This device is a prescription device and only for symptomatic relief of chronic intractable pain. (For EMS function): Relaxation of muscle spasms. Prevention or retardation of disuse atrophy. Increasing local blood circulation. Muscle re-education. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis. Maintaining or increasing range of motion.
Device Story
Digi-Pro TENS series (models WL-2203B, WL-2204B, WL-2204B-P1, WL-2204B-P2, WL-2205B) are transcutaneous electrical nerve stimulators and powered muscle stimulators. Device consists of a stimulus generator and electrodes attached to patient skin. Microprocessor-controlled software generates adjustable electrical waveforms and output currents to stimulate nerves or muscles. Used for pain relief or muscle therapy. Operated by healthcare providers or patients (prescription required). Output affects muscle contraction or nerve signaling to achieve therapeutic goals like pain reduction, spasm relaxation, or circulation improvement.
Clinical Evidence
Bench testing only. Compliance with ANSI/AAMI NS4-1985, EN 60601-1, EN 60601-1-1, and EN 60601-1-2 standards. Software verification performed per FDA software guidance.
Technological Characteristics
Microprocessor-controlled electrical stimulator. Output: adjustable electrical current/waveform. Connectivity: wired electrodes. Standards: ANSI/AAMI NS4-1985, EN 60601-1, EN 60601-1-1, EN 60601-1-2. Software: microprocessor-based control unit.
Indications for Use
Indicated for patients requiring symptomatic relief of chronic intractable pain (TENS function) or muscle stimulation for relaxation of spasms, prevention/retardation of disuse atrophy, increased local blood circulation, muscle re-education, post-surgical calf muscle stimulation to prevent venous thrombosis, and maintenance/increase of range of motion (EMS function).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
K03053.7
### 510(K) SUMMARY
MAR 20 2003
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: ________
- 1. Submitter's Identifications: Well-Life Healthcare Inc. Room 6C01, No. 5, Sec. 5, Hsin Yi Rd., Taipei, Taiwan, R.O.C.
Contact: Ms. Grace Chang Sales Manager
Date of Summary Preparation: February 11, 2003.
- 2. Name of the Device:
Well-Life TENS (Transcutaneous Electrical Nerve Stimulation Device), Model Digi-Pro TENS series, including WL-2203B, WL-2204B, WL-2204B-P1, WL-2204B-P2, and WL-2205B,
- 3. Information of the 510(k) Cleared Device (Predicate Device): K021359 & K020314.
#### 4. Device Description:
The Digi-Pro TENS series, including WL-2203B, WL-2204B-P1, WL-2204B-P1, WL-2204B-P2, and WL-2205B are transcutaneous electrical nerve stimulator used for pain relief and/or powered muscle stimulator by applying an electrical current to electrodes, which are attached on the patient's skin. The output and waveform is adjustable according to the situation of patient.
Digi-Pro TENS series, models WL-2203B, WL-2204B, WL-2204B-P1, WL-2204B-P2, and WL-2205B, consist mainly of two parts: the stimulus generator, electrode. The stimulus qenerator generates the output current specified as the input of controller. The output port transmits the output current to the electrode, which is attached to the patient's skin so as to transmit this stimulus current to the patient for pain relief.
The stimulation mode for Digi-Pro TENS includes several different operation modes as mentioned on the comparison table. These operation modes are generated from the software control by using the microprocessor as its main control unit.
- 5. Intended Use:
On the instruction manual of each model, the intended uses and contraindication are defined very clearly. Please see the information of instruction manuals in clause 7.6 of this submission document.
In addition, the standard format for the statement of indications and contraindication for use are provided hereafter.
{1}------------------------------------------------
Well-Life Healthcare Inc.
Page_1__of__1_________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Digi-Pro TENS / Model: WL-2203B, WL-2204B, WL-2204B-P1, WL-2204B-P2, and WL-2205B.
## Indications For Use (Available for WL-2203B & 'TENS function' of WL-2205B):
This device is a prescription device and only for symptomatic relief of chronic intractable pain.
Indications For Use (Available for WL-2204B, 2204B-P1, 2204B-P2 & 'EMS function' of WL-2205B):
- Relaxation of muscle spasms. .
- Prevention or retardation of disuse atrophy. ●
- Increasing local blood circulation. ●
- . Muscle re-education.
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis. ●
- Maintaining or increasing range of motion. .
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 3-10-98)
{2}------------------------------------------------
Well-Life Healthcare Inc.
- 6. Comparison to the 510(k) Cleared Device (Predicate Device)
- (1) The new model WL-2203B is substantially equivalent to the Well-Life clear model Wil-2203 (K021359).
- (2) The new model WL-2204B, 2204B-P1, and 2204B-P2 are substantially equivalent to the Well-Life clear model WL-2204 (K020314).
- (3) The new model WL-2205B is substantially equivalent to the Well-Life clear model Wt.-2205 (K021359 & K020314).
- 7. Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards includes ANSI/AAMI, NS4-1985, as well as EN 60601-1, EN 60601-1-1, and EN 60601-1-2 requirement.
In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.
#### 8. Conclusions
The Digi-Pro TENS series, including WL-2203B, WL-2204B-P1, WL-2204B-P1, WL-2204B-P2, and WL-2205B, have the same intended use and technological characteristics as the cleared device of SD603/Model III, SD-605/program and Q-MS. Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device.
In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three stylized human profiles facing to the right, with three lines above them that resemble a bird in flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAR 2 0 2003
Well-Life Healthcare, Inc. C/O Tony C.S. Chang Wincet Consultant Co., Ltd. No. 5, Alley 5, Lane Cheng Hsing Chung Ching Rd., Pei Tun District Taichung, Taiwan, R.O.C.
Re: K030537
Trade/Device Name: Digi-Pro TENS Model WL-2203B Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator Regulatory Class: Class II Product Code: GZJ
Trade/Device Name: Digi-Pro TENS Models WL-2204B, WL-2204B-P1 and WL-2204B-P2 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF
Trade/Device Name: Digi-Pro TENS Model WL-2205B Regulation Number: 21 CFR 882.5890 and 21 CFR 890.5850 Regulation Name: Transcutaneous Electrical Nerve Stimulator and Powered Muscle Stimulator Regulatory Class: Class II Product Code: GZJ and IPF
Dated: February 11, 2003 Received: February 20, 2003
Dear Mr. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
{4}------------------------------------------------
#### Page 2 -- Mr. Tony C.S. Chang
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N Milkus
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Well-Life Healthcare Inc.
Page 1 _of_1
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Digi-Pro TENS / Model: WL-2203B, WL-2204B, WL-2204B-P1, WL-2204B-P2, and WL-2205B.
Indications For Use (Available for WL-2203B & 'TENS function' of WL-2205B):
This device is a prescription device and only for symptomatic relief of chronic intractable pain.
Indications For Use (Available for WL-2204B, 2204B-P1, 2204B-P2 & 'EMS function' of WL-2205B):
- Relaxation of muscle spasms. ●
- Prevention or retardation of disuse atrophy. .
- Increasing local blood circulation. .
- Muscle re-education. ●
:
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis. ●
- Maintaining or increasing range of motion. .
Mark N. Milkersen
1. Resto rive
710(k) Number _
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 3-10-98)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.