CANNON DIGITAL RADIOGRAPHY MODEL CXDI-40G

K023750 · Canon USA, Inc. · MQB · Nov 22, 2002 · Radiology

Device Facts

Record IDK023750
Device NameCANNON DIGITAL RADIOGRAPHY MODEL CXDI-40G
ApplicantCanon USA, Inc.
Product CodeMQB · Radiology
Decision DateNov 22, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1680
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

CANON DIGITAL RADIOGRAPHY CXDI-40G provides digital image capture for conventional film/screen radiographic examinations. The device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures.

Device Story

The Canon CXDI-40G is a digital radiography system that captures and converts conventional projection X-rays into digital images. It utilizes an amorphous silicon TFT array sensor to detect X-radiation generated by an external diagnostic X-ray source. The system includes a removable stationary grid to reduce scatter. It operates with various support structures, including upright stands, tables, and universal stands. The device processes captured signals to optimize gray-scale and displays sub-sampled preview images on a monitor. Diagnostic images are transmitted via a DICOM-compatible network for printing or storage. The system is intended for use by healthcare professionals in clinical settings to perform general radiographic examinations. By providing immediate digital image capture, the device eliminates the need for traditional film processing, potentially improving workflow efficiency and reducing patient exposure through optimized digital image management.

Clinical Evidence

Bench testing only. The submission relies on technical performance comparisons, including image size, pixel pitch, MTF, and dynamic range, demonstrating equivalence to predicate devices. No clinical data was required or provided.

Technological Characteristics

Amorphous silicon TFT array sensor; fluorescent screen (Gd2O2S:TB3+) X-ray absorber; 2688 x 2688 pixel resolution; 160 μm pixel pitch; 14-bit A/D conversion; 12-bit output data; DICOM network connectivity; stationary removable grid; complies with IEC 601-1-2.

