I-125 RAPID STRAND, MODEL 7000

K030594 · Amersham Health · KXK · Apr 10, 2003 · Radiology

Device Facts

Record IDK030594
Device NameI-125 RAPID STRAND, MODEL 7000
ApplicantAmersham Health
Product CodeKXK · Radiology
Decision DateApr 10, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.5730
Device ClassClass 2
AttributesTherapeutic

Indications for Use

RAPID Strand™ consists of absorbable seeding spacers and Model 6711 OncoSeed™ seeds, embedded within Polyglactin 910 Synthetic Absorbable Braided Carrier and is intended for medical purposes to be placed into a body cavity or tissue as a source of nuclear radiation for therapy.

Device Story

RAPID Strand™ is a brachytherapy device consisting of I-125 radioactive seeds (Model 6711 OncoSeed™) and absorbable seeding spacers arranged at fixed intervals within a Polyglactin 910 braided carrier. The device is stiffened and sterilized via Ethylene Oxide. It is delivered in a stainless steel shielding tube that attenuates >99% of I-125 photons. Physicians determine the required dose and seed count based on patient-specific prescriptions; the customized strand contains a variable number of seeds and spacers. The device is implanted into body cavities or tissue to provide localized nuclear radiation therapy for tumors. It serves as a permanent implant to deliver radiation directly to the tumor site, potentially sparing surrounding healthy tissue compared to external radiation sources.

Clinical Evidence

Bench testing only. Evidence consists of comparative analysis of device materials and infrared spectroscopy (FTIR) to confirm chemical composition and structural integrity of the modified RAPID Strand™ compared to the predicate device.

Technological Characteristics

Device consists of I-125 seeds (welded titanium capsule, silver rod) and absorbable spacers (90% Glycolide, 10% L-lactide, D&C Violet No. 2). Carrier is Polyglactin 910 (90% Glycolide, 10% L-lactide, coated with Polyglactin 370 and calcium stearate). Sterilization via Ethylene Oxide. Delivered in stainless steel shielding tube.

Indications for Use

Indicated for permanent interstitial implantation of localized tumors with low to moderate radiosensitivity. Used as primary treatment (e.g., prostate cancer, unresectable tumors) or for residual disease post-excision. May be used concurrently or following external beam radiation or chemotherapy.

Regulatory Classification

Identification

A radionuclide brachytherapy source is a device that consists of a radionuclide which may be enclosed in a sealed container made of gold, titanium, stainless steel, or platinum and intended for medical purposes to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.

