XIVE 3.0 DENTAL IMPLANT SYSTEM

K030639 · Friandent GmbH · DZE · Aug 12, 2003 · Dental

Device Facts

Record IDK030639
Device NameXIVE 3.0 DENTAL IMPLANT SYSTEM
ApplicantFriandent GmbH
Product CodeDZE · Dental
Decision DateAug 12, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The XiVE 3.0 Dental Implant System is indicated for single tooth restorations and splinted tooth restorations in the region of 7 to 10 and 23 to 26.

Device Story

XiVE 3.0 Dental Implant System consists of subgingival threaded dental implants (3.25mm diameter, 11-15mm length) with FRIOS Deep Profile Surface. System includes surgical/laboratory instruments and prosthetic components (e.g., EstheticBase, AuroBase, Select, Telescopic Abutments). Used by clinicians in two-stage procedures for single or splinted tooth replacement in anterior regions. Implants are surgically placed into the jawbone to provide a foundation for prosthetic restorations. Benefits include restoration of tooth function and aesthetics in the anterior mouth.

Clinical Evidence

No clinical data. Evidence consists of bench testing, including compressive and static strength testing and finite element analysis.

Technological Characteristics

Screw-type endosseous dental implant; 3.25mm diameter; 11-15mm length; constructed of CP-2 titanium; FRIOS Deep Profile Surface coating. Prosthetic options include EstheticBase, AuroBase, Select, and Telescopic Abutments. Mechanical testing performed via compressive/static strength and finite element analysis.

Indications for Use

Indicated for single and splinted tooth restorations in the anterior regions of the mouth (teeth 7-10 and 23-26).

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ FRIADENT GmbH XiVE® 3.0 Dental Implant System AUG 1 2 2003 Original Premarket 510(k) Notification ## SECTION 16: # SUMMARY OF SAFETY AND EFFECTIVENESS This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92. #### SUBMITTER INFORMATION 16.1 | a. | Company Name: | FRIADENT GmbH. | |----|--------------------------------------|------------------------------------------------------| | b. | Company Address: | Steinzeugstrasse 50<br>Mannheim D-68229<br>Germany | | c. | Company Phone:<br>Company Facsimile: | (011) 49 06 21 4 86 1549<br>(011) 49 06 21 4 86 1866 | | d. | Contact Person: | Heike Dietzler<br>Regulatory Affairs Manager | | e. | Date Summary Prepared: | February 27, 2003 | #### DEVICE IDENTIFICATION 16.2. | a. Trade/Proprietary Name: | XiVE® 3.0 Dental Implant System | |----------------------------|-----------------------------------------------| | b. Classification Name: | Endosseous Dental Implants<br>21 CFR 872.3640 | #### IDENTIFICATION OF PREDICATE DEVICES 16.3 | Company | Device | 510(k) No. | Date Cleared | |---------------|----------------------------------------------------------------|------------|--------------| | FRIADENT GmbH | FRIALIT-2 3.4mm Dental<br>Implant with Deep Profile<br>Surface | K994376 | 03/24/2000 | #### 16.4 DEVICE DESCRIPTION The XiVE 3.0 Dental Implant System consists of subgingival threaded dental implants with a 3.25mm diameter and lengths of 11 -- 15mm . The implants are coated with the FRIOS Deep Profile Surface. The XiVE Dental Implant System {2}------------------------------------------------ is comprised of dental implants, surgical and laboratory instruments and prosthetic components. The system is designed for two stage procedures for single tooth replacement in the anterior regions of the mouth. #### ા રેજરી રહ્યું હતું. સંદર્ભ SUBSTANTIAL EQUIVALENCE The XiVE® 3.0 dental implant is substantially equivalent to the current FRIALIT-2® 3.4mm Dental Implant Systems in terms of design, materials, coatings, mechanical strength, prosthetic options and single tooth intended use. #### 16.6 INTENDED USE The XIVE 3.0 Dental Implant System is indicated for single tooth restorations in the region of 7 to 10 and 23 to 26. #### 16.7 TECHNOLOGICAL CHARACTERISTICS The XiVE® 3.0 dental implant is identical to the current FRIALIT-2® 3.4mm dental implants in terms of coatings, materials and prosthetic options. The XiVE® 3.0 dental implant is 3.25mm diameter screw-type implant with FRIOS® Deep Profile Surface. The lengths of the implants range from 11 - 15mm. The XiVE 3.0 dental implants are constructed of CP-2 titanium. A variety of prosthetic options are available for the XiVE 3.0 system including EstheticBase, AuroBase, Select and Telescopic Abutments. The XiVE 3.0 dental implant system was tested for compressive and static strength and finite element analysis. #### 16.8 CLASS III CERTIFICATION AND SUMMARY This notification contains a Class III certification and summary of adverse safety and effectiveness information pursuant to 513(f) of the Federal Food, Drug, and Cosmetic Act. {3}------------------------------------------------ #### 16.9 CONCLUSIONS This notification contains all information required by 21 CFR 807.87. A completed copy of the Premarket Notification 510(k) Reviewer's Checklist is provided in this submission. Performance evaluations of the XiVE 3.0 dental implant system show that the device performs as intended for the anterior region of the mouth. Comparison of the XiVE 3.0 dental implant system to the predicate device show that the device is substantially equivalent. {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. AUG 1 2 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Friadent GmbH C/O Ms. Carol Patterson Patterson Consulting Group. Incorporated 21911 Erie Lane Lake Forest, California 92630 Re: K030639 Trade/Device Name: XiVE® 3.0 Dental Implant System Regulation Number: 872.3640 Regulation Name: Endosseous Implant Regulatory Class: III Product Code: DZE Dated: July 23, 2003 Received: July 25, 2003 Dear Ms. Patterson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ ### Page 2 - Ms. Patterson Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Suser Rums Susan Runner, DDS, MA, Interim Director Division of Anesthesiology, General Hospital Infection Control and Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## INDICATION FOR USE K030639 510(k) Number: Device Name: XiVE® 3.0 Dental Implant System Indications for Use: The XiVE 3.0 Dental Implant System is indicated for single tooth restorations and splinted tooth restorations in the region of 7 to 10 and 23 to 26. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use Kein Marley for MSN (Division Sign-Off) General Hospital. 510/ CONFIDENTIAL
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...