I-STAT KAOLIN ACT TEST

K023582 · I-Stat Corporation · JBP · Sep 8, 2003 · Hematology

Device Facts

Record IDK023582
Device NameI-STAT KAOLIN ACT TEST
ApplicantI-Stat Corporation
Product CodeJBP · Hematology
Decision DateSep 8, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7140
Device ClassClass 2

Indications for Use

The i-STAT Kaolin Activated Clotting Time (ACT) test is an in vitro diagnostic test used to monitor high-dose heparin anticoagulation frequently associated with cardiovascular surgery. The test is to be used with the i-STAT Portable Clinical Analyzer (Models 200 and 300), but not the Philips Medical Systems (formerly Agilent Technologies) Blood Analysis Module (BAM). As part of the i-STAT System, the Kaolin ACT test is to be used by trained and certified health care professionals in accordance with a facility's policies and procedures.

Device Story

i-STAT Kaolin ACT Test is a single-use cartridge for point-of-care monitoring of high-dose heparin anticoagulation; used during cardiovascular surgery. Input: 40 ul fresh whole blood sample. Operation: sample mixed with kaolin activator and synthetic thrombin substrate within cartridge; coagulation cascade activation leads to thrombin-mediated conversion of electrogenic substrate (MeoPADA). Output: clotting time in seconds, measured via electrochemical sensor detecting current plateau. Used by clinicians in point-of-care settings; analyzer provides quantitative results to guide heparin dosing and anticoagulation management, potentially improving patient safety during surgery.

Clinical Evidence

Method comparison studies conducted at three clinical sites (N=976 total) using cardiovascular patients (ages 44-85). Results showed strong correlation (R=0.905-0.974) with predicate devices. Precision testing (NCCLS EP-5A) across three lots and operators yielded 0.1-5.4% CV. No clinical sensitivity/specificity data provided as the device is an in vitro diagnostic test for monitoring.

Technological Characteristics

Single-use cartridge containing kaolin activator, thrombin substrate, and inert matrix. Electrochemical sensing principle. Dimensions/form factor: cartridge for portable analyzer. Connectivity: standalone analyzer. Software: automated analysis cycle. Sterilization: not specified.

Indications for Use

Indicated for monitoring high-dose heparin anticoagulation in patients undergoing cardiovascular surgery. For use by trained and certified healthcare professionals.

Regulatory Classification

Identification

An activated whole blood clotting time tests is a device, used to monitor heparin therapy for the treatment of venous thrombosis or pulmonary embolism by measuring the coagulation time of whole blood.

Predicate Devices

Submission Summary (Full Text)