Special Controls

*Classification.* Class II (special controls). A prostate seeding kit intended for use with a radionuclide brachytherapy source only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # APR 1 0 2003 ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Kp3p594 The following 510(k) Summary of Safety and Effectiveness is prepared and provided in accordance with the requirements of 21 CFR 807.92 as amended under the Safe Medical Devices Act of 1990 (SMDA). ## Submitter's Information | Company Name: | Medi-Physics, Inc DBA Amersham Health | |----------------------------|---------------------------------------------------------| | Address: | 101 Carnegie Center<br>Princeton, NJ 08540-6231 | | Contact Name: | Subhash Patel<br>Associate Director, Regulatory Affairs | | Telephone Number: | (609) 514-6846 | | Summary Prepared Date: | February 19, 2003 | | Subject Device Information | | | Trade Name: | RAPID Strand™ | | Model: | 7000 | | Common Name: | Radionuclide Brachytherapy Source | | |-----------------|-----------------------------------|----------------------| | Class: | II | | | Classification: | 21 CFR 892.5730 | Product Code: 90-IWI | Legally marketed devices to which equivalence is claimed. Predicate Devices | 1. | Trade Name: | RAPID Strand™ | Model: 7000 | |----|---------------------|------------------------------------------------------------|----------------------| | | Common Name: | Radionuclide Brachytherapy Source | | | | Class: | II | | | | Classification: | 21 CFR 892.5730 | Product Code: 90-IWI | | | Cleared 510(k) No.: | K940632 | | | | Submitted by: | Medi-Physics, Inc., Arlington Heights, IL | | | 2. | Trade Name: | OncoSeed™ | Model: 6711 | | | Common Name: | Radionuclide Brachytherapy Source | | | | Class: | II | | | | Classification: | 21 CFR 892.5730 | Product Code: 90-KXK | | | Cleared 510(k) No.: | K914281 | | | | Submitted by: | Medi-Physics, Inc., Arlington Heights, IL | | | 3. | Trade Name: | Coated VICRYL (Polyglactin 910) Synthetic Absorbable Sutur | | | | Common Name: | Absorbable Poly (glycolide/L-lactide) surgical suture | | | | Class: | II | | | | Classification: | 21 CFR 892.4493 | Product Code: GAM | | | Cleared 510(k) No.: | K022269 (September 13, 2002) | | | | Submitted by: | Ethicon. Inc.. Somerville. NJ | | {1}------------------------------------------------ Ky3\$\phi\$594 CP Medical, Portland, OR Trade Name: Common Name: 4. Class: Classification: Cleared 510(k) No .: Submitted by: Absorbable Seeding Spacers Accessory to applicator and accessory to radionuclide Brachytherapy Sources II 21 CFR 892.5730 Product Code: 90 KXK K013964 (January 24, 2002) Description of Device RAPID Strand™ consists of 11 absorbable seeding spacers and 10 Model 6711 OncoSeed™ seeds (welded titanium capsule containing I-125 adsorbed onto a silver rod) spaced at a fixed distance within absorbable braided carrier. The absorbable braided carrier containing OncoSeeds and absorbable seeding spacer is stiffened and then sterilized by Ethylene Oxide. The seeds are housed in a plastic spacing jig within a stainless steel shielding tube which attenuates >99% of the I-125 photons. RAPID Strand™ is STERILE when shipped. The Customized RAPID Strand allows the variable seed spacing to reflect individual patient prescriptions as specified by the treating physician. This Customized RAPID Strand will have less than ten (10) OncoSeed™ seeds and a variable number of Absorbable Seeding Spacers. The treating physician will determine the required dose and total number of active seeds for each Customized RAPID Strand. Since it will contain less than 10 Brachytherapy seeds, it does not raise any issues of safety. Please refer to Seed Spacing Diagram in Section-E #### Intended Use RAPID Strand™ consists of absorbable seeding spacers and Model 6711 OncoSeed™ seeds, embedded within Polyglactin 910 Synthetic Absorbable Braided Carrier and is intended for medical purposes to be placed into a body cavity or tissue as a source of nuclear radiation for therapy. #### Indications for Use RAPID Strand ™ is indicated for permanent interstitial implantation of selected localized turnors which are low to moderate radiosensitivity. They may be used either as primary treatment (such as prostate cancer or unresectable turnors) or for treatment of residual disease after excision of the primary tumor. RAPID Strand™ may be indicated for use concurrent with or at the completion of other treatment modalities, such as external beam radiation therapy or chemotherapy. When compared with the Intended Use and Indications for Use of the predicate model 7000 RAPID Strand™, the Intended Use and Indications for Use of the subject model 7000 RAPID Strand™ have not changed. {2}------------------------------------------------ y30594 #### Technological Characteristics The following modifications have been incorporated in the subject device - Model 7000 RAPID StrandTM when compared to predicate device: | Modified<br>Characteristics | 7000 RAPID Strand<br>(Subject Device) | 7000 RAPID Strand<br>(Predicate Device)<br>Cleared in 510(k)# K940632 | |----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------| | Absorbable Seeding Spacers | | | | Regulatory Status | Cleared in 510(k)# K013964 | Not applicable | | Number of Spacers | 11 (Eleven) | None | | Base Material | 90% Glycolide + 10% L-lactide | None | | Color | D&C Violet No.2<br>(21 CFR 74.3602) | None | | Coated VICRYL (Polyglactin 910) Braided Carrier Material | | | | Regulatory Status | Cleared in 510(k) # K022269 | Cleared in 510(k) # K940632 | | Base Material | Ethicon JB947 - Coated VICRYL<br>(Polyglactin 910) Synthetic<br>Absorbable Suture | Ethicon J617 - Coated VICRYL<br>(Polyglactin 910) Synthetic<br>Absorbable Suture | | Formulation | Polyglactin 910<br>[90% Glycolide + 10% L-lactide<br>coated with Polyglactin 370 and<br>calcium stearate] | Polyglactin 910<br>[90% Glycolide + 10% L-lactide<br>coated with Polyglactin 370 and<br>calcium stearate] | | Appearance | Undyed (Natural) | Coated Vicryl Violet (D&C Violet<br>#2) | | Sterilization | Sterile (EtO sterilization<br>performed on finished device) | Sterile (EtO sterilization<br>performed on finished device) | All other technological characteristics of the subject 7000 RAPID Strand™ remain unchanged compared to the predicate RAPID Strand™ that was cleared in the 510(k) # K940632. #### Nonclinical Test Data The copies of the following Nonclinical Test Data reports have been provided in the Section-G of this submission. - Report of Analysis: Infrared Spectroscopy FTIR Spectra of the predicate RAPID Strand 1. (finished device) by Northview Laboratories, Inc., Northbrook, IL - Report of Analysis: Infrared Spectroscopy FTIR Spectra of the subject RAPID Strand 1.5 2. (finished device) by Northview Laboratories, Inc., Northbrook, IL ### Conclusion: Upon reviewing the safety and effectiveness information provided in this submission and comparing the intended use, indications for use, method of use and other technological characteristics, it can be concluded that the subject RAPID Strand™ is substantially equivalent to the predicate RAPID Strand™, which was cleared under 510(k)# K940632. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them, resembling a wave or fabric. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## APR 1 0 2003 Mr. Subhash Patel Associate Director Regulatory Affairs Amersham Health 101 Carnegie Center PRINCETON NJ 08540 Re: K030594 Trade/Device Name: I-125 RAPID Strand Model 7000 Regulation Number: 21 CFR 892.5730 Regulation Name: Radionuclide brachytherapy source Regulatory Class: II Product Code: 90 KXK Dated: March 25, 2003 Received: March 27, 2003 Dear Mr. Patel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## EXHIBIT-2 ## Indications for Use Form Page 1 of 1 | <strong>Labels</strong> | <strong>Values</strong> | |---------------------------|-------------------------| | 510(k) Number (if known): | K\$\phi\$5\$\rho\$59i4 | Device Name: Model: 7000 RAPID Strand™ ## Indications For Use: RAPID Strand ™ is indicated for permanent interstitial implantation of selected localized tumors which are low to moderate radiosensitivity. They may be used either as primary treatment (such as prostate cancer or unresectable tumors) or for treatment of residual disease after excision of the primary tumor. RAPID Strand™ may be indicated for use concurrent with or at the completion of other treatment modalities, such as external beam radiation therapy or chemotherapy. #### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH2 Office of Device Evaluation (ODE) OR David H. Legum (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K030594 Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use _
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...