IMMULISA ANTI-BETA2 GLYCOPROTEIN I (B2GP1) IGG ELISA

K030701 · Immco Diagnostics, Inc. · MSV · Sep 9, 2003 · Immunology

Device Facts

Record IDK030701
Device NameIMMULISA ANTI-BETA2 GLYCOPROTEIN I (B2GP1) IGG ELISA
ApplicantImmco Diagnostics, Inc.
Product CodeMSV · Immunology
Decision DateSep 9, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5660
Device ClassClass 2

Indications for Use

An enzyme linked immunoassay (ELISA) for the detection and semi-quantitation of IgG antibodies to β₂-GPI, as an aid in assessing the risk of thrombosis in patients with Systemic Lupus Erythromatosus (SLE) or lupus like disorders.

Device Story

ImmuLisa Anti-β₂ GPI IgG ELISA is a solid-phase immunoassay for semi-quantitative detection of IgG antibodies to β₂-GPI in human serum. The device utilizes microplates coated with β₂-GPI antigen. Patient serum samples, calibrators, and controls are incubated in wells; antibodies bind to the antigen. After washing, an alkaline phosphatase-labeled anti-human IgG conjugate is added, followed by pNPP substrate. The resulting color change, proportional to antibody concentration, is measured via spectrophotometer at 405 nm. The test is performed in clinical laboratory settings by trained personnel. Results assist clinicians in evaluating thrombosis risk in patients with SLE or lupus-like disorders.

Clinical Evidence

No clinical data provided in the document.

Technological Characteristics

Solid-phase ELISA; β₂-GPI antigen-coated microplate; alkaline phosphatase-labeled anti-human IgG conjugate; pNPP substrate; spectrophotometric detection at 405 nm. Requires microplate reader. Semi-quantitative output in EU/mL.

Indications for Use

Indicated for the detection and semi-quantitation of IgG antibodies to ß2-GP1 in patients with Systemic Lupus Erythematosus (SLE) or lupus-like disorders to aid in assessing thrombosis risk.

