The Neuroscan Nuamps® system is intended for the measuring, recording and analysis of the electrical activity of a patient's brain and/or through the attachment of multiple electrodes at various locations to aid in monitoring and diagnosis as routinely found in clinical settings for EEG. The system is intended for the EEG and long/middle-latency EP registration in the research environment
Device Story
Nuamps is a 40-channel EEG amplifier; performs signal amplification, analog-to-digital conversion, and filtering. Inputs: electrical signals from patient brain via multiple electrodes. Operation: device samples, acquires, and transfers data to a host computer; software controls channels, performs real-time digital filtering, calculates compressed spectrum arrays (CSA), and generates annotated signal plots or 2/3-D topographic/tomographic maps. Hardware includes a 12-bit trigger board for stimulus timing. Used in clinical and research settings by trained personnel. Output: visual EEG data and analysis results displayed on host computer. Clinical utility: aids in monitoring and diagnosis of neurological activity; provides high-resolution data acquisition.
Clinical Evidence
Bench testing only. Performance verified through compliance with IEC-60601-1, IEC-60601-1-1, IEC-60601-1-1-2, IEC-60601-1-1-4, IEC-60601-1-2-26, EN46001, EN ISO9001:2000, MDD 93/42/EEC, AAMI EC53-1995, and CDRH guidance on electrode lead wires. No clinical data presented.
Indicated for measuring, recording, and analyzing brain electrical activity in adults, children, and infants to aid in clinical EEG monitoring and diagnosis, and for EEG and long/middle-latency EP registration in research environments.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
#### V. 510(k) Summary
Safety and effectiveness information concerning this device is summarized below. Because this is not a Class III device, the special certification defined in this section is not required.
| Submitted by: | Neuroscan<br>7850 Paseo Del Norte, Suite 101<br>El Paso, TX 79912<br>Phone: 915-845-5600 ext 217<br>Fax: 915-845-2965 | JAN 1 5 2003 |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------|--------------|
| Contact Person: | Elvira Garcia | |
| Date Prepared: | October 9, 2002 | |
| Proprietary Name: | Nuamps® | |
| Common Name: | Electroencephalograph (EEG) | |
| Classification Name: | Electroencephalograph (GWQ), OLT | |
| Device Classification<br>Class II: | 21 CFR § 882.1400 Electroencephalograph | |
| Predicate Device: | SCAN LT<br>510(k) # K001564 | |
Description of the Device: The Neuroscan Nuamps® is a 40-channel EEG amplifier capable of direct current (DC) recording, including signal amplification, analog-to-digital conversion, and filtering. Nuamps® permits high-speed simultaneous sampling, acquisition and transfer of data host computer that controls, displays, and stores the acquired data. Nuamps®'s software routines separately control each channel and perform real-time digital filtering. Nuamps® software measures and analyzes EEG signals and performs analysis of complete data sets, calculates compressed spectrum arrays (CSA), and presents results as annotated signal plots or topographic/tomographic maps in real-time two or threedimensional (2/3-D) context. Nuamps® is optically isolated and transformers are available for line voltages of 100, 120, 230, VAC.
The Neuroscan Nuamps® systems works in the same manner as the approved and predicate device, and:
- permits 1 to 40 channels configuration, and .
- simplifies the acquisition, recording and analysis of the data generated in . high-resolution
{1}------------------------------------------------
The Neuroscan SCAN LT was slightly modified to make the Neuroscan Nuamps. The modification does not affect the safety or effectiveness of the device. The modification consisted of a hardware change. We added a 12-bit trigger board that adds timing information on stimulus presentation to the EEG data collected by the amplifier.
## Statement of Indications For Use:
The Neuroscan Nuamps® system is intended for the measuring, recording and analysis of the electrical activity of a patient's brain and/or through the attachment of multiple electrodes at various locations to aid in monitoring and diagnosis as routinely found in clinical setting for EEG.
Patient Population: Adults, children and infants
### 510(k) Summary of Non-Clinical Testing
The following is a list of test performed on the Neuroscan Nuamps® system. These tests demonstrate that the performance of the system is equivalent to that of the predicate devices in the terms of safety and effectiveness, and that the additional features provide utility and product performance which exceed that of the predicate devices. All tests were completed satisfactorily without adverse report.
The Neuroscan Nuamps® system was designed and is manufactured to comply with:
- IEC-60601-1 ●
- IEC-60601-1-1 ●
- IEC-60601-1-1-2
- IEC-60601-1-1-4 .
- IEC-60601-1-2-26 .
- EN46001 ●
- EN ISO9001:2000 ●
- MDD 93/42/EEC .
- AAMI EC53-1995 ●
- CDRH Guidance Document on the "Performance Standard for Electrode ● Lead Wire and Patient Cables," March 9, 1998
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Neuroscan Elvira Garcia Quality Assurance Manager 7850 Paseo del Norte, Suite 101 El Paso, Texas 79912
Re: K023536
Trade/Device Name: NuAmps Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: OLT, GWQ Dated (Date on orig SE ltr): October 9, 2002 Received (Date on orig SE ltr): October 21, 2002
APR - 9 2012
Dear Ms. Garcia:
This letter corrects our substantially equivalent letter of January 15, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Statement of Indications for Use IV.
| Applicant: | Neuroscan<br>7850 Paseo Del Norte, Suite 101<br>El Paso, TX 79912<br>Phone: 915-845-5600 ext. 217<br>Fax: 915-845-2965 |
|------------|------------------------------------------------------------------------------------------------------------------------|
|------------|------------------------------------------------------------------------------------------------------------------------|
51(k) Number: Ko23536
Neuroscan Nuamps® Device Name:
Indications For Use: The Neuroscan Nuamps® system is intended for the measuring recording and analysis of the electrical activity of a patient's brain and/or through the attachment of multiple electrodes at various locations to aid in monitoring and diagnosis as routinely found in clinical settings for EEG. The system is intended for the EEG and long/middle-latency EP registration in the research environment
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of device Evaluation (ODE)
R. Mark N. Melkman
(Division Sign-Off) Division of Coneral, Res rative and Neurological Devices
:10(k) Number K 023536
K023536 mnr
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
Nuamps 510(k) Application
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.