K023934 · Avacore Technologies, Inc. · DWJ · Jan 23, 2003 · Cardiovascular
Device Facts
Record ID
K023934
Device Name
RAPID THERMAL EXCHANGE SYSTEM
Applicant
Avacore Technologies, Inc.
Product Code
DWJ · Cardiovascular
Decision Date
Jan 23, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5900
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Intended Use
The Rapid Thermal Exchange System is designed to noninvasively lower or raise a patient's temperature and/or maintain a desired patient temperature. This is accomplished with local application of negative pressure and heating/cooling to a distal appendage.
Device Story
Rapid Thermal Exchange (RTX) System provides noninvasive core body temperature regulation. System components: Field Controller Unit (FCU), attachment hoses, and Thermal Exchange Chamber. Operation: Patient inserts hand or foot into chamber; light vacuum (25 inches water/47mmHg) applied to increase blood flow in appendage; temperature-controlled water circulated through chamber surface. FCU contains vacuum pump and water heater/cooler. Used in clinical settings; operated by healthcare professionals. Output: Controlled thermal exchange with blood flowing through appendage, resulting in core temperature change. Benefits: Rapid, noninvasive temperature management for medical conditions requiring thermal regulation.
Clinical Evidence
Bench testing confirms RTX System maintains desired operating characteristics and complies with IEC 60601-1 safety standards. Clinical evidence relies on human studies reported in the predicate submission, which demonstrated that simultaneous application of light vacuum and thermal exchange effectively and noninvasively changes patient core body temperature.
Technological Characteristics
System comprises a Field Controller Unit (vacuum pump, water heater/cooler), attachment hoses, and a Thermal Exchange Chamber. Operates via local negative pressure (25 inches water) and conductive thermal exchange. Available in small, medium, and large sizes for hand/wrist or foot. Complies with IEC 60601-1. No software-based automation described.
Indications for Use
Indicated for patients requiring noninvasive core temperature regulation (lowering, raising, or maintenance) via local application of negative pressure and thermal exchange to a distal appendage (hand or foot). Requires physician prescription.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
K014210 — PALMO THERMOREGULATION INTERFACE ACCESSORY · Avacore Technologies, Inc. · Aug 16, 2002
K101589 — BLANKETROL III, MODEL 233 AND COOLBLUE, MODEL 25-01 HYPER-HYPOTHERMIA SYSTEMS · Cincinnati Sub-Zero Products, Inc. · Sep 20, 2010
K033021 — KIMBERLY-CLARK PATIENT WARMING SYSTEM - MODEL 100 CONTROL UNIT AND ENERGY TRANSFER PADS · Kimberly-Clark Corp. · Nov 12, 2003
K070112 — RAPIDCOOL PATIENT TEMPERATURE MANAGEMENT SYSTEM · Medcool, Inc. · Jul 13, 2007
K963261 — EBI THERMAL THERAPY SYSTEM · Electro-Biology, Inc. · Sep 27, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
## SECTION 2 - 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 8807.92
Contact Person JAN 2 3 2003 Craig Coombs Vice President, Regulatory/Clinical Affairs & Quality Assurance 415.531.1932 650.321.3908 fax cicoombs@aol.com
· Device Name Trade Name: Rapid Thermal Exchange System Common Name: System, Thermal Regulating Classification Name: Thermal Regulation System Accessory (21 CFR 870.5900)
Predicate Devices Palmo Thermoregulating Accessory
K014210
Device Description
The Rapid Thermal Exchange (RTX) System consists of three main parts. The 1) Field Controller unit is the vacuum source; it also regulates and pumps the temperaturecontrolled water through the 2) Attachment Hoses that are connected to 3) the Thermal Exchange Chamber (identified in earlier submissions as the Palmo Thermoregulating Accessory).
The patient inserts his/her hand through the arm seal of the Thermal Exchange Chamber. The palm is placed on the temperature-controlled thermal exchange surface. A light vacuum is created by connecting the Thermal Exchange Chamber to the Field Controller Unit's vacuum pump via the inline vacuum control valve. The physician can monitor the adequacy of the vacuum and seal from the dial on the mechanical vacuum gauge mounted to the top of the Thermal Exchange Chamber. A light vacuum (25 inches of water, 47mmHg) is applied to increase the amount of blood available in the appendage.