{1} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: #K023582 B. Analyte: Activated Whole Blood Clotting Time (ACT) C. Type of Test: Quantitative D. Applicant: i-STAT Corporation E. Proprietary and Established Names: i-STAT® System Kaolin ACT Test F. Regulatory Information: 1. Regulation section: CFR 864.7140 2. Classification: Class II 3. Product Code: JBP 4. Panel: Hematology (81) G. Intended Use: 1. Intended use(s): The i-STAT Kaolin ACT is an *in vitro* diagnostic test which uses fresh whole blood to monitor high-dose heparin anticoagulation frequently associated with cardiovascular surgery. 2. Indication(s) for use: [Same as Intended Use] 3. Special condition for use statement(s): N/A 4. Special instrument Requirements: i-STAT Portable Clinical Analyzers, Models 200 (#K940918 and #K001154); and 300 (#K001387). H. Device Description: The i-STAT Kaolin ACT Test consists of a single test cartridge, upon which a fresh whole blood sample of 40 ul is placed. The Kaolin ACT Test cartridge uses the same electrogenic substrate technology as the i-STAT Celite ACT Test cartridge. Both are composed of plastic components that house sensor chips and fluid conduits. Reagents are coated on the cartridge cover and consist of an activator, thrombin substrate and inert matrix components. The sample is mixed with the reagents in the sensor channel, where activation of the coagulation cascade and detection of clot formation occur. {2} Page 2 of 5 # I. Substantial Equivalence Information: 1. Predicate device name(s): International Technidyne Corporation (ITC) Hemochron® Activated Whole Blood Clotting Time (K-ACT Test tube); i-STAT Corporation Celite Activated Clotting Time Test cartridge. 2. Predicate K number(s): #K913861; #K992571 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | For in vitro diagnostic use in monitoring high-dose heparin anticoagulation frequently associated with cardiovascular surgery. | For in vitro diagnostic use in monitoring heparin anti-coagulation during cardio-vascular procedures. | | Samples | Fresh whole blood | Whole blood | | Measuring temper- ature | 37° C. | 37° C. | | Measuring units | Seconds | Seconds | | Reference range – (normal donor, venous) | 89 – 137 seconds | 91 – 151 seconds | | Reference range (CPB patients, arterial) | 107 – 160 seconds | 110 – 154 seconds | | Heparin linearity/ prolongation | 50 – 1000 seconds/30 - 118 seconds per U/mL heparin. | 0 – 1500 seconds/33 – 109 seconds per U/mL heparin. | | Inhibited platelet effect | Extension | Extension | | Aprotinin effect (extension) | Approximately 3.0% | Approximately 3.2% | | | | | {3} Page 3 of 5 | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Sample volume | 0.04 mL | 2.0 mL | | Detection method | Electrochemical marker | Mechanical (magnetic) | | Reagent format | Test cartridge with kaolin activator and synthetic thrombin substrate. | Glass test tube with kaolin activator | | Display for ‘out-of-range’ results | >1000 seconds | flashing 1500 seconds | | Imprecision (controls) (whole blood, total) | 2.1 – 4.8 % CV 9.8 –13.2 % | 5.3 – 6.2% CV 11.0 –19.2% | | Sample temperature effect | None | Inversely proportional: ~1 % prolongation for each 1° C. decrease below 37° C. | | Fibrinogen sensitivity | Insensitive between 105 – 514 mg/dL. | Insensitive < 100 mg/dL. | J. Standard/Guidance Document Referenced (if applicable): N/A K. Test Principle: The i-STAT ACT Test endpoint is indicated by the appearance of an electroactive marker generated by thrombin-mediated conversion of a synthetic substrate included in the reagent. It employs the same electrogenic substrate technology as the company’s Celite ACT Test (#K992571). A whole blood sample is introduced into the sample well of the cartridge which is then placed into the analyzer. The test endpoint is indicated by the time that corresponds to the appearance of a current plateau, and is due to the conversion of the “MeoPADA” substrate detected by the electrochemical sensor. L. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Various imprecision components were tested in-house and at (3) clinical sites, according to NCCLS EP-5A document. Two levels of controls were tested on (3) cartridge lots by (3) different operators. All results ranged from 0.1 – 5.4% CV. {4} Page 4 of 5 b. Linearity/assay reportable range: Heparin sensitivity was determined by a dose response of 1.5 – 8 U/mL heparin over a range of 50 – 1000 seconds. c. Traceability (controls, calibrators, or method): The i-STAT Celite ACT substrate, NIST certified reference electrode and gold film sensor for the amperometric endpoint. d. Detection limit: 1.5 mL heparin e. Analytical specificity: No fibrinogen interference &lt; 100 mg/dL; Factor deficiencies and other coagulopathies may cause interference. f. Assay cut-off: N/A 2. Comparison studies: a. Method comparison with predicate device: Studies were performed at (3) representative point-of-care clinical sites, using (3) lots of i-STAT Kaolin cartridges. Samples were obtained from male and female cardiovascular patients, ranging from 44 – 85 years of age. Some samples contained aprotinin and/or protamine; and all hematocrits ranged between 14 – 42%. i. Data from the University of Michigan Medical Center (N = 311) ranged: Y = 111 – 118 S = 0.949 – 0.982 I = -14 to -10 R = 0.905 – 0.910 ii. Data from the Baptist Hospital (N = 352) ranged: Y = 76 – 87 S = 1.040 – 1.075 I = -46 to -34 R = 0.934 – 0.946 iii. Data from the Florida Hospital (N = 313) ranged: Y = 60 – 61 S = 0.913 – 1.047 I = -56 to -29 R = 0.971 – 0.974 b. Matrix comparison: N/A {5} Page 5 of 5 3. Clinical studies: a. Clinical sensitivity: N/A b. Clinical specificity: N/A c. Other clinical supportive data (when a and b are not applicable): Other clinical data were presented in various literature references included in the Original submission and the Supplement, dated 7/24/03. 4. Clinical cut-off: N/A 5. Expected values/Reference range: The reference range for normal volunteers (N = 120) on venous blood is 89 – 137 seconds (95% C.I.). The range for pre-procedural baseline (arterial) blood (N = 79) is 107 – 160 seconds (90% C.I.). M. Conclusion: There is no reference standard for the ACT Test, however, based upon extensive studies performed with the proposed i-STAT Kaolin ACT, and from data and information submitted in several literature references, the device is considered substantial equivalent to the ITC Hemochron ® Activated Whole Blood Clotting Time (K-ACT) Test.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...