Indications for Use

Indicated for general purpose diagnostic radiographic procedures in the general population to replace conventional film/screen systems.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary #### Prepared: August 8, 2002 K023750 #### Submitter: Company Name: Canon USA, Inc. (U.S. agent/official correspondent for Canon Inc.) Company Address: One Canon Plaza Contact Person: Phone Number: Fax Number: Lake Success, NY 11042 Ms. Sheila Driscoll (516) 328-5602 (216) 328-2169 #### Proposed Device: | Reason For 510(k): | New Model | |----------------------|-------------------------------| | Manufacturer: | Canon Inc. | | Trade Name: | Canon | | Model Name: | CXDI-40G | | Classification Name: | MQB, Solid State X-ray Imager | | FDA 510(k) #: | To be assigned | #### Predicate Device: | Manufacturer: | Canon Inc. | |----------------------|---------------------------------| | Trade Name: | Canon | | Model Name: | CXDI-11/CXDI-22 | | Classification Name: | 90MQB, Solid State X-ray Image- | | FDA 510(k) #: | K981556/K992547 | #### Description Of Device: The Canon digital radiography CXDI-40G is used to directly capture and convert conventional projection X-ray images to digital images. A sub-sampled image can be displayed on a preview monitor for viewing. The diagnostic image can be transmitted through a DICOM compatible digital network for printing. The device provides digital image capture for conventional film/screen radiographic examinations. The Canon digital radiography CXDI-40G is different from CXDI-11 and CXDI-22 in the following respect: · The CXDI-40G operates in conjunction with an upright stand, table, and universal stand while the CXDI-11 operates only in conjunction with an upright stand, and the CXDI-22 only operates in conjunction with a table. #### Intended Use: Canon digital radiography CXDI-40G provides digital image capture for conventional film/screen radiographic examinations. The device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. #### Technical Characteristics: Please refer to the attached COMPARISON CHART.(Table 10) {1}------------------------------------------------ ## Descriptive Comparison The predicate device is the Canon digital radiography CXDI-11 cleared under Document Number K981556 on November 4, 1998 and CXDI-22 cleared under Document Number K992547 on October 13, 1999. The CXDI-40G's technical specifications (including image size, pixel pitch, number of pixels), imaging principle, physical characteristics and intended use are the same as those of CXDI-11 and CXDI-22. However, the differences in the design are as follows: - · The mechanical structure and physical appearance of the CXDI-40G are different than the CXDI-11 and CXDI-22. Please refer to the attached CXDI-40G comparison table provided in this section. Additional information can also be found in the CXDI-40G Operation Manual provided in this submission. - · The CXDI-40G can be used with upright stand, table, and universal stand by using the installation unit for each application, while the CXDI-11 is used with upright stand only and the CXDI-22 is only used with a table. - · A removable, fixed grid is used for the CXDI-40G, while an implemented, moving grid is used for the CXDI-11 and a removable moving grid is used for the CXDI-22. Both types of grids are used for eliminating the dispersed X-ray in exposures that use films. (See attached Table 4.2) Regarding the software: - The system software for controlling CXDI-40G is released as V5.0. 8 - l V5.0 includes some changes from V4.1. - · The main changes of the V5.0 are the addition of the control of CXDI-40G sensor and some change of GUI. - · V4. I was first introduced and cleared under K003689 and is currently used in Canon models CXDI-11, CXDI-22, and CXDI-31. - · The comparison between V5.0 and V4.1 is described in the Software Information for model CXDI-40G section. Based on the information in this submission, similarity to the predicate devices (Canon digital radiography CXDI-11 and CXDI-22), and the results of our design control activities, it is our opinion that the Canon digital radiography CXDI-40G described in this submission is substantially equivalent to the predicate devices {2}------------------------------------------------ | comparison<br>of<br>ble<br>C<br>- | |-----------------------------------| | 0<br>T<br>ರ | | able | . | | Item | CXDI-11 | CXDI-22 | CXDI-40G | |--------------------------|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|---------------------------------------|---------------------------------------| | | FDA 510(k)# | K981556 | K992547 | Proposed Device | | | Intended Use | Provide diagnostic images for general | Provide diagnostic images for general | Provide diagnostic images for general | | | Design | radiography with upright system,<br>Digital acquisition,<br>electronic processing | radiography with table system | radiography | | | Energy Uses | Receives x-radiation generated by external diagnostic x-ray generator | Same | Same | | | X-ray Absorber | Fluorescent screen (Gd2O2S:TB3+) | Same | Same | | | Materials | Visible emission peak: 545nm | | | | Item | CXDI-11 | CXDI-22 | CXDI-40G | | | Sensing | Amorphous Silicon W/TFT Array | Same | Amorphous Silicon W/TFT Array | | | | Detection peak: 540- 620nm | Same | Detection peak: 540- 620nm | | | Anatomical Sites | General radiography | Same | Same | | | Target Population | General population | Same | Same | | | Physical Safety | Minimize exposure to x-radiation | Same | Same | | | Compliance with Standard | Complies with IEC 601-1-2 | Same | Same | | | Biocompatibility | N/A | N/A | N/A | | | Performance | After digital processing ( optimize the | Same | Same | | | | gray-scale) | Same | Same | | | Labeling | Approved 510(K) | Approved 510(K) | Approved 510(K) | | | Pixel | 2688 × 2688 pixels | Same | Same | | | Item | CXDI-11 | CXDI-22 | CXDI-40G | | | | (7,200,000 pixels) | Same | Same | | | Image size | 43cm x 43cm | Same | Same | | | Pixel pitch | 160 μ m | Same | Same | | | MTF | MTF@2lp/mm 42% | Same | Same | | | Dynamic Range | Dynamic range:<br>approximately 4 digit<br>( linear A/D : 14bit)<br>( output data : 12bit) | Same | Same | | | Grid | Moving grid | Moving grid<br>( removable ) | Stationary grid<br>( removable ) | | | Sensor Unit Dimension | 552 x 598 x 231mm | 604 x 645 x 73.50r69mm | 550x 550 x 68mm | | | Item | CXDI-11 | CXDI-22 | CXDI-40G | | | Power Supply Dimension | 580 x 489 x 275mm | 390 x 160 x 110mm | 390 x 160 x 110mm | | | Control PC Dimension | 483.5 x 594 x 300mm | 453 x 594 x 300mm | 453 x 594 x 300mm | | | Operation Unit Dimension | 298 x 209.5 x 130mm | Same | 398 x 395 x 150 | | | Card Reader Dimension | 50 x 180 x 39mm | Same | Same | | | Stand Dimension | 900 x 475 x 2100mm | N/A | (900 x 480 x 2000) * | | | Table Dimension | N/A | (2025 x 820 x 890) * | (2025 x 820 x 890) * | | {3}------------------------------------------------ : . {4}------------------------------------------------ · {5}------------------------------------------------ : *: These dimensions are anylicable for the stand and the table made by canon inc. It is possible to use another stands and tables. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with stylized wings and body. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Canon. Inc. % Ms. Pamela K. Gwynn Engineering Team Leader Underwriters Laboratories, Inc. 12 Laboratory Drive Research Triangle Park, NC 27709 AUG 23 2013 Re: K023750 Trade/Device Name: Canon Digital Radiography Model CXDI-40G Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: November 7, 2002 Received: November 8, 2002 Dear Ms. Gwynn: This letter corrects our substantially equivalent letter of November 22 2002. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {7}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Janine M. Morris Janine M. Mdrri Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ Section 3: Statement # Indications Statement Page l of I 510(K)Number(if known): KU23 Device Name: Indications for Use: CANON DIGITAL RADIOGRAPHY CXDI-40G provides digital image capture for conventional film/screen radiographic examinations. The device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHERT PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation(ODE) Prescription Use OR (Per 21 CFR 801.109) Over-The-Counter Use (Optional Format 1-2-96) David h. Lippman (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number 3-1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...