← Product Code [JJY](/productcode/JJY) · K030801

# BAYER ADVIA 1650 SPECIAL CHEMISTRY CONTROL (K030801)

_Bayer Healthcare, LLC · JJY · Apr 9, 2003 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K030801

## Device Facts

- **Applicant:** Bayer Healthcare, LLC
- **Product Code:** [JJY](/productcode/JJY.md)
- **Decision Date:** Apr 9, 2003
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Special Chemistry Controls are assayed control materials for in vitro diagnostic use to monitor the precision and accuracy of certain chemistry test procedures for the ADVIA 1650 Chemistry analyzer.

## Device Story

Special Chemistry Controls are lyophilized, human serum-based quality control materials with non-serum constituents. Used in clinical laboratory settings to monitor precision and accuracy of chemistry test procedures on the Bayer ADVIA 1650 Chemistry analyzer. Requires reconstitution with 5.0 mL distilled water. Analytes include Acid Phosphatase, Lactate, Pancreatic Amylase, and Lipase. Healthcare providers use the control results to verify system performance before patient sample analysis; ensures reliability of clinical diagnostic testing.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Lyophilized human serum-based control material; requires reconstitution with 5.0 mL distilled water. Designed for use on the ADVIA 1650 Chemistry analyzer. No electronic or software components.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- Bayer Special Chemistry Control ([K023840](/device/K023840.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) Number is:

## Submitter's Name and Address

Bayer Healthcare LLC 511 Benedict Avenue Tarrytown, NY 10591 Establishment Registration Number: 2432235

Contact Person: Andres Holle Telephone: 914-524-3494 Fax: 914-524-2500 e-mail: andres.holle.b(@bayer.com

### Contract Manufacturer

Randox Laboratories 55 Diamond Road Crumlin, County Antrim, UK Establishment Registration: 8020890

Device Name:

Special Chemistry Control

Proprietary/Trade Name

Common Name: Classification Name:

Bayer ADVIA 1650 Special Chemistry Control Quality Control Material Enzyme Controls (assayed and unassayed)

Classification: Class I Regulation Number: 21 CFR 862.1660 Panel: Chemistry (75) Product Code: JJY Predicate Device: Bayer Special Chemistry Control Premarket Notification Number: K023840

{1}------------------------------------------------

#### Device Description:

The Bayer Special Chemistry Controls are two separate levels of quality control material prepared from human serum with nonserum constituents added.

All the analytes currently in the calibrator and control material are: Acid Phosphatase Lactate Pancreatic Amylase Lipase

The intention of this submission is to add the assigned values to the labeling claims for: Pancreatic Amylase Lipase

### Intended Use:

The Special Chemistry Controls are assayed control materials for in vitro diagnostic use to monitor the precision and accuracy of certain chemistry test procedures for the ADVIA 1650 Chemistry analyzer.

#### Substantial Equivalence:

The Special Chemistry Controls are identical in intended use, storage and handling, stability, source material (human serum), and instructions for use as the previously cleared Special Chemistry Controls. The only difference in these controls is the addition of the assigned values in the labeling of two new analytes: Pancreatic Amylase and Lipase.

As with the predicate device, the control materials are lyophilized and require reconstitution with 5.0 mL distilled water. These controls are only for use on the Bayer ADVIA 1650 Chemistry Analyzer.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its wings and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

APR a 9 2003

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Andres Holle Manager, Regulatory Affairs Bayer Healthcare LLC 511 Benedict Avenue Tarrytown. NY 10591

> k030801 Trade/Device Name: Special Chemistry Control Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed). Regulatory Class: Class I Product Code: JJY Dated: March 10, 2003 Received: March 13, 2003

Dear Mr. Holle:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Special Chemistry Control

# Indications for Use:

For in vitro diagnostic use in the control of ADVIA 1650 Chemistry system for certain chemistry methods.

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE, IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation |  |
|--------------------------------------------------|--|
| (ODE)                                            |  |

  

| Prescription Use     | <div style="text-align:left;">✓</div> | OR | Over-The-Counter  |
|----------------------|---------------------------------------|----|-------------------|
| Use                  |                                       |    |                   |
| (Per 21 CFR 801.109) |                                       |    | (Optional Format) |
| 1-2-96)              |                                       |    |                   |

  

|                                         | <div style="text-align:left;">Sean Cooper</div> |
|-----------------------------------------|-------------------------------------------------|
| (Division Sign-Off)                     |                                                 |
| Division of Clinical Laboratory Devices |                                                 |
| 510(k) Number                           | A030801                                         |

---

**Source:** [https://fda.innolitics.com/device/K030801](https://fda.innolitics.com/device/K030801)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
