SUNDER CENTRAL VENOUS CATHETER KIT, MODELS SD-3L55F30J AND SD-3L70F30J
Applicant
Sunder Biomedical Tech. Co., Ltd.
Product Code
FOZ · General Hospital
Decision Date
Jul 14, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The triple lumen central venous Catheter is intended for vascular access infusion and withdrawal of blood, blood products, and fluids, central venous blood pressure monitoring (CVP) plasma pheresis, acute hyperalimentation, continuous or int4ermittent drug infusion.
Device Story
Triple-lumen polyurethane central venous catheter (5.5 and 7.0 French); features three independent non-communicating lumens, extension lines, luer hubs, and slide clamps. Placed in a vein with distal end in vena cava. Used for vascular access, infusion/withdrawal of blood/fluids, CVP monitoring, and drug delivery. Operated by clinicians in clinical settings. Kit includes catheter in PET tray, sealed with Tyvek, and sterilized. Provides multi-lumen access for patients requiring complex intravenous therapy or monitoring.
Clinical Evidence
Bench testing only. Performance testing included tensile strength, elongation, anti-corrosive properties, flexural fatigue, stiffness, flow rate, burst pressure, collapse, leakage, and ink adhesion. Microbiological testing included sterilization process validation and E.O. residual testing. Biocompatibility testing was performed. No clinical data provided.
Technological Characteristics
Triple-lumen polyurethane catheter; 5.5 and 7.0 French sizes. Features three independent non-communicating lumens, extension lines, luer hubs, and slide clamps. Sterilized via E.O. gas. Packaged in PET tray with Tyvek lid.
Indications for Use
Indicated for patients requiring multiple IV access sites, lacking usable peripheral sites, requiring central venous pressure monitoring, total parenteral nutrition, incompatible medications, multiple fluid/medication/chemotherapy infusions, frequent blood sampling, blood transfusions, or hypertonic/hyperosmolar/divergent pH infusions.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
ARROW g* and blue Plus™ Multi-Lumen Central Venous Catheter (K993691)
Submission Summary (Full Text)
{1}------------------------------------------------
K024007
## Section 1. -510 (k) Summary
- a) Submitter:
SUDER
Sunder Biomedical Tech. Co., Ltd
Contact person: Tony Hung
Address: 10F-1. 1-67, Wu-Chuan Rd., Taichung, Taiwan 403 R.O.C.
886-4-23755650 Phone:
Fax: 886-4-23755651
Email: tonv@sunder.com.tw
Date Summary prepared: 11/15/2002
- b) Device
| Device trade name: | Sunder Central Venous Catheter Kit |
|-----------------------------|----------------------------------------------------------------------------------|
| Device common name: | Central Venous Catheter (CVC) |
| Device Classification name: | Short-term catheters (less than 30 days) -Percutaneous<br>Intravascular catheter |
c) Legally marketed device to which the device is substantially equivalent
ARROW g* and blue Plus™ Multi-Lumen Central Venous Catheter 510(k) No.: K993691
- d) Description of the device
The medical device made by Sunder Biomedical Tech, CO., LTD are triple-lumen. polyurethane catheter, 5.5 and 7.0 French in size, with three independent non-communication lumens, extension lines, luer hubs and slide clamps, placed within a vein and whose distal end is intended to be located within the vena cava (inferior of superior). Central Venous Catheter (CVC) - a tubular device with single lumen or rnulti-lumen catheter, or cannula, placed within a vein and whose distal end is intended to be located within the vena cava (inferior of superior).
The catheterization kit components are configured in a Polyester Terephthalate (PET) tray and sealed with a Tyvek lid stock, and sterilized.
- e) Intended use of the device
The triple lumen central venous Catheter is intended for vascular access infusion and withdrawal of blood, blood products, and fluids, central venous blood pressure monitoring (CVP) plasma pheresis, acute hyperalimentation, continuous or int4ermittent drug infusion.
- f) Performance tests:
The following performance tests are included in the submission:
- 1. Functional testing
- ﺳﺮ ﺍ Tensile strength and elongation test
- Anti-corrosive -
- , ( Catheter-flexural fatigue tolerance test
{2}------------------------------------------------
K0241007
- Stiffness test ﺮ
- ア Flow rate
- ア Catheter burst pressure test. Catheter collapse test
- تر Leakage test
- تر Ink adhesion test
- 2. Validation (Microbiological test)
- ア Sterilization Process Validation
- ア E.O. residual test
- 3. Biocompatibility (Biological tests)
The result of the laboratory tests demonstrate that the device is as safe as and is effective with the legally marketed predicate device.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and features an abstract image of an eagle. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged around the eagle in a circular fashion. The logo is black and white.
Public Health Service
JUL 1 4 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Tony Hung President Sunder Biomedical Tech. Co. Ltd. 10F-1, 1-67, Wu-Chung Road Taichung City 403 TAIWAN R.O.C.
Re: K024007
Trade/Device Name: Sunder Central Venous Catheter Kits (SD-3L70F30J and SD-3L55F30J) Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: May 26, 2003 Received: May 28, 2003
Dear Mr. Hung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Hung
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Susan Runner
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Attachment 4.
510(k) Number (if known): 서스2400 7
Device Name: SUNDER CENTRAL VENOUS CATHETER KITS (SD-3L70F30J AND SD-3L55F30J)
Indications for Use:
The triple lumen central venous Catheter is intended for vascular access infusion and withdrawal of blood, blood products, and fluids, central venous blood pressure monitoring and continuous or intermittent drug infusion.
- Patients requiring multiple sites for IV access.
- Patients lacking useable peripheral IV sites.
- Patients requiring central venous pressure monitoring.
- Patients requiring total parenteral nutrition.
- Patients receiving incompatible medications.
- Patients requiring multiple infusions of fluids, medications, or chemotherapy.
- . Patients subject to frequent blood sampling or receiving blood transfusions.
- Patients receiving infusions that are hypertonic, hyperosmolar or infusions that have divergent pH values.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 3-10-98)
1
Pollara Cuceniti
510(k) Number: K024007
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.