The intended use of this incubator is to provide an environment with controlled temperature, CO2, elevated humidity, an automatic decontamination mode, and O2 at suppressed levels, for the development of ova or embryos at or near body temperature.
Device Story
The HERAcell 240 is a benchtop or floor-standing incubator used in clinical laboratory settings to support the development of ova or embryos. It maintains physiological conditions by controlling temperature, CO2 levels, humidity, and O2 concentrations (including hypoxic conditions). The device features an automatic decontamination mode and a water level alarm. Inputs include user-defined setpoints for environmental parameters; these are processed by an internal microprocessor to regulate the chamber environment. Options include stainless steel or copper inner chambers, an infra-red CO2 sensor, and a roller bottle system. The device provides a stable, controlled environment that mimics physiological conditions, assisting healthcare professionals in assisted reproduction procedures. Output is the physical environment itself, which facilitates embryo development.
Clinical Evidence
Bench testing only. The device complies with internationally accepted electrical safety standards and CE marking requirements. No clinical data provided.
Technological Characteristics
Benchtop/floor-standing incubator. Materials: stainless steel or solid copper inner chamber. Sensing: Thermal conductivity (TC) sensor (standard) or optional Infra-red CO2 sensor. Energy: Electrical. Control: Microprocessor-based regulation of temperature, CO2, humidity, and O2. Features: Automatic decontamination mode, water level alarm, optional roller bottle system. Connectivity: Not specified.
Indications for Use
Indicated for the development of ova or embryos at or near body temperature in a controlled environment.
Regulatory Classification
Identification
Assisted reproduction accessories are a group of devices used during assisted reproduction procedures, in conjunction with assisted reproduction needles and/or assisted reproduction catheters, to aspirate, incubate, infuse, and/or maintain temperature. This generic type of device may include:(1) Powered aspiration pumps used to provide low flow, intermittent vacuum for the aspiration of eggs (ova). (2) Syringe pumps (powered or manual) used to activate a syringe to infuse or aspirate small volumes of fluid during assisted reproduction procedures. (3) Collection tube warmers, used to maintain the temperature of egg (oocyte) collection tubes at or near body temperature. A dish/plate/microscope stage warmer is a device used to maintain the temperature of the egg (oocyte) during manipulation. (4) Embryo incubators, used to store and preserve gametes and/or embryos at or near body temperature. (5) Cryopreservation instrumentation and devices, used to contain, freeze, and maintain gametes and/or embryos at an appropriate freezing temperature.
Special Controls
*Classification.* Class II (special controls) (design specifications, labeling requirements, and clinical testing). The device, when it is a simple embryo incubator with only temperature, gas, and humidity control; a syringe pump; a collection tube warmer; a dish/plate/microscope stage warmer; a controlled-rate cryopreservation freezer; or an assisted reproduction laminar flow workstation is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
# KO23465
510 (k) Summary fas required by 21 CFR 807.92]
Date Prepared [21 CFR 807.92(a)(1)|
October 14, 2002
### Submitter's Information [21 CFR 807.92(a)(1)]
Joseph M. Azary C/o Kendro Laboratory Products L.P. P.O. Box 2156 Huntington, CT. 06484
Azary Technologies has received authorization to submit this 510(k) on behalf of the sponsor Kendro Laboratory Products L.P., 31 Pecks Lane, Newtown, CT 06470.
## Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
The device trade names are: Kendro Laboratory Products HERAcell 240 Common Name: Assisted Reproduction Accessories, Embryo Incubator
## Predicate Device [21 CFR 807.92(a)(3)]
- . Kendro HERAcell CO2 Incubator - K002805
There are a few minor differences between the subject device (HERAcell 240) and the predicate device (HERAcell). The differences revolve around the options that are now offered with the HERAcell 240 including Oxygen O2 control ranges (1-21% for applications requiring hypoxic conditions or a wide range of 5-90%), Infra-red CO2 Sensor, Roller Bottle System, and the Water Level Alarm.
The predicate device allowed the user to control the Carbon Dioxide levels directly and indirectly control Oxygen and Nitrogen levels by the manipulation of CO2 levels. The subject device allows for the direct control of Carbon Dioxide and Oxygen. The concept of controlling Oxygen levels in incultators is not new. The Forma Scientific Incubator (510k# K991408) allows for the direct control of oxygen levels (2-20%).
#### Description of the Device [21 CFR 807.92(a)(4)]
The Kendro HERAcell 240 incubators are bench top or floor standing units. The devices control CO2, temperature, provide elevated humidity, feature a decontamination mode, and control O- at suppressed levels to ensure the proper development of ova or embryos at or near body temperature. The conditions produced by the subject device are an exact simulation of physiological conditions. Controlled parameters and alarin functions are microprocessor controlled.
{1}------------------------------------------------
The basic model includes the auto-zero self-balancing CO2 control system, which ensures long term stability of CO2 levels. Each unit is equipped with a water level alarm to remind the user to refill the water reservoir.
As with the predicate device the user has the choice of a stainless steel or solid copper inner chamber. Additionally, the user can choose from additional options such as Oxygen control, Infra-Red Sensor, and the roller bottle system. The basic model includes the TC (thermal conductivity) sensor, however the user can choose the Infra-Red Sensor option.
The main differences between the subject device and the predicate device are related to new options that are offered with the new device as listed below:
- Oxygen O2 control ranges (1-21% for applications requiring hypoxic conditions or a wide . range of 5-90%).
- Infra-red CO2 Sensor. .
- Roller Bottle System (a system with up to 4 bottle turning devices for roller between . 58-186mm in diameter).
- . Water Level Alarm (included in all HERAcell 240 units)
## Intended Use [21 CFR 807.92(a)(5)]
The intended use of this incubator is to provide an environment with controlled temperature, CO2, elevated humidity, an automatic decontamination mode, and O2 at suppressed levels, for the development of ova or embryos at or near body temperature.
#### Technological Characteristics |21 CFR 807.92(a)(6)]
Kendro Laboratory Products L.P. believes that the addition of a few options in the subject device is a minor change from the predicate device. Therefore, it is believed that the substantially equivalent to the predicate device.
#### Performance Data [21 CFR 807.92(b)(1)]
The subject device complies with internationally accepted electrical safety standards as well as other requirements such as CE marking.
## Conclusion [21 CFR 807.92(b)(3)]
We believe the changes are minor and conclude that the subject devices are as safe and effective as the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a human figure with three overlapping profiles, suggesting a sense of community and service.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Kendro Laboratory Products, L.P. % Mr. Joseph M. Azary Azary Technologies, LLC P.O. Box 2156 HUNTINGTON CT 06484
Re: K023465
MAR 1 0 2003
Trade/Device Name: HERAcell 240 Regulation Number: 21 CFR 884.6120 Regulation Name: Assisted reproduction accessories Regulatory Class: II Product Code: 85 MQG Dated: January 18, 2003 Received: January 21, 2003
Dear Mr. Azary:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
5 10(k) Number (if known): _
): KD23465
Device Name: Kendro Laboratory Products L.P. HERAcell 240
Indications For Use: The intended use of this incubator is to provide an environment with controlled temperature, CO2 elevated humidity, an automatic decontamination mode, and O2 at suppressed levels, for the development of ova or embryos at or near body temperature.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
L Prescription Use (Per 21 CFR 801.109)
OR
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Page 4 HERAcell 240 510(k)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.