K030208 · Orthovita, Inc. · FMF · Apr 16, 2003 · General Hospital
Device Facts
Record ID
K030208
Device Name
IMBIBE II SYRINGE
Applicant
Orthovita, Inc.
Product Code
FMF · General Hospital
Decision Date
Apr 16, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IMBIBE II Syringe is intended for use as a piston syringe for the aspiration of bone marrow, autologous blood, plasma or other blood components. Prior to use, the syringe can be filled with the surgeon's choice of bone void filler (allograft, autograft or synthetic bone graft material). This syringe provides the surgeon with a convenient way to mix blood or marrow with bone void filler (bone graft) and deliver the material to the orthopaedic surgical site.
Device Story
IMBIBE II Syringe is a piston-style delivery device for orthopaedic bone graft materials. Device consists of a calibrated hollow barrel, moveable piston, and Luer-lock nozzles at both distal and proximal ends. Surgeon fills barrel with bone void filler; collects bone marrow or blood via aspiration needle attached to distal Luer-lock using vacuum from secondary syringe at proximal Luer-lock; or fills via proximal port. Device facilitates mixing of graft material with blood/marrow components within the barrel. After mixing, surgeon attaches pushrod to piston, removes distal cap, and extrudes mixture into surgical site. Used in clinical/surgical settings by physicians. Benefits include convenient, integrated mixing and delivery of bone graft composites.
Clinical Evidence
No clinical data. Substantial equivalence is based on design, intended use, and biocompatibility testing of materials (ISO 10993).
Technological Characteristics
Piston syringe; materials include polycarbonate, ABS, silicone, and Dymax UV adhesive 1187-M-T. Features Luer-lock nozzles at distal and proximal ends. Available in 5, 10, 15, and 30 cc volumes. Single-use; sterilized by gamma radiation. No lubricant used.
Indications for Use
Indicated for use in orthopaedic surgical procedures requiring the aspiration of bone marrow, autologous blood, plasma, or other blood components, and the subsequent mixing and delivery of these components with bone void fillers (allograft, autograft, or synthetic bone graft material).
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
K03020f
IMBIBE II Syringe 510(k) Notification Orthovita, Inc.
APR 1 6 2003
### 510(k) Summary
### IMBIBE II Syringe
| Submitted by | Address | Telephone | Contact Person |
|---------------------|-------------------------------------------------|----------------|-----------------------------------------------------|
| Orthovita, Inc. | 45 Great Valley Parkway<br>Malvern, PA 19355 | (610) 640-1775 | Andreina Ide<br>Sr. Director, Regulatory<br>Affairs |
| April 2001 | | | |
| Trade Name | Subject Device | | Predicate Device |
| | IMBIBE II Syringe | | IMBIBE Bone Marrow<br>Aspiration Syringe |
| Common Name | Bone Graft/Bone Void Filler<br>Delivery Syringe | | Bone Graft/Bone Void Filler<br>Delivery Syringe |
| Classification Name | Piston Syringe | | Piston Syringe |
### Device Description:
The IMBIBE II Syringe consists of a calibrated hollow barrel and a moveable piston. At the distal end of the syringe, there is a Luer-lock nozzle for fitting the connector (hub) of a single lumen aspiration needle. A Luer-lock nozzle makes a stable connection between the syringe and needle. A threaded screw cap (containing the Luer-lock nozzle) at the distal end of the syringe can be removed to allow the user to fill the syringe with his or her choice of bone void filler. The proximal end of the syringe contains a moveable piston. A Luerlock nozzle is located in the center of the piston. This connector provides a mechanism for either of two actions.
- Fitting the male connector of a vacuum application device, such as a secondary empty I piston syringe.
- . Fitting the male connector of a secondary pre-filled piston syringe containing bone marrow, autologous blood, plasma or other blood components.
Prior to use, the IMBIBE II syringe is filled with the surgeon's choice of bone void filler (allograft, autograft or synthetic bone graft material). Two methods can be used to imbibe the bone void filler with bone marrow or autologous blood. Method I collects blood or marrow by way of a needle attached to the distal Luer-lock and vacuum/aspiration applied by way of a secondary syringe attached to the proximal Luer-lock. Method II fills the IMBIBE II Syringe through the proximal port (Luer-lock) with blood or marrow collected by way of a secondary syringe.
