SIEMENS MULTIVIEW WORKSTATION MODIFICATIONS

K023569 · Siemens Medical Solutions USA, Inc. · MHX · Nov 4, 2002 · Cardiovascular

Device Facts

Record IDK023569
Device NameSIEMENS MULTIVIEW WORKSTATION MODIFICATIONS
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeMHX · Cardiovascular
Decision DateNov 4, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1025
Device ClassClass 2

Indications for Use

The intended use of Siemens MultiView WorkStation & INFINITY Network is to act as a communications network, central monitoring device, and remote display for Siemens Patient Monitoring Systems and recorders.

Device Story

INFINITY MultiView WorkStation (MVWS) functions as a central monitoring station, communications network, and remote display for Siemens patient monitoring systems. It receives physiological data from bedside monitors. Software version VF3 introduces: Waveform Export (25 hours of full disclosure waveforms via NFS protocol); Caliper application (manual measurement of 18-second ECG segments from full disclosure database); Selected Strip Report (user-defined channel segments); and support for a second SpO2/Pulse parameter. Used in clinical settings by healthcare providers to view patient data, perform measurements, and generate reports. Output facilitates clinical decision-making by providing centralized access to patient physiological status and historical waveform data.

Clinical Evidence

No clinical data. Bench testing performed according to internal design control procedures.

Technological Characteristics

Software-based central monitoring station. Connectivity via standard network protocols (NFS). Features include ECG waveform measurement tools (calipers), data export functionality, and multi-parameter (SpO2/Pulse) support. Not MRI compatible.

Indications for Use

Indicated for use as a central monitoring device, communications network, and remote display for Siemens Patient Monitoring Systems and recorders.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 1. 510(k) SUMMARY as required per 807.92(c) ### Submitters Name, Address: Siemens Medical Solutions, Inc. Electromedical Systems Group, PCS Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879 Establishment Registration Number: 1220063 Official Correspondent: Connie Hertel, Director, QA/RA Contact person for this submission: Penelope H. Greco Date submission was prepared: October 21, 2002 Trade Name, Common Name and Classification Name: - A. Trade Name: INFINITY MultiView WorkStation - B. Common Name, Classification Name, Class and Regulation Number: | Common Name | Product<br>Code | Class | Regulation<br>Number | |--------------------------------------------------------------|-----------------|-------|----------------------| | Detector and Alarm, Arrhythmia | DSI | III | 870.1025 | | System, Network and Communication,<br>Physiological Monitors | MSX | II | 870.2300 | #### Legally Marketed Device Identification: Olympus Communications Network, SC 3000 WorkStation and Remote Display K955029 (now referred to as the Infinity Network, MultiView WorkStation and Remote Display) K003246 Infinity VentViewer (now known as VentCentral) ## Device Description: ### Introduction Some features included in the INFINITY MultiView WorkStation VF3 software release are locked options that will be submitted separately. Page 1 of 2 ## Siemens Medical Solutions, Inc. Electromedical Systems Group, PCS 16 Electronics Avenue Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879 {1}------------------------------------------------ K023569 page 2 of 2 The INFINITY MultiView WorkStation (MVWS) software version VF3 modifications include: ## Waveform Export Waveform Export allows the user to export 25 hours of the four full disclosure waveforms to any client device. The export protocol uses Native File System (NFS), a standard file transfer protocol. #### Calipers The release of software version VF3 includes the caliper application, which is an extension of the Full Disclosure database application. The caliper application allows the user to select an 18-second ECG waveform segment from the Full Disclosure database and perform simple caliper measurements. ### Selected Strip Report The Selected Strip Report allows the user to select an area of the topmost channel within Full Disclosure by defining the beginning and ending cursor times. ### Second SpO2 and Pulse Parameter Support Minor software modifications have been implemented to support a second SpO2 and Pulse parameter received from the bedside. The Indications for Use have not changed with the implementation of Waveform Export, Calipers, or support for a second SpO2 and Pulse parameter received from the bedside. Testing has been performed according to internal design control procedures and indicates no affect on the safety or efficacy of the MultiView WorkStation. #### Intended Use: The intended use of Siemens MultiView WorkStation & INFINITY Network is to act as a communications network, central monitoring device, and remote display for Siemens Patient Monitoring Systems and recorders. Assessment of non-clinical performance data for equivalence: Section J Assessment of clinical performance data for equivalence: Not applicable Biocompatability: Not applicable Sterilization: Not applicable Standards and Guidances: Section E Page 2 of 2 ## Siemens Medical Solutions, Inc. Electromedical Systems Group, PCS 16 Electronics Avenue Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is an abstract image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville .MD 20850 NOV 4 2002 Siemens Medical Solutions, Inc. c/o Ms. Penelope H. Greco Regulatory Submissions Manager Electromedical Systems Group, PCS 16 Electronics Avenue Danvers, MA 01923 Re: K023569 Trade Name: INFINITY MultiView WorkStation Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: Class III (three) Product Code: MHX Dated: October 21, 2002 Received: October 24, 2002 Dear Ms. Greco: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 - Ms. Penelope H. Greco Please be advised that F.DA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Oktaten B. Zeller, M.P. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 . 510(k) Number (if known): Device Name: INFINITY MultiView WorkStation Indications for Use: Siemens INFINITY MultiView WorkStation, INFINITY Network and Remote Display are indicated for use as a central monitoring device, communications network, and remote display for Siemens Patient Monitoring Systems and recorders. and the country of the country of the county of MRI Compatibility Statement: The MultiView WorkStation and Infinity Network are not compatible for use in a MRI magnetic field. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96) Athin Division of Cardiovascular & Respiratory Devices 510(k) Number K023564
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...