ASCENSIA DEX 2 DIABETES CARE SYSTEM

K023584 · Bayer Corp. · NBW · Nov 20, 2002 · Clinical Chemistry

Device Facts

Record IDK023584
Device NameASCENSIA DEX 2 DIABETES CARE SYSTEM
ApplicantBayer Corp.
Product CodeNBW · Clinical Chemistry
Decision DateNov 20, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2

Intended Use

The ASCENSIA™ DEX®2 Diabetes Care System is for the Self-Monitoring of Blood Glucose as an adjunct to the care of person with diabetes. 4

Device Story

The ASCENSIA™ DEX®2 Diabetes Care System is a blood glucose meter used by patients and healthcare professionals in home or clinical settings. It utilizes electrochemical amperometric glucose oxidase technology to measure glucose levels in whole blood samples. The system uses test sensor discs, which are individually sealed in packages of ten. The meter processes the electrochemical signal from the test strip to provide a quantitative blood glucose result, which is referenced to plasma glucose. The device provides a linear response for glucose concentrations ranging from 10-600 mg/dL. Healthcare providers use these results to monitor blood glucose levels, which serves as an adjunct to the management of diabetes. The system is designed for over-the-counter (OTC) use.

Clinical Evidence

In-house and clinical evaluations were conducted by healthcare professionals and persons with diabetes. Studies demonstrated that users obtain blood glucose results substantially equivalent to current blood glucose monitoring methods. No specific sensitivity, specificity, or p-values were provided in the summary.

Technological Characteristics

Amperometric glucose oxidase method; electrochemical sensing; linear range 10-600 mg/dL; results referenced to plasma glucose; test sensor discs packaged in sets of ten.

Indications for Use

Indicated for persons with diabetes and healthcare professionals in home and clinical settings for the measurement of glucose in whole blood as an adjunct to diabetes care.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ NOV 2 0 2002 Page 1 of 2 # 510(k) Summary of Safety and Effectiveness # Submitter Information | Contact person: | George M. Tancos | |-----------------|----------------------------------------| | Address: | 1884 Miles Avenue<br>Elkhart, IN 46515 | | Phone: | 574.262.6928 | | FAX: | 574.262.6945 | | E-mail: | george.tancos.b@bayer.com | | Date Prepared: | October 11, 2002 | ### Device Information | Proprietary Name: | ASCENSIA™ DEX® 2 Diabetes Care System | |-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Blood Glucose Meter | | Classification: | Division of clinical laboratory Devices Panel --- Clinical<br>Chemistry and Toxicology classification Code 75 CGA (Glucose<br>Oxidase, Glucose) | # Predicate Device Information | Name: | ASCENSIA™ DEX® 2 Blood Glucose Meter | |---------------|-------------------------------------------------------------------| | Manufacturer: | Bayer Diagnostics<br>430 S. Beiger Street<br>Mishawaka, In 46544 | 510(k) Number: K020210 # Device Description The ASCENSIA™ DEX®2 Diabetes Care System consists of an electrochemical method-based meter and dry reagent sensor (test strips) designed for testing glucose by persons with diabetes or by healthcare professionals in the home or in healthcare facilities. {1}------------------------------------------------ #### Statement of Intended Use: The ASCENSIA™ DEX®2 Diabetes Care System is for the Self-Monitoring of Blood Glucose as an adjunct to the care of person with diabetes. 4 #### Summary of Technological Characteristics: The ASCENSIA™ DEX®2 Diabetes Care System employs an amperometric glucose oxidase method to measure glucose in blood. It is conceptually the same as other blood glucose monitoring products available for blood glucose testing. The test sensor discs are individually sealed in a package of ten. Blood glucose results are referenced to plasma glucose. The System has a linear response to glucose from 10-600 mg/dL. # Performance Data: An evaluation of the ASCENSIA™ DEX® 2 Diabetes Care System was studied in-house and in clinical settings by healthcare professionals and by persons with diabetes. The studies demonstrated that users can obtain blood glucose results that are substantially equivalent to current methods for blood glucose. # Conclusion: The results of in-house and clinical evaluations of the ASCENSIA™ DEX® 2 Diabetes Care System demonstrate that the device is equivalent in performance to the predicate devices and suitable of its intended use. <sup>&</sup>quot;Consensus Statement on Self-Monitoring of Blood Glucose, " Diabetes Care, Vol. 10, No. 1, January-February 1987, pp. 95.99 {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with three parallel lines above them, resembling an abstract representation of a bird or eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Mr. George M. Tancos R.A.C. Manager Regulatory Affairs Bayer Corporation 1884 Miles Avenue P.O. Box 70 Elkhart, IN 46515-0070 Re: k023584 . Trade/Device Name: ASCENSIA™ DEX®2 Diabetes Care System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW; CGA Dated: October 22, 2002 Received: October 25, 2002 Dear Mr. Tancos: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. NOV 2 0 2002 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): KOB3584 Device Name: ASCENSIA™ DEX®2 Diabetes Care System Indications for Use: The ASCENSIA™ DEX®2 Blood Glucose Meter is used with ASCENSIA™AUTODISC™ Blood Glucose Test Strips GLUCOMETER® DEX® Test Sensor and Controls for the measurement of glucose in whole blood. The ASCENSIA™ DEX®2 Diabetes Care System is an Over-The-Counter (OTC) device used by persons with diabetes and by healthcare professionals in home settings and in healthcare facilities. The frequent monitoring of blood glucose is an adjunct to the care of persons with diabetes. (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Stan Cooper (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number k623594 Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use V
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%