3.0T TORSO PHASED ARRAY COIL
K030495 · Ge Medical Systems, Inc. · MOS · Mar 5, 2003 · Radiology
Device Facts
| Record ID | K030495 |
| Device Name | 3.0T TORSO PHASED ARRAY COIL |
| Applicant | Ge Medical Systems, Inc. |
| Product Code | MOS · Radiology |
| Decision Date | Mar 5, 2003 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1000 |
| Device Class | Class 2 |
| Attributes | 3rd-Party Reviewed |
Intended Use
The coil is indicated for use, on the order of a physician in conjunction with a 3.0T MR scanner, as an accessory to produce images of the abdominal and pelvic regions in 2D and 3D. The primary applications associated with evaluation of the anatomy are as follows: Thorax Abdomen Male and Female Pelvis Prostate
Device Story
The 3.0T Torso Phased Array Coil is a four-channel, receive-only MR coil accessory designed for use with GE 3.0T MR systems. It captures magnetic resonance signals from the thorax, abdomen, and pelvic regions (including prostate) to produce 2D and 3D diagnostic images. The device is operated by trained medical personnel in a clinical MR environment. By providing higher field-of-view (FOV) coverage compared to previous 1.5T iterations, it enables high-resolution imaging of abdominal and pelvic anatomy, assisting physicians in clinical diagnosis and evaluation of disease states.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Four-channel phased array receive-only MR coil; tuned to 127.72MHz for 3.0T operation; dimensions 34cm x 32cm; hardware-based signal reception; compatible with GE 3.0T MR systems.
Indications for Use
Indicated for patients requiring MR imaging of the thorax, abdomen, male/female pelvis, and prostate; used in conjunction with a 3.0T MR scanner under physician order.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- GE Medical Systems 1.5T Cardiac Phased Array (K971667)
- USA Instruments Torso and Pelvis Phased Array Coil (K001209)
Related Devices
- K073236 — NEOCOIL 3.0T 32-CHANNEL TORSO ARRAY COIL FOR GE · Neocoil · Nov 30, 2007
- K023679 — GORE 3.0T TORSO ARRAY · W. L. Gore & Associates, Inc. · Nov 18, 2002
- K041185 — TORSO ARRAY COIL, MODEL 558GE-64 · Igc-Medical Advances, Inc. · May 14, 2004
- K173446 — Contour 24 · Quality Electrodynamics, LLC · Nov 17, 2017
- K982772 — SIGNA PROFILE QUADRATURE THORACIC/LUMBAR SPINE PHASED ARRAY COIL · Ge Medical Systems, Inc. · Oct 15, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a series of handwritten numbers. The numbers are 103 0495. The numbers are written in black ink on a white background.
# MAR 0 5 2003
# 510(K) SUMMARY
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92(c).
| Submitter: | GE Medical Systems<br>PO Box 414<br>Milwaukee, WI 53201 |
|-----------------|---------------------------------------------------------|
| Contact Person: | Larry A. Kroger Ph.D.<br>Manager, Regulatory Programs |
| Telephone: | 262- 544-3894 |
| Fax: | 262- 548-4768 |
| Date Prepared: | January 31, 2003 |
#### Device Name:
3.0T Torso Phased Array Coil Magnetic Resonance Diagnostic System, 21 CFR 892.1000, 90-MOS
#### Marketed Device:
The 3.0T Torso Phased Array Coil is substantially equivalent to the currently marketed GE Medical Systems 1.5T Cardiac Phased Array (K971667) and USA Instruments Torso and Pelvis Phased Array Coil (K001209).
### Device Description:
The 3.0T Torso Phased Array Coil is a four-channel phased array receive only MR coil used in conjunction with the GE 3.0T MR system.
#### Indications for Use:
The coil is indicated for use, on the order of a physician in conjunction with a 3.0T MR scanner, as an accessory to produce images of the abdominal and pelvic regions in 2D and 3D.
The primary applications associated with evaluation of the anatorny are as follows:
- u Thorax
- . Abdomen
- 트 Male and Female Pelvis
- . Prostate
#### Comparison with Predicate Device:
The 3.0T Torso Phased Array Coil is a modification of the GE Medical Systems 1.5T Cardiac Phased Array Coil (K971667) with the main differences being that the 3.0T
{1}------------------------------------------------
Torso Phased Array Coil is tuned to 127.72MHz for operation at 3.0T instead of 63.86MHz for operation at 1.5T. In addition the dimensions of the 3.0T Torso Phased Array Coil (34cm wide and 32cm long) are larger than the dimensions of the 1.5T Cardiac Phased Array Coil for a higher FOV coverage requirement in case of the 3.0T Torso Phased Array coil. The technological similarities to the USA Instruments Torso and Pelvis Phased Array Coil (K001209) include similar applications and indications for use.
## Summary of Studies:
Testing was performed to demonstrate that the design of the 4 channel 3.0T Torso Phased Array Coil meet predetermined acceptance criteria.
#### Conclusion:
It is the opinion of GE that the 3.0T Torso Phased Array Coil is substantially equivalent to the currently marketed GE Medical Systems 1.5T Cardiac Phased Array Coil (K971667) and USA Instruments Torso and Pelvis Phased Array Coil (K001209). Usage of the 3.0T Torso Phased Array Coil does not result in any new potential hazards.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
GE Medical Systems % Mr. Heinz Joerg Steneberg TUV Rheinland of North America 12 Commerce Road NEWTON CT 06470
Re: K030495
MAR 0 5 2003
Trade/Device Name: 3.0T Torso Phased Array Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 MOS Dated: February 14, 2003 Received: February 19, 2003
Dear Mr. Steneberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{4}------------------------------------------------
Ko30495 510(k) Number (if known):
Device Name: 3.0T Torso Phased Array Coil
## Indications For Use:
The coil is indicated for use, on the order of a physician in conjunction with a 3.0T MR scanner, as an accessory to produce images of the abdominal and pelvic regions in 2D and 3D.
The primary applications associated with evaluation of the anatomy are as follows:
- Thorax 비
- 트 Abdomen
- 미 Male and Female Pelvis
- Prostate
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C Broadon
informati 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
OR Over-The-Counter Use