This combination of light vacuum and a thermal exchange surface provides a rapid and noninvasive mechanism for changing the temperature of the blood flowing through the appendage. This, in turn, changes the temperature of the body core.
The arm seal of the Thermal Exchange Chamber is available in three different sizes (small, medium & large) to accommodate the range of possible hand and wrist sizes of athletes. A variation of the Chamber can be applied to the foot in a similar manner.
{1}------------------------------------------------
## Indications for Use
The Rapid Thermal Exchange System is designed to noninvasively lower or raise a patient's temperature and/or maintain a desired patient temperature. This is accomplished with local application of negative pressure and heating/cooling to a distal appendage.
### Testing in Support of Substantial Equivalence Determination
The thermal exchange chamber (aka the Palmo) is unchanged since the predicate submission. The operating characteristics are also unchanged.
The results of bench testing support the claims that the RTX System can maintain the desired operating characteristics. In addition, the manufacturer claims that the RTX system has been designed, tested and manufactured in compliance with the FDA consensus standard IEC 60601-1 (Medical Electrical Equipment -- Part 1: General requirements for safety).
Additionally, human studies reported in a predicate submission have demonstrated that the simultaneous application of light vacuum and thermal exchange can effectively and noninvasively change the body core temperature of a patient.
#### Substantial Equivalence Conclusion
Substantial equivalence is based on the fact that the RTX System has the same intended use as the Palmo Thermoregulating Accessory (K014210).
The RTX System uses the same technique and operating parameters as the Palmo Thermoregulating Accessory (Palmo). The Palmo consisted of a vacuum chamber for the distal appendage that served as the thermal exchange surface, an arm seal, an in-line vacuum relief valve and Attachment hoses. The RTX System uses the exact same components.
The Palmo Thermoregulating Accessory was cleared as an accessory to the Cincinnati SubZero Blanketrol II. The Blanketrol II pumped its temperature-controlled water through the Palmo thermal exchange chamber. The RTX System has a water heater/cooler in its Field Controller Unit (FCU) that pumps temperature-controlled water through the same Palmo thermal exchange chamber. The Palmo Thermoregulating Accessory was cleared to use a hospital's central vacuum as a vacuum source for the Palmo thermal exchange chamber. The FCU has its own vacuum pumps as a vacuum source for the same Palmo thermal exchange chamber. Since the RTX provides both temperature-controlled water and vacuum sources in a suitcase-sized, wheeled, Field Control Unit, it can be applied in a wider range of locations.
The RTX System requires the prescription or order of a physician to be applied in response to medical conditions.
{2}------------------------------------------------
The Palmo and the RTX System use identical technology, methods and temperature/vacuum settings to exchange thermal energy with the body core of the patient. The technology uses a local application of negative pressure and a thermal exchange surface to a distal appendage. There are no new questions of safety or efficacy raised between the two systems.
AVAcore Technologies certifies that the RTX System is in compliance with the following FDA recognized consensus standard: IEC 60601-1: Medical Electrical Equipment – Part 1: General requirements for safety.
The predicate device is classified as a "Thermal Regulating System" (per 21CFR870.5900), as is the RTX. Therefore, it can be concluded that the RTX System is substantially equivalent to the predicate device.
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
JAN 2 3 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AVACore Technologies, Inc. c/o Mr. Michael Kwan Underwriters Laboratories Inc. 1655 Scott Boulevard Santa Clara, CA 95050
Re: K023934
Trade Name: Rapid Thermal Exchange (RTX) System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulation System Regulatory Class: Class II (two) Product Code: DWJ Dated: January 13, 2002 Received: January 14, 2003
Dear Mr. Kwan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 – Mr. Michael Kwan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman. M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name: Rapid Thermal Exchange (RTX) System
Indications for Use
The Rapid Thermal Exchange System is designed to noninvasively lower or raise a patient's temperature and/or maintain a desired patient temperature. This is accomplished with local application of negative pressure and heating/cooling to a distal appendage.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | |
|------------------------------------------|--|
|------------------------------------------|--|
OR
| Over-the-Counter Use<br>(Optional Format 1-2-96) | |
|--------------------------------------------------|--|
|--------------------------------------------------|--|
(Division Sign-Off)
Division of Cardiovascula cos
| 510(k) Number | K023954 |
|---------------|---------|
|---------------|---------|
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.