← Product Code [LLZ](/productcode/LLZ) · K024159

# STRYKERWARE SYSTEM (K024159)

_Stryker Endoscopy · LLZ · Mar 10, 2003 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K024159

## Device Facts

- **Applicant:** Stryker Endoscopy
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Mar 10, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The Strykerware™ PACS System is a system designed to provide licensed medical professionals access to medical images. This product is intended to be used for reviewing and storing images from a variety of modalities. Strykerware™ is the system with software installed on a computer configured with connections to image acquisition devices, image storage locations and the internet. The system facilitates the transferal of images between local and remote locations (including but not limited to a licensed medical professional's operating suite). The users of Strykerware™ are general surgeons, gynecologists, cardiac surgeons, plastic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons, and urologists.

## Device Story

Strykerware™ PACS System; software-based medical image management system. Inputs: medical images and data from acquisition devices or other PACS networks. Processing: integrated web-based viewing and storage archival applications; facilitates image transfer between locations (e.g., office to operating suite). Outputs: accessible medical images for review. Used in office-based practices and operating suites by physicians and licensed professionals. Benefits: enables transition from paper-based to paperless environment; improves workflow and information flow between office and surgical settings.

## Clinical Evidence

Bench testing only. The device conforms to voluntary standards including ACR/NEMA DICOM, ANSI/AAMI SW68, and IEC 60601.

## Technological Characteristics

Software-based PACS; web-based viewing and storage archival applications. Connectivity: networked, internet-enabled, DICOM compliant. Standards: ACR/NEMA DICOM, ANSI/AAMI SW68 (software life cycle), IEC 60601 (medical electrical equipment).

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- RemoteImage™ System ([K994228](/device/K994228.md))
- AMICAS Web/Intranet Image Server ([K970064](/device/K970064.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K02 4159

MAR 1 0 2003

**stryker**
ENDOSCOPY

5900 Ontical Cour an Jose CA 9513 x (408) 754-2505

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

## Device Name

| Classification Name:   | System, Image Processing                  |
|------------------------|-------------------------------------------|
| Common and Usual Name: | StrykerwareTM Office Portal/Media Archive |

Strykerware™ System Proprietary Name:

This 510(k) summary and effectiveness is being submitted in accordance with requirements of SMDA 1990.

The Strykerware™ PACS System is substantially equivalent in safety and efficacy as the currently marketed Remotelmage™ System and AMICAS Web/Intranet Image Server which were cleared under 510K#'s K994228 and K970064 respectively.

The Stryker Strykerware™ PACS System is a medical system that is designed to allow access to medical images and related data for physicians and other licensed professionals. The system receives images and medical data from image acquisition devices or other PACS (Picture Archive and Communication System) networks. The system's software includes an integrated web-based viewing and storage archival applications. The Strykerware™ System allows for the transfer of medical images and data between locations such as but not limited to a physicians operating suite.

The intent of the Strykerware™ System is to allow an office-based practice to convert from a paper to paperless environment and to create a flow of medical images and information between the office and the operating suites of the physicians belonging to that practice.

The Strykerware™ System conforms to the following voluntary standards: ACR/NEMA DICOM, ANSI/AAMI SW68 - "Medical device software - Software life cycle processes", and IEC 60601 -"Medical Electrical Equipment."

The technological differences between the Strykerware™ System and the predicate Remotelmage™ and AMICAS devices do not raise new issues of safety and efficacy of the predicate devices. Therefore, the Strykerware™ System is substantially equivalent to the currently marketed Remotelmage™ and AMICAS Web/Intranet Image Server.

Lee Rich

Jerry Dickerson Senior Project Engineer Stryker Endoscopy

Date: 12-13-02

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Evan Norton Product Manager Stryker Endoscopy 5900 Optical Court SAN JOSE CA 95138 Re: K024159

MAR 1 0 2003

Trade/Device Name: Strykerware PACS System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: December 13, 2002 Received: December 17, 2002

Dear Mr. Norton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive. Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

December 12, 2002

KO2 4154 510(k) Number (if known):

Device Name

Indications for Use:

The Strykerware™ PACS System is a system designed to provide licensed medical professionals access to medical images. This product is intended to be used for reviewing and storing images from a variety of modalities. Strykerware™ is the system with software installed on a computer configured with connections to image acquisition devices, image storage locations and the internet. The system facilitates the transferal of images between local and remote locations (including but not limited to a licensed medical professional's operating suite). The users of Strykerware™ are general surgeons, gynecologists, cardiac surgeons, plastic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons, and urologists.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

510(k) Number

OR

Over-The-Counter-Use_

Marcye brogdon

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Dai

Stryker Endoscopy Strykerware™ PACS System SHUIKTSUOMISSION

2 of 15

---

**Source:** [https://fda.innolitics.com/device/K024159](https://fda.innolitics.com/device/K024159)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
