← Product Code [FII](/productcode/FII) · K030470

# NXSTAGE SYSTEM ONE (K030470)

_Nxstage Medical, Inc. · FII · Jul 7, 2003 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K030470

## Device Facts

- **Applicant:** Nxstage Medical, Inc.
- **Product Code:** [FII](/productcode/FII.md)
- **Decision Date:** Jul 7, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5860
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The NxStage System One is indicated for treatment of renal failure or fluid overload using hemofiltration, hemodialysis and/or ultrafiltration. All treatments must be administered by a health care provider, under physician prescription.

## Device Story

NxStage System One is an electro-mechanical hemodialysis system comprising the NxStage Cycler and a single-use NxStage Cartridge. The cartridge serves as an extracorporeal blood circuit and fluid management device, mounting integrally within the cycler. The system performs hemofiltration, hemodialysis, and ultrafiltration. Operated by healthcare providers in clinical settings, the device manages blood and fluid circuits to treat renal failure or fluid overload. The system design and configuration are based on established predicate hemodialysis technologies.

## Clinical Evidence

Bench testing only. Performance testing was conducted to characterize the device and provide a basis of comparison to predicate devices, demonstrating substantial equivalence.

## Technological Characteristics

Electro-mechanical hemodialysis system. Consists of a cycler and a single-use extracorporeal blood/fluid circuit cartridge. Designed using components common to predicate high permeability hemodialysis systems.

## Regulatory Identification

A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).

## Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”

## Predicate Devices

- NxStage Therapy System ([K020858](/device/K020858.md))
- B. Braun Diapact CRRT ([K973322](/device/K973322.md))
- Baxter (SPS), Model 550 ([K872364](/device/K872364.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K030470
page 1 of 2

## NxStage Medical, Inc. NxStage System One 510(k) Premarket Notification

# Section VIII: Summary of Safety and Effectiveness

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of SMDA 1990.

#### A. Submitter's Information:

JUL 0 7 2003

| Name:                | NxStage Medical, Inc.                                  |
|----------------------|--------------------------------------------------------|
| Address:             | 439 South Union Street, Suite 50<br>Lawrence, MA 01843 |
| Phone:               | (978) 687-4700                                         |
| Fax:                 | (978) 687-4800                                         |
| Contact Person:      | Norma LeMay<br>Sr. Regulatory Specialist               |
| Date of Preparation: | January 30, 2003                                       |

#### B. Device Name:

| Trade Name:          | NxStage System One                                                        |
|----------------------|---------------------------------------------------------------------------|
| Common/Usual Name:   | High Permeability Hemodialysis System                                     |
| Classification Name: | System, Dialysate Delivery, Sealed (21 CFR<br>875.5860, Product Code FII) |

#### C. Substantial Equivalence/Predicate Devices:

The proposed NxStage System One is substantially equivalent to the following legally marketed predicate devices previously cleared by FDA:

- NxStage Therapy System (K020858, cleared 04/17/02)
- B. Braun Diapact CRRT (K973322, cleared 11/10/98) .
- . Baxter (SPS), Model 550 (K872364, cleared 10/13/87)

{1}------------------------------------------------

# NxStage Medical, Inc. NxStage System One 510(k) Premarket Notification

**Section VIII: Summary of Safety and Effectiveness**

#### D. Device Description/Indications for Use:

The NxStage System One consists of the NxStage Cycler and the NxStage Cartridge Extracorporeal Blood and Fluid Circuit.

The NxStage Cycler is an electro-mechanical device that interfaces with the NxStage Cartridge. The NxStage Cartridge is a single-use extracorporeal blood circuit and fluid management device that mounts integrally within the NxStage Cycler. The system is designed to deliver hemofiltration, hemodialysis and/or ultrafiltration.

#### Indications for use:

The NxStage System One is indicated for treatment of renal failure or fluid overload using hemofiltration, hemodialysis and/or ultrafiltration. All treatments must be administered by a health care provider, under physician prescription.

#### E. Technological Characteristics:

The proposed device has the same technological characteristics and is similar in design and configuration as compared to the predicate devices. The proposed device is designed and assembled with components commonly found in the predicate devices.

#### F. Summary of Non-Clinical Test/Performance Testing

NxStage Medical believes that the information and data provided in this submission clearly describes the proposed device and demonstrates that the device is adequately designed for the labeled indications for use. Performance testing was conducted to characterize performance of the proposed NxStage System One to provide a basis of comparison to the predicate devices. Results of the performance testing have documented that the proposed NxStage System One is substantially equivalent to the predicate devices and is suitable for the labeled indications for use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle faces right and is enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL - 7 2003

NxStage Medical, Inc. c/o Intertek Testing Services Donald James Sherratt 70 Codman Hill Road BOXBOROUGH MA 01719

Re: K030470

Trade/Device Name: NxStage® System One Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Product Code: 78 FII Dated: July 3, 2003 Received: July 7, 2003

Dear Mr. Sherratt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE STATEMENT

Page 1 of 1

510(k) Number (if known):

510(k) Number (if known): `K030470

Device Name:

NxStage System One

Indications for Use:

The NxStage System One is indicated for treatment of renal failure or fluid overload using hemofiltration, hemodialysis and/or ultrafiltration. All treatments must be administered by a health care provider, under physician prescription.

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number Prescription Use Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ OR (Per 21 CFR 801.109)

---

**Source:** [https://fda.innolitics.com/device/K030470](https://fda.innolitics.com/device/K030470)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
