K030241 · Blackstone Medical, Inc. · KWP · Feb 21, 2003 · Orthopedic
Device Facts
Record ID
K030241
Device Name
MODIFICATION TO BLACKSTONE SPINAL FIXATION SYSTEM
Applicant
Blackstone Medical, Inc.
Product Code
KWP · Orthopedic
Decision Date
Feb 21, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Blackstone Spinal Fixation System is intended for non-cervical use in the spine. The Blackstone Spinal Fixation System, when used for pedicle screw fixation, is intended only for patients: a) Having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; b) Who are receiving fusion using autogenous bone graft only; c) Who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and d) Who are having the device removed after the development of a solid fusion mass. The Blackstone Spinal Fixation System, when used as a pedicle screw system in skeletally mature patients, is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: a) Degenerative spondylolistheses with objective evidence of neurologic impairment; b) Fracture; c) Dislocation; d) Scoliosis; e) Kyphosis; Spinal tumor; and f) g) Failed previous fusion (pseudarthrosis). The Blackstone Spinal Fixation System, when used for anterolateral non-pedicle screw fixation to the non-cervical spine, is intended for the following indications: a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by history and radiographic studies); b) spondylolistheses; c) spinal stenosis; d) spinal deformities (i.e., scoliosis, kyphosis, lordosis); e) tumor; pseudoarthrosis; f) g) previous failed fusion; and h) trauma (i.e., fracture or dislocation). The Blackstone Spinal Fixation System, when used for posterior non-pedicle screw fixation system of the non-cervical spine, is intended for the following indications: a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by history and radiographic studies); b) spondylolistheses; c) spinal stenosis; d) spinal deformities (i.e., scoliosis, kyphosis, lordosis); e) tumor; pseudoarthrosis; f) g) previous failed fusion; and h) trauma (i.e., fracture or dislocation).
Device Story
Blackstone Spinal Fixation System is a non-sterile, single-use spinal instrumentation system; comprised of titanium alloy (6AL-4V ELI) rods, screws, hooks, and domino rod connectors. Used by surgeons in clinical settings to build spinal implant constructs for immobilization and stabilization of thoracic, lumbar, and sacral spine segments. Domino component allows intra-operative rod attachment for construct extension. System provides mechanical support as an adjunct to fusion; helps restore spinal alignment and stability in patients with deformities or instabilities.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Titanium alloy (6AL-4V ELI, ASTM F136). Non-sterile, single-use components. Includes rods, screws, hooks, and domino rod connectors for spinal fixation.
Indications for Use
Indicated for non-cervical spinal stabilization/immobilization in skeletally mature patients. Conditions: degenerative spondylolisthesis, fracture, dislocation, scoliosis, kyphosis, spinal tumor, pseudarthrosis, degenerative disc disease, spinal stenosis, lordosis, trauma. Pedicle screw use restricted to L3-S1 for severe spondylolisthesis (Grades 3-4) at L5-S1 with autogenous bone graft.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
Blackstone™ Spinal Fixation System 4.5mm Multi-Axial Screws (K020674)
Blackstone™ Spinal Fixation System 4.5mm Mono-Axial Screws (K013558)
Blackstone™ Spinal Fixation System 2nd Gen. Cross-Connector (K003735)
Blackstone™ Spinal Fixation System Modified Multi-Axial Screws (K023498)
Blackstone™ Spinal Fixation System Hooks (K013885)
Blackstone™ Spinal Fixation System Spacers (K022399)
Blackstone™ Spinal Fixation System Staple & Washer (K022605)
Reference Devices
Danek device
Submission Summary (Full Text)
{0}------------------------------------------------
# FEB 2 1 2003
### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
| Name of Firm: | Blackstone Medical, Inc.<br>90 Brookdale Drive<br>Springfield, MA 01104 |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact: | Contact Person: Dean Ciporkin<br>Director of Regulatory Affairs & Quality Assurance |
| Trade Name: | Blackstone™ Spinal Fixation System<br>Spinal Fixation System<br>Domino |
| Common Name: | Rod and screw spinal instrumentation |
| Device Product Code<br>& Classification: | MNH – 888.3070 – Spondylolisthesis Spinal Fixation<br>Device System<br>KWQ – 888.3060 – Spinal Intervertebral Body Fixation<br>Orthosis<br>MNI – 888.3070 – Pedicle Screw Spinal System<br>KWP – 888.3050 – Spinal Interlaminal Fixation Orthosis |
#### Substantially
### Equivalent Devices:
Blackstone™ Spinal Fixation System (K994217) Blackstone™ Spinal Fixation System 4.5mm Multi-Axial Screws (K020674) Blackstone™ Spinal Fixation System 4.5mm Mono-Axial Screws (K013558) Blackstone™ Spinal Fixation System 2nd Gen. Cross-Connector (K003735) Blackstone™ Spinal Fixation System Modified Multi-Axial Screws (K023498) Blackstone™ Spinal Fixation System Hooks (K013885) Blackstone™ Spinal Fixation System Spacers (K022399) Blackstone™ Spinal Fixation System Staple & Washer (K022605)
## Device Description:
The Blackstone™ Spinal Fixation System is comprised of titanium alloy (6AL-4V ELI, per ASTM F136) devices in a variety of non-sterile, single use components. This system allows a surgeon to build a spinal implant construct. The system is attached to the vertebral body by means of screws and hooks to the non-cervical spine.
