← Product Code [LNH](/productcode/LNH) · K030874

# GE 3.0T SIGNA INFINITY TWINSPEED WITH EXCITE MR SYSTEM (K030874)

_General Electric Co. · LNH · Mar 28, 2003 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K030874

## Device Facts

- **Applicant:** General Electric Co.
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Mar 28, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The GE Signa® Infinity line of products are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio and short scan times. The Signa® Infinity System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa® Infinity systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

## Device Story

The Signa 3.0T Infinity TwinSpeed with Excite is a whole-body MRI system utilizing a 3.0T superconducting magnet. It acquires 2D single/multi-slice and 3D volume images in various planes (sagittal, coronal, axial, oblique, double oblique) using pulse sequences and Fourier transformation techniques. Operated by trained clinicians in a hospital or imaging center, the system provides high-resolution anatomical images and spectroscopic data. Physicians interpret these outputs to assist in clinical diagnosis. The system features eight independent receive channels and a dual-mode gradient coil to improve signal-to-noise ratio and reduce scan times compared to previous iterations.

## Clinical Evidence

Bench testing only. The device was evaluated for safety and performance according to NEMA standards and IEC 601-1 and IEC 601-2-33 international medical equipment safety standards. No clinical data was provided.

## Technological Characteristics

3.0T superconducting magnet; 8 independent receive channels; dual-mode gradient coil; 2D/3D Fourier transformation image reconstruction; supports transverse, sagittal, coronal, and oblique imaging planes; multinuclear spectroscopy capability. Complies with IEC 601-1 and IEC 601-2-33 safety standards.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Signa 3.0T MR system ([K022397](/device/K022397.md))
- Signa 3.0T MR system ([K003575](/device/K003575.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'GE' intertwined within a circular frame. The letters are stylized and connected, creating a monogram-like appearance. A small sprig or leaf-like element extends from the top of the circle, adding a distinctive touch to the design. The logo is presented in black and white, with the letters and frame appearing in black against a white background.

Image /page/0/Picture/1 description: The image contains a sequence of handwritten alphanumeric characters. The sequence appears to be "K030874". The characters are written in a simple, slightly irregular style, suggesting they were written quickly or without strict adherence to a specific font.

#### GE Medical Systems

General Electric Company P.O. Box 414, Milwaukee, WI 53201

MAR 2 8 2003

## 510(k) Summary

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92(c).

| Submitter:      | GE Medical Systems<br>PO Box 414<br>Milwaukee, WI 53201 |
|-----------------|---------------------------------------------------------|
| Contact Person: | Larry A. Kroger Ph.D.<br>Manager, Regulatory Programs   |
| Telephone:      | 262- 544-3894                                           |
| Fax:            | 262- 548-4768                                           |
| Date Prepared:  | March 10, 2003                                          |

#### Device Name:

GE 3.0T Signa® Infinity TwinSpeed with Excite Magnetic Resonance System Magnetic Resonance Diagnostic System, 21 CFR 892.1000, 90-LNH

#### Marketed Device:

The Signa® 3.0T Infinity TwinSpeed with Excite MR System is substantially equivalent to the currently marketed Signa® 3.0T MR system (K022397) with the main differences being an addition of eight independent receive channels and an update to the magnet design to increase the bore diameter.

#### Device Description:

The Signa® 3.0T Infinity TwinSpeed with Excite Magnetic Resonance System is a modification to the Signa® 3.0T Magnetic Resonance System (K022397) which utilizes a superconducting magnet to acquire 2D single-slice and multi-slice, and 3D volume images. The Signa® 3.0T Infinity Twin Speed with Excite Magnetic Resonance System features a superconducting magnet operating at 3.0T. The system can image in the sagittal, coronal, axial, oblique and double oblique planes, using various pulse sequences. Images are acquired and reconstructed using 2D and 3D Fourier transformation techniques. The system is intended for high-resolution anatomical applications, short scan times, and multinuclear spectroscopy.

#### Indications for Use:

The GE Signa® Infinity line of products are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio and short scan times. The Signa® Infinity System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa "Infinity systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circular border. The letters and border are black, contrasting with the white background.

General Electric Company P.O. Box 414, Milwaukee, WI 53201

resonance. These images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis.

#### Comparison with Predicate Device:

The Signa® 3.0T Infinity TwinSpeed with Excite MR System is a modification of the Signa 3.0T MR system (K022397) and (K003575) with the main differences being the addition of eight independent receive channels, the addition of the dual mode gradient coil, and a change to the magnet dimensions.

#### Summary of Studies:

The Signa® 3.0T Infinity TwinSpeed with Excite Magnetic Resonance System was evaluated to the appropriate NEMA performance standards as well as the IEC 601-1 International Medical Equipment Safety standard and IEC 601-2-33 Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis. The Signa® 3.0T Infinity TwinSpeed with Excite Magnetic Resonance System is comparable to the currently marketed Signa® 3.0T Magnetic Resonance System.

#### Conclusion:

It is the opinion of GE that the Signa® 3.0T Infinity TwinSpeed with Excite Magnetic Resonance System is substantially equivalent to the Signa® 3.0T Magnetic Resonance System. Usage of the Signa® Infinity TwinSpeed with Excite Magnetic Resonance System does not result in any new potential hazards.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes around a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### MAR 2 8 2003

Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager GE Medical Systems P.O. Box 414. W-400 MILWAUKEE WI 53201

Re: K030874 Trade/Device Name: GE 3.0T Signa® Infinity TwinSpeed with Excite MR System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 LNH Dated: March 19, 2003 Received: March 20, 2003

Dear Dr. Kroger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx. 1xxx                        | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

Page 2

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined and enclosed within a circular border. The letters and the border are in black, contrasting with the white background.

GE Medical Systems

General Electric Company P.O. Box 414, Milwaukee, WI 53201

## STATEMENT OF INTENDED USE

510(k) Number (if known): __ K 6 3 0 8 7 4 ___________________________________________________________________________________________________________________________________

Device Name: GE 3.0T Signa® Infinity TwinSpeed with Excite MR System

# Indications for Use

The GE Signa® Infinity line of products are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio and short scan times. The Signa® Infinity System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa® Infinity systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801-109)

OR

Over-The-Counter Use

David R. Ingram

(Division Sign-Off) Division of Reproductive, and Radiological Devic 510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K030874](https://fda.innolitics.com/device/K030874)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
