← Product Code [KNW](/productcode/KNW) · K024089

# VACUFLASH BIOPSY SYSTEM (K024089)

_Bip USA, Inc. · KNW · Mar 11, 2003 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K024089

## Device Facts

- **Applicant:** Bip USA, Inc.
- **Product Code:** [KNW](/productcode/KNW.md)
- **Decision Date:** Mar 11, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1075
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The BIP VacuFlash® Biopsy System is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. The instrument is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

## Device Story

Mechanical vacuum-assisted biopsy system; used with imaging guidance (ultrasound, X-ray, stereotaxy, CT, MR). Components: disposable trocar-tipped needle with vacuum cylinder, disposable introducer, reusable housing with control module, touch pad, LEDs, and vacuum/air pressure generators. Operation: tissue vacuum-aspirated into inner cannula; cut via sampling window edges and rotating outer cannula. Used in clinical settings by physicians. Provides semi-automatic and automatic modes for cutter advancement, rotation, and specimen retrieval. Enables radial tissue collection for retrospective correlation with imaging. Benefits: portable, integrated design minimizes user interaction variability, facilitates diagnostic tissue removal.

## Clinical Evidence

Bench testing only. Preclinical performance testing confirmed the device performs as intended for breast tissue biopsy and provides sample quality equivalent to predicate devices.

## Technological Characteristics

Mechanical biopsy system; vacuum-assisted core biopsy. Components: disposable needle/vacuum cylinder, reusable housing with integrated motor, control module, battery, and vacuum/air pressure generators. Features touch pad interface with LED status indicators. Portable form factor (approx. 14 oz). Compatible with various stereotactic/imaging mounting brackets. Sterilization: disposable components provided sterile.

## Regulatory Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Predicate Devices

- Ethicon Mammotome® Hand Held 8 Gauge Probe ([K003297](/device/K003297.md))
- Ethicon Mammotome® Hand Held System ([K991980](/device/K991980.md))
- Ethicon Mammotome® ([K970565](/device/K970565.md))
- Suros Surgical Systems ATEC® Vacuum Assisted Core Biopsy System ([K010400](/device/K010400.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Appendix VIII: 510(k) Summary of Safety and Effectiveness

# VacuFlash® Biopsy System

#### Company:

BIP USA Inc. 345 Third Street, Suite 400 Niagara Falls, NY 14303

MAR 11 2003

# Contact

Siegfried Gruchot, XYZ General Manager (716) 284-1581, telephone (716) 284-1583, fax

# Date Prepared

03 December 2002

# Name of Device

Trade Name: VacuFlash® Biopsy System Classification Name: Biopsy Needle

# Predicate Devices

Ethicon Mammotome® Hand Held 8 Gauge Probe Ethicon Mammotome® Hand Held System Ethicon Mammotome® Suros ATEC® Vacuum Assisted Core Biopsy System

# Device Description

The BIP VacuFlash® Biopsy System is a mechanical device that may be used with imaging quidance (such as ultrasound, X-Ray, stereotaxy, CT, or MR) to provide breast tissue samples for histologic examination with partial or complete removal of the imaged abnormality.

The VacuFlash® Biopsy System consists of three major components: a disposable trocar tipped needle with a vacuum cylinder, a disposable introducer with stylette, and a reusable device housing containing the control module, touch pad with control buttons and LEDs indicating device status, and vacuum/air pressure generators. The disposable needle/vacuum cylinder will be available in several different lengths. A battery recharger/table top stand is the only provided accessory. Mounting brackets to attach the VacuFlash® Biopsy System to the Fischer Mammotest® and to the Lorad stereotactic tables are optional accessories. In addition, brackets to attaché the VacuFlash to the Siemens stereotactic mammography and to the Siemens MR Mammography Biopsy systems are available. Sterile bushings for the above devices, including CytoGuide stereotactic and GE DMR Mammography systems are also available.

The needle inner cannula incorporates a distal sampling with sharpened edges. During the biopsy procedure, tissue is vacuum aspirated into the inner cannula and cut using both the sampling window edges and the rotating edge of the outer cannula. Once the needle is removed from the patient's body, the sampling window is opened and the specimen removed. Tissue samples may be collected in radial order so that when imaging guidance is used, sampling positions may be retrospectively correlated with the imaging display.

{1}------------------------------------------------

K024089 2/2

#### Intended Use

The BIP VacuFlash® Biopsy System is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. The instrument is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.

The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

#### Substantial Equivalence

The BIP VacuFlash® Biopsy System is substantially equivalent to the Mammotome® that was determined to be substantially equivalent on 28 March 1997 (K970565). The BIP VacuFlash® Biopsy System is also substantially equivalent to the Mammotome® Hand Held System that was determined to be substantially equivalent on 17 August 1999 (K991980) and the Mammotome® Hand Held 8 Gauge Probe that was determined to be substantially equivalent on 18 January 2001 (K003297). In addition, the VacuFlash® Biopsy System is also substantially equivalent to the Suros Surgical Systems ATEC® Vacuum Assisted Core Biopsy System that was determined to be substantially equivalent on 24 May 2001 (K010400).

The BIP VacuFlash® Biopsy System has similar indications for use and technological characteristics as the predicate devices. The patient contact components and component materials for obtaining core biopsy samples in both the new and predicate devices are equivalent. The packaging materials, packaging configuration, sterilization methods and sterility assurance level are also equivalent.

The hand-held or mounted biopsy device, used with or without imaging modalities, provides for the diagnostic removal of tissue with fluid management through a combination of vacuum and radial cutting functions. The proposed and predicate devices contain the same primary components to achieve these functions: a needle/probe, housing/handle/holster, and a control module.

In the proposed device, the housing contains the motor and control module as well as an integrated battery and vacuum/air pressure generation equipment. A touch pad is also integrated into the device with buttons for operational control and LED indicator lights provide feedback. The needle/probe and the device weigh approximately 14 oz. Integration of all components into the handle/housing makes the device portable and facilitates user and patient interface. The control module allows the VacuFlash® Biopsy System to provide semi-automatic and automatic modes of operation for cutter advancement, cutter rotation, and specimen retrieval. In addition, the variability of user interaction is minimized.

Preclinical testing confirms the quality of samples obtained with the BIP VacuFlash® Biopsy System is equivalent to those obtained with the predicate devices.

Based on the indications for use, technological characteristics and testing results, the BIP VacuFlash® Biopsy System does not raise significant new questions of safety and effectiveness.

#### Performance Data

Preclinical testing was performed to confirm the device performs as intended. Testing demonstrated satisfactory performance in breast tissue biopsy.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing the bird's head and body. The bird is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the border of the circle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 11 2003

Mr. Siegfried Gruchot General Manager BIP USA, Inc. 345 Third Street, Suite 400 Niagara Falls, New York 14303

Re: K024089

Trade/Device Name: VacuFlash Biopsy System Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: II Product Code: KNW Dated: December 3, 2002 Received: December 11, 2002

Dear Mr. Gruchot:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Siegfried Gruchot

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Muriam C. Provost

For Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Statement of Indications for Use

510(k) Number (if known): KO24089

Device Name:

VacuFlash Biopsy System

Indications for Use:

The BIP VacuFlash® Biopsy System is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. The instrument is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.

The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ***_*** OR Over-The-Counter Use **_**
Per 21 CFR 801.109

Muriam C. Provost

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K024089

510(k) Notification BIP USA, Inc.

03 Dec 2002

CONFIDENTIAL page 4

---

**Source:** [https://fda.innolitics.com/device/K024089](https://fda.innolitics.com/device/K024089)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
