Browse hierarchy Clinical Chemistry (CH) Subpart B — Clinical Chemistry Test Systems 21 CFR 862.1660 Product Code JJY K023731 — QUICKCHECK LIQUID CHEMISTRY CONTROL (ASSAYED AND UNASSAYED), LEVELS 1, 2, AND 3
QUICKCHECK LIQUID CHEMISTRY CONTROL (ASSAYED AND UNASSAYED), LEVELS 1, 2, AND 3
K023731 · Consolidated Technology · JJY · Dec 24, 2002 · Clinical Chemistry
Device Facts
Record ID K023731
Device Name QUICKCHECK LIQUID CHEMISTRY CONTROL (ASSAYED AND UNASSAYED), LEVELS 1, 2, AND 3
Applicant Consolidated Technology
Product Code JJY · Clinical Chemistry
Decision Date Dec 24, 2002
Decision SESE
Submission Type Traditional
Regulation 21 CFR 862.1660
Device Class Class 1
Intended Use
QUICKCHECK LIQUID CHEMISTRY CONTROL, Levels 1, 2 and 3, is a liquid human serum based assayed quality control material intended to monitor the performance of clinical test procedures that analyze routine chemistry analytes as listed in this package insert.
Device Story
QUICKCHECK LIQUID CHEMISTRY CONTROL is a human serum-based quality control material used to monitor the performance of clinical chemistry test procedures. The device is provided in three levels (1, 2, and 3) and is available as either assayed or unassayed. It contains 59 analytes of clinical significance. The control is stored frozen at -20°C. In clinical practice, laboratory personnel use the control to verify the accuracy and precision of routine chemistry analyzers. By comparing the measured values of the control against the target ranges provided in the package insert or certificate of analysis, clinicians can assess the reliability of patient test results. This monitoring helps ensure the analytical performance of laboratory instrumentation, thereby supporting accurate clinical decision-making and patient diagnosis.
Clinical Evidence
No clinical data. The device is a quality control material; substantial equivalence is based on bench testing and comparison of physical/chemical properties to the predicate device.
Technological Characteristics
Human serum-based liquid control; preserved with ciprofloxacin; stored at -20°C. Available in three levels with 59 analytes. Configuration includes various fill sizes (4ml, 5ml, 10ml) in amber glass or plastic dropper vials. No software or electronic components.
Indications for Use
Indicated for use as a quality control material to monitor the performance of clinical laboratory test procedures for routine chemistry analytes. Intended for professional use in clinical laboratory settings.
Regulatory Classification
Identification A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
QUICKCHECK UNASSAYED CHEMISTRY CONTROL, Levels 1 and 2 (K973469 )
Related Devices
K140971 — LIQUID ASSAYED CHEMISTRY CONTROL PREMIUM PLUS LEVEL 1,2 AND LEVEL 3. · Randox Laboratories Limited · Dec 8, 2014
K100727 — LIQUID ASSAYED MULTIQUAL 1, LIQUID ASSAYED MULTIQUAL 2, LIQUID ASSAYED MULTIQUAL 3, LIQUID ASSAYED MULTIQUAL MINIPAK · Bio-Rad, Diagnostics Grp. · Sep 14, 2010
K043208 — LIQUID ASSAYED MULTIQUAL CONTROL · Bio-Rad · Jan 7, 2005
K033387 — QUEST DIAGNOSTICS SERUM CHEMISTRY CONTROL, MODELS 935 AND 945 · Bio-Rad · Dec 2, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
## Product Performance and Substantial Equivalency 510k Summarv
DEC 2 4 2002
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:_KOZ 37 31
- Submitter: Consolidated Technologies 4401 Freidrich Lane Building 1, Suite 100 Austin, TX 78744-1832 Phone: (512) 445-5100 (512) 445-5515 Fax:
Contact: Candice Betz Preparation date: October 31, 2002
### Product name (trade & common):
- QUICKCHECK LIQUID CHEMISTRY CONTROL (Assayed and Unassayed), Proprietary: Levels 1, 2 and 3
Common: N/A
#### Classification name:
Class I. Product code: JJY 21 CFR 862:1660: Multi-Analyte Controls, All Kinds (Assayed and Unassayed)
#### Predicate device:
QUICKCHECK UNASSAYED CHEMISTRY CONTROL, Levels 1 and 2. Consolidated Technologies, Inc. K-973469
#### Device description:
QUICKCHECK LIQUID CHEMISTRY CONTROL is designed to monitor the performance of test procedures that analyze a number of routine chemistry analytes as listed in the package insert.
This product will be offered as an assayed and unassayed control. In addition, the control is being expanded to three levels.
#### Intended use:
QUICKCHECK LIQUID CHEMISTRY CONTROL is a liquid human serum based assayed quality control material intended to monitor the performance of test procedures that analyze routine chemistry analytes as listed in the package insert.
