DEPILASE TWIN YAG LASER SYSTEM
K030510 · Depilase Group , Ltd. · GEX · May 14, 2003 · General, Plastic Surgery
Device Facts
| Record ID | K030510 |
| Device Name | DEPILASE TWIN YAG LASER SYSTEM |
| Applicant | Depilase Group , Ltd. |
| Product Code | GEX · General, Plastic Surgery |
| Decision Date | May 14, 2003 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4810 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The DEPILASE TWIN YAG Laser System is intended for the coagulation and haemostasis of vascular lesions (1064 and 532 nm wavelengths), the removal and permanent reduction of unwanted hair in Fitzpatrick skin type I - VI and the dermatological use for the non invasive treatment of facial wrinkles (1064 nm wavelength).
Device Story
The DEPILASE TWIN YAG Laser System is a flashlamp-pumped Nd:YAG laser for dermatological procedures. It utilizes an optical cavity with an Nd:YAG crystal, optionally frequency-doubled to 532 nm via a Second Harmonic Crystal (SHC). A red diode aiming beam is co-aligned via a beamsplitter. Energy is delivered via optical fiber to a focusing handpiece for non-contact treatment. The system includes a high-voltage power supply, water-to-air heat exchanger cooling, and a microprocessor-based controller for parameter selection. Operated by clinicians in a medical setting, the device allows precise control of laser pulses (up to 3 Hz). The output is used for vascular coagulation, hair reduction, and wrinkle treatment. The microprocessor ensures regulated power and thermal control, providing consistent energy delivery to target tissues, which facilitates clinical outcomes in dermatological applications.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and intended use comparison to the predicate device.
Technological Characteristics
Flashlamp-pumped Nd:YAG laser; 1064 nm fundamental wavelength, 532 nm frequency-doubled via Second Harmonic Crystal (SHC). Includes red diode aiming beam (Class I). Features high-voltage power supply, internal closed-loop water-to-air heat exchanger, and optical fiber delivery with focusing handpiece. Microprocessor-controlled. Pulse frequency up to 3 Hz. Non-contact operation.
Indications for Use
Indicated for coagulation and haemostasis of vascular lesions (1064/532 nm), permanent hair reduction (Fitzpatrick skin types I-VI), and non-invasive treatment of facial wrinkles (1064 nm).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
- Laserscope Lyra (K020021)
Related Devices
- K020697 — DEPILASE TWIN YAG LASER SYSTEM · Depilase Group , Ltd. · Apr 12, 2002
- K030506 — DEPILASE YAG LASE PLUS LASER SYSTEM · Depilase Group , Ltd. · May 9, 2003
- K020412 — DEPILASE TWIN LASE LASER SYSTEM · Depilase Group , Ltd. · Feb 22, 2002
- K020463 — DEPILASE YAG LASE PLUS LASER SYSTEM · Depilase Group , Ltd. · Feb 27, 2002
- K030508 — DEPILASE TRIO LASE LASER SYSTEM · Depilase Group , Ltd. · May 20, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 1 4 2003
K0 30510
1 ஆ 2
# Appendix E
#### 510 (k) Summary
- 1. General Information:
| Company: | Depilase Group Ltd<br>One Canada Square<br>Canary Wharf<br>London E14 5DY<br>United Kingdom |
|--------------------|---------------------------------------------------------------------------------------------|
| Contact Person: | Dr. Mario Luca Russo |
| Preparation: | 11-15-02 |
| Device Trade Name: | DEPILASE TWIN YAG Laser System |
| Common Name: | Long Pulsed Nd:YAG Surgical, Powered, Lase<br>79-GEX<br>21 CFR 878-48 |
#### I. Description
The DEPILASE TWIN YAG Laser System is based on Nd: YAG laser technology. Within the system, an optical cavity contains the Nd: YAG crystal, which is activated by means of the use of a flashlamp, with the option of frequency doubling the fundamental laser wavelength, from 1064 nm to 532 nm. The frequency doubling wavelength is obtained by passing with the laser radiation through a special crystal called Second Harmonic Crystal (SHC). After the cavity, a red diode aiming beam is reflected onto a coaxial beam path using a beamsplitter assembly. The combined therapeutic and aiming beams are guided down an optical fibre delivery system to a focusing handpiece. The laser is used in a non-contact mode.
The DEPILASE TWIN YAG Laser System is designed with 5 major sub-systems:
- a) A high voltage power supply which converts and rectifies the a.c. mains current to provide regulated power for the flashlamp simmer current and main triggering pulse.
- b) A cooling system consisting of an internal water flow circuit together with water to air heat exchanger.
- c) An Nd: YAG laser rod, capable of generating optical pulses at a frequency up to 3 Hz
- d) An optical delivery system, interfacing the energy from the laser to the patient via an optical fibre and focusing handpiece.
- e) The microprocessor based controller which regulates the functions of the laser and allows parameter selection by the user.
{1}------------------------------------------------
K0 30510 log 2
## II. Intended Use
The DEPILASE TWIN YAG Laser System is indicated for the use in dermatological applications for the non invasive treatment of facial wrinkles ( 1064 nm laser emission).
### III. Summary of Substantial Equivalence
Depilase believes that its DEPILASE TWIN YAG Laser System is substantially equivalent. regarding the 1064 nm laser emission , to the Laserscope Lyra (K020021) Nd: YAG laser. previously cleared for the use in dermatological applications for the non invasive treatment of facial wrinkles.
It therefore has the same Intended Use as the DEPILASE TWIN YAG Laser System.
Technologically, the predicate device has identical characteristics to the DEPILASE TWIN Y AG Laser System, comprising a flashlamp pumped Nd: Y AG laser rod generating light at a wavelength of 1064 nm, which is subsequently delivered to the patient via an optical fibre delivery system and focusing handpiece.
The DEPILASE TWIN YAG Laser System output characteristics are very similar to those of the predicate device.
All lasers are microprocessor controlled devices.
All lasers utilize Class I aiming beams which pose no hazard to the user.
All systems utilize an internal closed loop water-air heat exchange circuit for optimal thermal control of the laser cavity.
The risk and benefits of the DEPILASE TWIN YAG Laser System are comparable to the predicate device when used for similar clinical applications.
It is therefore believed that there are no new questions of Safety or Effectiveness raised by the introduction of this device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white seal for the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around the perimeter. In the center is a stylized image of an eagle with three lines representing its wings and body. The eagle faces to the right.
MAY 1 4 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Dr. Mario Luca Russo Director of Research and Development Depilase Group Ltd. One Canada Square Canary Wharf London E14 5DY United Kingdom
Re: K030510
Trade/Device Name: DEPILASE TWIN YAG Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: January 5, 2003 Received: February 19, 2003
Dear Dr. Russo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Dr. Mario Luca Russo
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# APPENDIX F
#### INDICATIONS FOR USE STATEMENT
K030510 510 (K) Number (if known): _
Device Name:
DEPILASE TWIN YAG LASER SYSTEM
Indications For Use:
The DEPILASE TWIN YAG Laser System is intended for the coagulation and haemostasis of vascular lesions (1064 and 532 nm wavelengths), the removal and permanent reduction of unwanted hair in Fitzpatrick skin type I - VI and the dermatological use for the non invasive treatment of facial wrinkles (1064 nm wavelength).
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative and Neurological Devices KO30510 510(k) Number Prescription Use OR Over The Counter Use (per 21 CFR 801.109)
Premarket Notification for DEPILASE TWIN YAG Laser System