← Product Code [HLI](/productcode/HLI) · K030433

# STRATUSOCT WITH RNFL NORMATIVE DATABASE (K030433)

_Carl Zeiss Meditec, Inc. · HLI · May 1, 2003 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K030433

## Device Facts

- **Applicant:** Carl Zeiss Meditec, Inc.
- **Product Code:** [HLI](/productcode/HLI.md)
- **Decision Date:** May 1, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.1570
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The STRATUSOCT™ is a high resolution tomographic device for the viewing and axial cross sectional imaging of posterior ocular structures. It is used for in vivo imaging and measurement of the retinal nerve fiber layer, and optic disk. It is intended for use as a diagnostic device to aid in the detection and management of ocular diseases, including but not limited to macular edema, central serous retinopathy, and glaucoma. The STRATUSOCT™ with Retinal Nerve Fiber Layer ("RNFL") Normative Database is a quantitative tool for the comparison of retinal nerve fiber layer in the human retina to a database of known normal subjects. The RNFL Normative Database is intended for use as a diagnostic aid in the detection and management of ocular diseases, especially glaucoma.

## Device Story

Microprocessor-based low-coherence digital instrument; uses low-coherence interferometry to measure optical reflections/backscatter from internal ocular tissue. Generates high-resolution axial cross-sectional images of posterior ocular structures. Includes RNFL Normative Database for quantitative comparison of patient retinal nerve fiber layer measurements against known normal human subjects. Used by clinicians for diagnostic aid in detecting/managing ocular diseases like glaucoma. Output allows clinicians to visualize ocular microstructures and assess quantitative data against normative values to inform clinical decision-making.

## Clinical Evidence

Clinical data collected on a statistically significant number of normal human subjects to validate the RNFL Normative Database inclusion.

## Technological Characteristics

Microprocessor-based; low-coherence interferometry; optical beam sensing; digital imaging; RNFL Normative Database software integration.

## Regulatory Identification

An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

## Predicate Devices

- Humphrey® OCT 3 ([K012727](/device/K012727.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# MAY - 1 2003

## K030433

## 510(k) Summary Carl Zeiss Meditec, Incorporated STRATUSocT™ with RNFL Normative Database

This 510(k) summary for the STRATUS cr™ with RNFL Normative Database is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

#### GENERAL INFORMATION

| Manufacturer:        | Carl Zeiss Meditec, Inc.<br>5160 Hacienda Drive<br>Dublin, California 94568<br>(925) 557-4353 (phone)<br>(925) 557-4481 (fax)<br>Est. Reg. No. 2918630 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:      | R. Michael Crompton<br>Vice President, Regulatory / Clinical Affairs<br>and Quality Assurance                                                          |
| DEVICE DESCRIPTION   |                                                                                                                                                        |
| Classification:      | Class II                                                                                                                                               |
| Trade Name:          | STRATUSOCT™ with Retinal Nerve Fiber Layer Normative Database                                                                                          |
| Generic/Common Name: | Ophthalmoscope (21 CFR § 886.1570)<br>Ultrasonic pulsed echo imaging system (21 CFR § 892.1560)                                                        |

## PREDICATE DEVICE

- Humphrey® OCT 3 (K012727) (1)
## INTENDED USE

The STRATUSocr™ is a high resolution tomographic device for the viewing and axial cross sectional imaging of posterior ocular structures. It is used for in vivo imaging and measurement of the retinal nerve fiber layer, and optic disk. It is intended for use as a diagnostic device to aid in the detection and management of ocular diseases, including but not limited to macular edema, central serous retinopathy, and glaucoma. The STRATUSOCT™ with Retinal Nerve Fiber Layer ("RNFL") Normative Database is a quantitative tool for the comparison of retinal nerve fiber layer in the human retina to a database of known normal subjects. The RNFL Normative Database is intended for use as a diagnostic aid in the detection and management of ocular diseases, especially glaucoma.

{1}------------------------------------------------

## DEVICE DESCRIPTION

The STRATUSOCT™ with RNFL Normative Database is a microprocessor-based low-coherence digital intrument employing low coherence interferometry to generate images of internal ocular tissue microstructures. The device measures optical reflections or backscatter from tissue using optical beam. Results obtained for Retinal Nerve Fiber Layer can be compared to a database of known normal human subjects.

## SUBSTANTIAL EQUIVALENCE

The STRATUSOCT™ with Retinal Nerve Fiber Layer ("RNFL") Normative Database is substantially equivalent to the predicate device identified previously. The STRATUSOCT™ with RNFL Normative Database is substantially equivalent to the predicate device with regard to intended use, operating principle, function, and materials.

Clinical evaluation performed on the STRATUS oct™ with RNFL Normative Database supports the expanded indications for use statement and demonstrates the device is substantially equivalent to the predicate device and does not raise new questions regarding safety and effectiveness with respect to ophthalmoscopes and ultrasonic pulsed echo imaging systems.

## CLINICAL EVALUATION

Clinical data was collected on a statistically significant number of normal human subjects and analyzed to support the inclusion of the RNFL Normative Database on the STRATUSOCT™

## CONCLUSION

As described in this 510(k) Summary, all testing deemed necessary was conducted on the STRATUSOCT™ with RNFL Normative Database to ensure that the device is safe and effective for its intended use when used in accordance with its Instructions for Use.

{2}------------------------------------------------

MAY 2 9 2007

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is positioned to the right of the text, which reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the symbol. The text is in a simple, sans-serif font.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Carl Zeiss Meditec, Inc. c/o Ms. Judith A. Brimacombe 5160 Hacienda Drive Dublin, CA 94568

Re: K030433 and K033123

Trade/Device Name: STRATUSocn™ with Retinal Nerve Fiber Layer (RNFL) & Macula Normative Database Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: II Product Code: OBO Dated: February 7, 2003 and September 26, 2003 Received: February 10, 2003 and September 30, 2003

Dear Ms. Brimacombe:

This letter updates our substantially equivalent letters of May 1, 2003 and April 20, 2004, respectively.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your

{3}------------------------------------------------

#### Page 2 - Ms. Judith A. Brimacombe

device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Everette W. Bees PhD

Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Statement of Indications for Use

510(k) Number (if known): K < 30 4 3 3

Device Name: STRATUSOCT™ with RNFL Normative Database

Indications for Use: The STRATUSOCT™ is a high resolution tomographic device for the viewing and axial cross sectional imaging of posterior ocular structures, It is used for in vivo imaging and measurement of the retinal nerve fiber layer, and optic disk. It is intended for use as a diagnostic device to aid in the detection and management of ocular diseases, including but not limited to macular edema, central serous retinopathy, and glaucoma. The STRATUSOCT™ with Retinal Nerve Fiber Layer ("RNFL") Normative Database is a quantitative tool for the comparison of retinal nerve fiber layer in the human retina to a database of known normal subjects. The RNFL Normative Database is intended for use as a diagnostic aid in the detection and management of ocular diseases, especially glaucoma.

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Jan C Callaway

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Ophthalmic Ear, Nose and Throat Devises

| 510(k) Number                              | K030433                     |
|--------------------------------------------|-----------------------------|
| Prescription Use (Per 21 C.F.R. § 801.109) | <div> <span>✓</span> </div> |
| Over-the-Counter Use                       |                             |

---

**Source:** [https://fda.innolitics.com/device/K030433](https://fda.innolitics.com/device/K030433)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
