 1 5 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Peggy Hansen, RAC Director, Clinical, Regulatory, and Ouality Assurance Collagen Matrix, Inc. 509 Commerce Street Franklin Lakes, New Jersey 07417
Re: K030921
Trade/Device Name: Collagen Topical Wound Dressing Regulatory Class: Unclassified Product Code: KGN Dated: March 21, 2003 Received: March 24, 2003
Dear Ms. Hansen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Peggy Hansen, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
| 510(k) Number (if known): | | - " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " | |
|---------------------------|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
|---------------------------|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
Device Name:
Indications for Use:
Collagen Topical Wound Dressing is indicated for the management of moderately to heavily exudating wounds and to control minor bleeding.
Collagen Topical Wound Dressing may be used for the management of exudating wounds such as:
- Pressure ulcers .
- Venous stasis ulcers .
- Diabetic ulcers .
- Acute wounds, for example trauma and surgical wounds
- Partial-thickness burns .
Muriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K030921
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div></div> |
|----------------------|-------------|
| (Per 21 CFR 801.109) | <div></div> |
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)