Once the desired volume of blood or marrow has been collected and mixed with the bone void filler, the secondary syringe and/or aspiration needle are removed. A pushrod is
G 01
-3
{1}------------------------------------------------
KOSOLO 8 IMBIBE II Syringe 510(k) Notification Orthovita, Inc.
attached to the piston at the proximal end of the IMBIBE II Syringe by way of the female Luer-lock, the distal cap is unscrewed and the contents of the syringe are delivered to the surgical site by extruding the mixed bone void filler.
#### Intended Use:
The IMBIBE II Syringe is intended for use as a piston syringe for the aspiration of bone marrow, autologous blood, plasma or other blood components. Prior to use, the syringe is filled with the surgeon's choice of bone void filler (allograft, autograft or synthetic bone graft material). This syringe provides the surgeon with a convenient way to mix blood or marrow with bone void filler (bone graft) and deliver the material to the orthopaedic surgical site.
| COMPARISON TO PREDICATE<br>Table I | | |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | IMBIBE 11 Syringe | Predicates IMBIBE Bone<br>Marrow Aspiration<br>Syringe NML1087 |
| Syringe Type | Piston Syringe | Piston Syringe |
| Intended Use | To collect blood<br>components/bone marrow for<br>mixing with bone graft and<br>subsequent delivery to the<br>surgical site. | To collect blood<br>components/bone marrow for<br>mixing with bone graft and<br>subsequent delivery to the<br>surgical site. |
| Principle of Operation | Syringe used to collect<br>blood, marrow Removable cap allows<br>syringe to be filled with graft<br>material Syringe provides for mixing<br>of blood, marrow with graft<br>material Removable screw cap allows<br>for delivery of blood,<br>marrow filled graft to<br>surgical site Luer-lock (nozzle) at the<br>proximal end of the barrel<br>allows connection of a<br>secondary piston syringe. | Syringe used to collect blood,<br>marrow Removable cap allows<br>syringe to be filled with graft<br>material Syringe provides for mixing<br>of blood, marrow with graft<br>material Removable screw cap allows<br>for delivery of blood, marrow<br>filled graft to surgical site |
### Comparison to Predicate:
G
02
{2}------------------------------------------------
K030208f
IMBIBE II Syringe
510(k) Notification
Orthovita, Inc.
| Overall Length | 3.5, 4.6, 3.9, 5.3 inches<br>(respectfully for a 5cc, 10cc, 15cc, 30cc internal volume) | 2.9 inches |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Barrel Diameter | 0.73, 0.73, 1.05, 1.05 inches, OD<br>(respectfully for a 5cc, 10cc, 15cc, 30cc internal volume) | 0.73 inches, OD |
| Tip Type | Gasket | Gasket |
| Volume | 5, 10, 15, 30 cc | 10 cc |
| Nozzle Type | Luer-lock | Luer-lock |
| Barrel Markings | Graduated scale | Graduated scale |
| Lubricant Composition | None | Silicone |
| Lubricant amt/cm² | Not Applicable | 100mg ± 5mg |
| Barrel Transparency | Transparent, no radiopacifiers | Transparent, no radiopacifiers |
| Reuse | Single use only | Single use only |
| Biocompatibility | Established by way of legally marketed device, IMBIBE Bone<br>Marrow Aspiration Syringe, K011087 and ISO 10993 testing of adhesive | Established |
| Materials | Polycarbonate, ABS, Silicone,<br>Dymax UV adhesive 1187-M-T | Polycarbonate, ABS, Silicone |
| Labeling | See Exhibit B. Same as predicate except additional information regarding directions for use. | See Exhibit C |
| Sterility | Sterilized by gamma radiation | Sterilized by gamma radiation |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 6 2003
Ms. Andreina Ide Senior Director, Regulatory Affairs Orthovita, Inc. 45 Great Valley Parkway Malvern, Pennsylvania 19355
Re: K030208
Trade/Device Name: IMBIBE II Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: II Product Codes: FMF Dated: January 17, 2003 Received: January 21, 2003
Dear Ms. Ide:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Andreina Ide
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmaldsmamain.html
Sincerely yours,
Miriam C. Provost
tot Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
# 510(k) Number:
### Device Name: IMBIBE II Syringe
### Indications For Use:
The IMBIBE II Syringe is intended for use as a piston syringe for the aspiration of bone marrow, autologous blood, plasma or other blood components. Prior to use, the syringe can be filled with the surgeon's choice of bone void filler (allograft, autograft or synthetic bone graft material). This syringe provides the surgeon with a convenient way to mix blood or marrow with bone void filler (bone graft) and deliver the material to the orthopaedic surgical site.
#### PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Prescription Use**
*(Per 21 CFR 801.109)*
OR
Over-The-Counter Use
Miriaml. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K030208
01
D
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.