Regulatory Classification

Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: K030701 B. Analyte: IgG Anti-β₂ Glycoprotein I C. Type of Test: Semi-quantitative (ELISA) D. Applicant: IMMCO Diagnostics E. Proprietary and Established Names: ImmuLisa Anti-β₂ GPI IgG ELISA F. Regulatory Information: 1. Regulation section: 21 CFR 866.5660 Multiple Autoantibodies Immunological Test System 2. Classification: Class II 3. Product Code: MSV; B2-Glycoprotein I (B2-GPI) Antibodies Test System 4. Panel: 82 Immunology G. Intended Use: 1. Indication(s) for use: An enzyme linked immunoassay (ELISA) for the detection and semi-quantitation of IgG antibodies to β₂-GPI, as an aid in assessing the risk of thrombosis in patients with Systemic Lupus Erythromatosus (SLE) or lupus like disorders. 2. Special condition for use statement(s): NA 3. Special instrument Requirements: Microplate reader capable of reading absorbance values at 405 nm. If dual microplate reader is available, the reference filter should be set at 600-650 nm H. Device Description: The ImmuLisa Anti-Beta₂ Glycoprotein I (β₂GPI) IgG ELISA test kit contains a microplate coated with β₂GPI, calibrators, positive and negative controls, enzyme-labeled conjugate, substrate and substrate stop solution, buffered diluent and wash buffer. I. Substantial Equivalence Information: 1. Predicate device name(s): INOVA Quanta Lite anti-β₂ GPI IgG Antibody Test kit {1} 2. Predicate K number(s): K970551 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Detection and semi-quantitation of IgG antibodies to B2-GPI | Same | | Methodology | Enzyme linked immunosorbent assay | Same | | Sample Dilution | 1:101 | Same | | Differences | | | | Item | Device | Predicate | | Conjugate | Alkaline Phosphatase labeled | Horseradish peroxidase labeled | | Calibrators | Set of 4 pre-diluted calibrators | Set of 5 pre-diluted calibrators | | Substrate | p-NPP | TMB | | Absorbance | 405 nm | 450 nm | # J. Standard/Guidance Document Referenced (if applicable): NA # K. Test Principle: The test is performed as a solid phase immunoassay (ELISA) in $\beta_{2}$ - GPI coated microwells. Controls, calibrators and patient serum samples are incubated in the antigen coated microwells to allow antibodies present in the serum to bind. Unbound antibody and other serum proteins are removed by washing the microwells. Antibodies bound to the microwells are detected by adding an enzyme labeled anti-human IgG conjugate to the microwells. These enzyme-conjugated antibodies bind specifically to the human immunoglobulin of the appropriate class. Unbound enzyme-labeled conjugate is removed by washing. Specific enzyme substrate (pNPP) is then added to the wells and the presence of antibodies is detected by a color change produced by the conversion of the pNPP substrate. The reaction is stopped and the intensity of color change, which is proportional to the concentration of antibody, is read by a spectrophotometer # L. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: Three samples with known concentrations of anti $\beta_{2}$ -GPI were assayed in 10 replicates over a two week period. This study included negatives (less than $20\mathrm{EU / mL}$ ), weak positives and other values extended through the high-end calibrator range of $160\mathrm{EU / mL}$ . Intra- and inter-assay coefficients of variation (CV) were as follows: {2} | | Sample 1 | Sample 2 | Sample 3 | | --- | --- | --- | --- | | Inter-assay CV | 8.7% | 5.0% | 10.7% | | Intra-assay CV | 7.3% | 13.9% | 5.6% | # b. Linearity/assay reportable range: Linearity was established in a variety of calibrator studies during development. Absorbance values versus calibration were established in a linear to linear plot using known values of calibrators using standard statistical methods. The r-squared value for the curve is 0.997. Graph showed linearity up to 337 EU/mL # c. Traceability (controls, calibrators, or method): Positive controls and calibrators were derived from serum obtained from various commercial plasma centers. Samples were selected on the basis of the specific antibody reactivity and the concentration. For assignment of values, the samples were tested at various dilutions on at least two different lots of the $\beta_{2}$ -GPI antigen coated plates # d. Detection limit: 0.543 EU/mL The limits of detection fall well below the established cutoff. # e. Analytical specificity: Interference studies were provided using positive specimens diluted into hemolyzed and lipemic serum. Slight depression of O.D. was noted in specimens diluted in normal serum. The product insert included a statement warning users against the use of grossly hemolyzed or lipemic sera. # f. Assay cut-off: Less than 20 EU/mL. Cut-off data is provided. It was calculated using Mean +2 standard deviations. # 2. Comparison studies: # a. Method comparison with predicate device: The IMMCO ImmuLisa Anti-Beta $_2$ Glycoprotein I ( $\beta_2$ GPI) IgG assay was compared with INOVA Anti $\beta_2$ -GPI IgG ELISA as a predicate device. A range of positives with antibody levels spanning from low positive to high positive values, disease controls from patients with syphilis, and normal patients were tested. Patients with syphilis may be positive for cardiolipin but negative for B2GPI antibodies. {3} Page 4 of 5 | | ImmuLisa Anti B2GPI IgG | | | | --- | --- | --- | --- | | INOVA Anti-B2-GPI IgG ELISA | Positive | Negative | Total | | | 17 | 0 | 17 | | | 1 | 74 | 75 | | | 18 | 74 | 92 | Relative Agreement: 99% Relative Sensitivity: 100% Relative Specificity: 99% b. Matrix comparison: Both tests are for human serum only 3. Clinical studies: a. Clinical sensitivity: NA b. Clinical specificity: It is known that patients suffering from infectious diseases like syphilis have cardiolipin antibodies and lupus anti-coagulant. To test the clinical specificity of the assay, sera from patients with syphilis were tested for the presence of anti-cardiolipin and anti-B2GPI antibodies. The study suggests that B2-GPI antibodies are more specific than anti-cardiolipin antibodies as only a small percentage of the patients with syphilis are positive for B2-GPI. c. Other clinical supportive data (when a and b are not applicable): Clinical studies of anti-B2GPI antibodies have been predicated upon the relationship between the occurrence of B2GPI antibodies and anti-cardiolipin antibodies in Anti-phospholipid Syndrome (APS) patient populations. A reference demonstrating the correlation is provided. 4. Clinical cut-off: Negative <20 Borderline 20-25 Positive >25 5. Expected values/Reference range: The normal range was established by testing 64 serum samples from apparently healthy donors obtained from the local Red Cross. The mean plus three SD of the mean of this normal was used to determine the cut off of the normal to the abnormal. This was assigned an arbitrary unit value of 20 EU/mL {4} # M. Conclusion: Based on the review of the information provided in this 510(k), the ImmuLisa Anti-Beta $_2$ Glycoprotein I ( $\beta_2$ GPI) IgG appears to be Substantially Equivalent to devices regulated under 21CFR 866.5660, Multiple Autoantibodies Immunological Test System, Product code MSV, Class II.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...