Consolidated Technologies, Inc. 4401 Freidrich Lane Building 1, Suite 100 Austin Tx. 78744
{1}------------------------------------------------
#### Product Performance and Substantial Equivalency
510(k) Summary (continued)
#### Labeling/Packaging:
Consolidated Technologies is a contract manufacturer. The expected values listed on the package insert for QuickCheck Liquid Chemistry Control are target ranges developed for manufacturing purposes only. These expected values are based on the product having three distinct levels. As such, CTI will manufacture, at the request of a customer, either unassayed or assayed QuickCheck Chemistry Control product.
If the customer requests an unassayed product, the expected manufacturing targets for the product will be presented for reference and a certificate of analysis will accompany the final product for completion. The certificate of analysis will list the actual values obtained for analytes present in the control (see exhibit I for example)
If the customer requests an assayed control the package insert will list the analytes present in the control and the values obtained for those analytes on various test methods and/or instrumentation. The customer determines the test methods and instrumentation. The value assignment data (assayed) are provided to the customer to develop their own package insert. See exhibit II for an example of an assay sheet.
The customer is responsible for labeling and/or package insert for the finished device.
This product is offered to the customer in the following fill sizes:
5ml fill in a 7ml amber glass vial 4ml fill in a 7 ml amber glass vial 10ml fill in a 15ml amber glass vial 10ml fill in a 15 ml plastic dropper vial
Unlabeled vials in labeled flats Labeled vials in flats Labeled vials in Labeled kits
Consolidated Technologies is also a manufacturer and will sell this product as QuickCheck Liquid Chemistry Control, Level 1, 2, and 3. The package insert will list the analytes present in the control and the values obtained for those analytes on various test methods and/or instrumentation. The assay sheet included in this 510k is an example of potential methods used to assay the product. See Exhibit II, III, IV and V, for all labeling and package insert and example assay sheet.
QuickCheck Liquid Chemistry Control is sold in the following configurations:
5ml in a 7ml labeled vials 10ml in a 15ml labeled vials The above kitted are 5 of each level to equal 15 vials of the same level Package Insert
Consolidated Technologies, Inc. 4401 Freidrich Lane Building 1. Suite 100 Austin Tx. 78744
{2}------------------------------------------------
# Product Performance and Substantial Equivalency
510(k) Summary (continued)
## Comparative analysis:
The table below provides a summary of the technological characteristics between QUICKCHECK LIQUID CHEMISTRY CONTROL and the predicate device.
| Device<br>Characteristic | Proposed Device<br>QUICKCHECK LIQUID<br>CHEMISTRY CONTROL | Predicate Device<br>QUICKCHECK UNASSAYED<br>CHEMISTRY CONTROL |
|--------------------------|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Intended use | Assayed quality control serum for<br>monitoring performance of routine<br>chemistry test procedures. | Uassayed quality control serum for<br>monitoring performance of routine<br>chemistry test procedures |
| Matrix | Human Serum | Human Serum |
| Form | Liquid, Frozen | Liquid, Frozen |
| Analytes | 59 analytes of clinical significance<br>that may be found in serum. | 43 analytes of clinical significance<br>that may be found in serum |
| Levels | Three (3) levels | Two (2) levels |
| Preservatives | Ciprofloxacin | Ciprofloxacin |
| Storage | -20°C | -20°C |
| Stability | Until expiration date noted on vial<br>label. | Until expiration date noted on vial<br>label |
## Conclusions:
The information provided in the pre-market notification demonstrates that QUICKCHECK LIQUID CHEMISTRY CONTROL (Assayed and Unassayed) is substantially equivalent to the predicate device, for which there is FDA clearance. This equivalence was demonstrated through comparison of intended uses and physical properties to a commercially available The information supplied in the pre-market notification provides reasonable device. assurance that QUICKCHECK LIQUID CHEMISTRY CONTROL is safe and effective for the stated intended use.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" around the perimeter. Inside the circle is a stylized design featuring four wavy lines that resemble a human figure.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
DEC 24 2002
ood and Drug Administrati 2098 Gaither Road Rockville MD 20850
Ms. Candice Bentz Quality Manager Consolidated Technologies, Inc 4401 Freidrich Lane Building 1, Suite 100 Austin, TX 78744
Re: k023731
> Trade/Device Name: QuickCheck Liquid Chemistry Control (Assayed and Unassayed). Levels 1. 2. and 3 Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I Product Code: JJY Dated: October 31, 2002 Received: November 6, 2002
Dear Ms. Bentz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS OF USE STATEMENT
© 2373
510(k) number (if known): K
Device name:
QUICKCHECK LIQUID CHEMISTRY CONTROL (Assayed and Unassayed), Levels 1, 2 and 3
Indications for use:
QUICKCHECK LIQUID CHEMISTRY CONTROL, Levels 1, 2 and 3, is a liquid human serum based assayed quality control material intended to monitor the performance of clinical test procedures that analyze routine chemistry analytes as listed in this package insert.
Sea
ean
Division Sign-Off)
Clinical Laboratory Device.
510(K) Nr. K033731
.
prescription use
Consolidated Technologies, Inc. 4401 Freidrich Lane Building 1, Suite 100 Austin Tx. 78744