The domino addition will function as a rod connector. There are clinical applications in which a surgeon will need to have the inter-operative ability to attach rods when extending a construct.
{1}------------------------------------------------
KO30241 age 2 of 3
### Intended Use / Indications for Use:
The Blackstone Spinal Fixation System is intended for non-cervical use in the spine. The Blackstone Spinal Fixation System, when used for pedicle screw fixation, is intended only for patients:
- a) Having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint;
- b) Who are receiving fusion using autogenous bone graft only;
- c) Who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and
- d) Who are having the device removed after the development of a solid fusion mass.
The Blackstone Spinal Fixation System, when used as a pedicle screw system in skeletally mature patients, is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine:
- a) Degenerative spondylolistheses with objective evidence of neurologic impairment;
- b) Fracture;
- c) Dislocation;
- d) Scoliosis;
- e) Kyphosis;
- Spinal tumor; and f)
- g) Failed previous fusion (pseudarthrosis).
The Blackstone Spinal Fixation System, when used for anterolateral non-pedicle screw fixation to the non-cervical spine, is intended for the following indications:
- a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by history and radiographic studies);
- b) spondylolistheses;
- c) spinal stenosis;
- d) spinal deformities (i.e., scoliosis, kyphosis, lordosis);
- e) tumor:
- pseudoarthrosis; f)
- g) previous failed fusion; and
- h) trauma (i.e., fracture or dislocation).
The Blackstone Spinal Fixation System, when used for posterior non-pedicle screw fixation system of the non-cervical spine, is intended for the following indications:
- a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by history and radiographic studies);
- b) spondylolistheses;
- c) spinal stenosis;
- d) spinal deformities (i.e., scoliosis, kyphosis, lordosis);
- e) tumor;
- pseudoarthrosis; f)
- g) previous failed fusion; and
- h) trauma (i.e., fracture or dislocation).
{2}------------------------------------------------
K030241
Page 3 of 3
# BASIS OF SUBSTANTIAL EQUIVALENCE:
The Blackstone™ Domino by its very nature are substantially equivalent to the Danek device which has been cleared by FDA for certain anterior and pedicle fixation use indications.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized image of a human figure with three lines extending from the head, resembling a bird in flight. The text is arranged in a circular pattern around the symbol.
FEB 2 1 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Dean E. Ciporkin Director, Regulatory Affairs and Quality Assurance Blackstone Medical, Inc. 90 Brookdale Drive Springfield, MA 01104
Re: K030241
Trade/Device Name: Blackstone Spinal Fixation System Domino Regulation Number: 21 CFR 888.3050, 888.3060, 888.3070 Regulation Name: Spinal interlaminal fixation orthosi, Spinal intervetebral body fixation Orthosis, Spondylolisthesis spinal fixation device system, Pedicle screw spinal system Regulatory Class: II
Product Code: KWP, KWQ, MNH, MNI Dated: January 22, 2003 Received: January 23, 2003
Dear Mr. Ciporkin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Dean E. Ciporkin
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
for Calia M. Witter, Ph.D., M.D.
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
1030241 Page 1 of 2
# 510(k) Number: K 03024 (
### Device Name: Blackstone™ Spinal Fixation System
#### Indications for Use:
The Blackstone Spinal Fixation System is intended for non-cervical use in the spine. The Blackstone Spinal Fixation System, when used for pedicle screw fixation, is intended only for patients:
- a) Having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint;
- b) Who are receiving fusion using autogenous bone graft only;
- c) Who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and
- d) Who are having the device removed after the development of a solid fusion mass.
The Blackstone Spinal Fixation System, when used as a pedicle screw system in skeletally mature patients, is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine:
- a) Degenerative spondylolistheses with objective evidence of neurologic impairment;
- b) Fracture:
- c) Dislocation:
- d) Scoliosis;
- e) Kyphosis;
- Spinal tumor; and f)
- g) Failed previous fusion (pseudarthrosis).
The Blackstone Spinal Fixation System, when used for anterolateral non-pedicle screw fixation to the non-cervical spine, is intended for the following indications:
- a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by history and radiographic studies);
- b) spondylolistheses;
- c) spinal stenosis;
- d) spinal deformities (i.e., scoliosis, kyphosis, lordosis);
- e) tumor;
- () pseudoarthrosis;
- g) previous failed fusion; and
- h) trauma (i.e., fracture or dislocation).
The Blackstone Spinal Fixation System, when used for posterior non-pedicle screw fixation system of the non-cervical spine, is intended for the following indications:
- a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by history and radiographic studies);
- b) spondylolistheses;
{6}------------------------------------------------
(030241 ige 2 o.
c) spinal stenosis;
d) spinal deformities (i.e., scoliosis, kyphosis, lordosis);
e) tumor;
pseudoarthrosis; f)
g) previous failed fusion; and
h) trauma (i.e., fracture or dislocation).
Concurrence of CDRH, Office of device Evaluation
Prescription Use
OR
Over-The-Counter Use
Miriam C. Provost
(Division Sign-Off) (Division of General, Restorative and Neurological Devices
510(k) Number K630241
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.