ABBOTT PLUM A+ INFUSION PUMP, MODEL 12391

K024084 · Abbott Laboratories · FRN · Dec 31, 2002 · General Hospital

Device Facts

Record IDK024084
Device NameABBOTT PLUM A+ INFUSION PUMP, MODEL 12391
ApplicantAbbott Laboratories
Product CodeFRN · General Hospital
Decision DateDec 31, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The pump is intended for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products. The pump must be used with sterile, dedicated, intravenous PlumSet® administration sets.

Device Story

Abbott Plum A+ Infusion Pump is an electromechanical device using a stepper motor and in-line cassette to meter IV fluids via dedicated PlumSet administration sets. Used in clinical settings for parenteral, enteral, epidural, and blood product delivery. Healthcare practitioners program fluid delivery using weight-based or medication-based units (e.g., mcg/kg/hr). Device features include nurse call, computer data port, drug list, and bar code wand. Display provides status of active operations, alarms, and flow parameters for single or concurrent lines. Software version 11.3 modifications include user-selectable units of measure, updated drug list, therapy confirmation screen, standby function, and password-protected keypad to aid in medication error reduction. Device supports standard, piggyback, and concurrent delivery.

Clinical Evidence

Bench testing only. Functional performance testing was conducted to verify and validate design modifications, confirming acceptable performance for the device with version 11.3 software.

Technological Characteristics

Electromechanical infusion pump; stepper motor actuation; in-line cassette mechanism; dedicated PlumSet administration sets; software-controlled; includes nurse call, data port, and bar code wand; password-protected keypad; version 11.3 software.

Indications for Use

Indicated for use in parenteral, enteral, and epidural therapies, and the administration of whole blood and blood products.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Koz 4084 # DEC 31 2002 Special 510(k): Device Modification Abbott Plum A+® Infusion Pump Page 1 of 3 # 510(k) SUMMARY Abbott Plum A+® Infusion Pump #### 1. Submitted by: Frank Pokrop Associate Director, Requlatory Affairs Abbott Laboratories D-389. Bldg. LFJ45 Abbott Park, IL 60064 #### 2. Date Prepared: December 10, 2002 #### 3. Name/Classification of Device: Infusion Pump. Class II 80 FRN -- 21 CFR Parts 880.5725 #### 4. Proposed Device: Abbott Plum A+® Infusion Pump, List Number, 12391 #### 5. Predicate Device: Abbott Plum A+® Infusion Pump #### 6. Proposed Device Description: The Abbott Plum A+ Infusion Pump is an electromechanical infusion pump that uses a stepper motor in conjunction with an in-line cassette to meter IV fluids through a dedicated intravenous administration set that is also manufactured and distributed by Abbott Laboratories. The subject and predicate devices have identical indications for use. Abbott proposes to modify the predicate device through software changes, minor changes to mechanical parts and labeling changes. The featured software changes include: providing users with a choice to select the manufacturer's recommended units of measure for drug administration, updating the drug list, adding a confirmation screen before allowing therapy to begin, including a standby function and adding a password protected keypad. {1}------------------------------------------------ Special 510(k): Device Modification Abbott Plum A+® Infusion Pump Page 2 of 3 ## 6. Proposed Device Description: (cont'd) These proposed changes along with the updated drug list are intended to potentially aid in medication error reduction by highlighting the drug to be infused and the units of measure for drug administration as shown in the manufacturer's package insert. The user interface of the infusion pump allows the healthcare practitioner to program fluid delivery through a variety of weight and medication based units such as micrograms/kq/hour, grams/hr and other delivery specifications. Like the predicate device. the proposed device includes: a nurse call function, a computer data port, a druq list, and a bar code wand. The display on the pump provides visible indication of several functions including active pump operations, alarm and program status and the parameters of fluid flow for one or both incoming fluid lines. Both the predicate and the proposed devices can be used for standard, piggyback, or concurrent fluid delivery. #### 7. Statement of Intended Use: The pump is intended for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products. This is the same intended use as the predicate device. The pump must be used with sterile, dedicated, intravenous PlumSet® administration sets. ## 8. Summary of Technological Characteristics of New Device Compared to Predicate Device Both pumps use the same software lanquage. The proposed pump has similar design, materials of construction, components, labeling and manufacturing processes as the currently marketed Abbott Plum A+® Infusion Pump. These differences do not raise new issues of safety and effectiveness nor do they alter the fundamental technology of the predicate device. {2}------------------------------------------------ Special 510(k): Device Modification Page 3 of 3 # 9. Discussion and Conclusions from Nonclinical Tests: Data regarding the functional performance of the proposed Abbott Plum A+® Infusion Pump has been generated and reviewed. The results of testing conducted to validate and verify the design modifications demonstrate acceptable performance of the device. There are no new issues of safety or effectiveness raised by the Plum A+® with version 11.3 software. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of flowing lines, possibly representing health and well-being. ## DEPARTMENT OF HEALTH & HUMAN SERVICES od and Drug Administrat 9200 Corporate Boulevaro Rockville MD 20850 # DEC 3 1 2002 Mr. Frank Pokrop Associate Director, R/A Abbott Laboratories 200 Abbott Park Road Dept. 0389, Bldg. LFJ45 Abbott Park, Illinois 60064-6133 Re: K024084 Trade/Device Name: Abbott Plum A+® Infusion Pump Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: December 10, 2002 Received: December 11, 2002 Dear Mr. Pokrop: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ## Page 2 -- Mr. Pokrop Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Jo. Albert Louise for, Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ # Indications for Use Statement | 510(k) Number (if known) | 12024084 | |--------------------------|---------------------------------------------------------------------------------------------------------------------------| | Device Name: | Abbott Plum A+ <sup> ® </sup> Infusion Pump | | Indications For Use: | Abbott Plum A+ <sup>®</sup> Infusion Pump has the following indications for use: | | | Indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products. | # PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED ### Concurrence of CDRH, Office of Device Evaluation (ODE) OR Prescription Use (per 21 CFR 801.109) Over-The_Counter Use__________________________________________________________________________________________________________________________________________________________ Kittie Cucenite (Division Sign-Off) (Division Sign-On) Division of Anesthesiology, General Hospital, Division Control, Dental Devices 510(k) Number. K024084
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...