The Automatic X-Ray Film Processor is intended to be used to process films exposed for medical purposes. The automatic and continuous process contains developing, fixing, washing and dryng of films.
Device Story
LifeRay PRO 1000 is an automatic radiographic film processor; processes exposed medical films via continuous mechanical transport. Input: exposed radiographic film. Operation: film fed into processor; roller transport system moves film through developing, fixing, rinsing, and drying chemical baths. Microprocessor controls processing conditions, temperature regulation, circulation, and automatic replenishment of solutions. Output: processed, dried film ready for clinical viewing. Used in clinical settings by trained medical professionals (physicians, nurses, technicians). Benefits: automates film development workflow, ensuring consistent processing conditions for diagnostic imaging.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Automatic radiographic film processor; mechanical roller transport system; chemical immersion (developer/fixer) with water rinse and drying; microprocessor-controlled temperature regulation, circulation, and replenishment; anti-crystallization and stand-by modes; integrated software control; 35cm feed width; 10x10cm minimum film format.
Indications for Use
Indicated for processing medical radiographic films in general diagnostic imaging procedures. Intended for use by trained medical professionals, including physicians, nurses, and lab technicians.
Regulatory Classification
Identification
An automatic radiographic film processor is a device intended to be used to develop, fix, wash, and dry automatically and continuously film exposed for medical purposes.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Protec Optimax Automatic X-Ray Film Processor
Submission Summary (Full Text)
{0}------------------------------------------------
KO23801
# 510(k) Safety and Effectiveness Summary.
The following safety and effectiveness summary has been prepared pursuant to requirement for 510(k) summaries specified in 21CFR 807.92(a)
#### Submitter Information
FEB 1 2 2003
FERRANIA USA INC., Official Corrispondent Ferrania Imaging Technologies 2700-EAST FRONTAGE ROAD WEATHERFORD, OK 73096 Phone: (580) 774-3623 Fax: (580) 774-3658 Contact Person: Mr Doug Hale
S.a.i.e.p. S.r.l., Manufacturer Via Stalingrado 4/32 17014 Cairo Montenotte - GENOVA - ITALY Phone: 0039 019 506108 Fax: 0039 019 5091323 Contact Person: Mario Daniel
Date: 2002-10-29
Trade Name:
LIFERAY PRO 1000
Common Name: Automatic Radiographic Film Processor
Classification Name: Automatic Radiographic Film Processor
Regulation Number: 892.1900
Product Code: IXW
#### Predicate Device
Additional Substantial Equivalence Information is provided in the following Substantial Equivalence Comparison Table
{1}------------------------------------------------
## · Device Description:
Due to the precise roller transport system, both sheet and roller films can be processed. The Automatic film registration is activated immediately when a film is fed in. The transport system starts running. The film material is developed , fixed, rinsed and dried. With the easy to operate micro-processor, the processing conditions can be adjusted to suit the various film and chemical types. The developing solutions are temperatureregulated, circulated and automatically replenished. The feed with of the LifeRay PRO 1000 is 35cm, the smallest film format of the processor is 10x10cm.
#### • Intended Use:
The Automatic X-Ray Film Processor is intended to be used to process films exposed for medical purposes. The automatic and continuous process contains developing, fixing, washing and dryng of films.
#### Summary of Substantial Equivalence Comparison: .
The comparison of similarities and differences shows that the new device LifeRay PRO 1000 Automatic Radiographic Film Processor, which is intended to market and the predicate device:
- 1. Have the same intended use.
- 2. Have the same target user group.
- 3. Have the same technological characteristics.
There are no new questions about safety and effectiveness. The new device is a safe and effective as the predicate device.
The Automatic Radiographic Film Processor is substantially equivalent to the Protec Optimax Automatic X-Ray Film Processor.
### · Technological Characteristics:
The photographic processing (developing) technique employed by the Automatic Radiographic Film Processor is the same as the predicate device. The film medium is mechanically transported for immersion in two chemical baths (developer and fixer), is rinsed in water, dried, and then ejected for viewing. The processors use mechanical rollers and guides for transportation, the solutions are temperature-regulated, circulated and automatically replenished, and same additional functions (anti-crystallization cycle, stand-by mode). Both are controlled by an integrated software.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. The text is written in all capital letters.
FEB 1 2 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
S.A.I.E.P. S.r.1. % Mr. Mario Daniel Ferrania USA, Inc. Ferrania Imaging Technologies 2700 East Frontage Road WEATHERFORD OK 73096-6104 Re: K023801 Trade/Device Name: Film Processor, Life Ray Pro 1000 Regulation Number: 21 CFR 892.1900 Regulation Name: Automatic radiographic film processor Regulatory Class: II Product Code: 90 IXW Dated: October 29, 2002 Received: November 14, 2002
Dear Mr. Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indication for use.
Applicant:
510(k) Number (if known):
Device Name: Automatic Radiographic Film Processor LifeRay PRO 1000.
Indication For Use: The Automatic Radiographic Film Processor is intended to be used to process films exposed for medical purposes. The automatic and continuous process contains developing, fixing, washing and drying of films. This may be used in all general radiographic, diagnostic imaging procedures. Typical users of this system are trained medical professionals, including but not limited to physicians, nurses, and lab technicians.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED).
Concurrence of CDRH, Office of Device Evaluation (ODE).
1 Prescription Use (Per 21 CFR 801.109)
Daniel A. Legum
(Division Sign-Off) Division of Reproductive, Ab and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.