OXYBOX SYSTEM

K023456 · Oxyfast Corporation · KPJ · Feb 24, 2003 · General, Plastic Surgery

Device Facts

Record IDK023456
Device NameOXYBOX SYSTEM
ApplicantOxyfast Corporation
Product CodeKPJ · General, Plastic Surgery
Decision DateFeb 24, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.5650
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The OxyBox System is intended to provide topical oxygen to treat 1) skin ulcerations due to diabetes, venous stasis, post surgical infections and gangrenous lesions, 2) decubitis ulcers, 3) amputations/infected stumps, 4) skin grafts, 5) burns, and 6) frostbite.

Device Story

OxyBox is a single-patient, disposable, battery-operated topical oxygen bandage system. It comprises an oxygen concentrator, sterile cannula, and wound dressing. The concentrator generates oxygen in situ, delivering 98-100% oxygen at ~3.0 ml/hour for seven days. The cannula directs oxygen beneath the bandage to the wound site. Used in clinical or home settings to treat various wounds; provides topical oxygen at ambient pressure (1 atmosphere) to promote healing. Benefits include increased epithelization compared to control dressings.

Clinical Evidence

Bench testing validated device specifications. Animal testing demonstrated that topical oxygen delivered by the OxyBox provides increased epithelization compared to a control.

Technological Characteristics

Single-use, battery-operated oxygen concentrator; delivers 98-100% oxygen at ~3.0 ml/hour. Ambient pressure delivery (1 atm). Components include sterile cannula and wound dressing. No software or complex algorithms described.

Indications for Use

Indicated for patients with skin ulcerations (diabetic, venous stasis, post-surgical, gangrenous), decubitus ulcers, amputations/infected stumps, skin grafts, burns, and frostbite.

Regulatory Classification

Identification

A topical oxygen chamber for extremities is a device that is intended to surround a patient's limb and apply humidified oxygen topically at a pressure slightly greater than atmospheric pressure to aid healing of chronic skin ulcers such as bedsores.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance: Topical Oxygen Chamber for Extremities.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # 510(k) Summary for FEB 2 4 2003 # OxyBox #### 1. SPONSOR Oxyfast Corporation 11000 Cedar Avenue, Suite 449 Cleveland, OH 44106 Contact Person: Melvyn Burk Telephone:216-795-3060 Date Prepared: January 6, 2003 #### 2. Device Name | Proprietary Name: | OxyBox | |----------------------|-----------------------------------------------| | Common/Usual Name: | Topical oxygen bandage | | Classification Name: | Topical oxygen chamber CFR878.5650, Class III | #### 3. PREDICATE DEVICES O2Boot™: K971507 #### 4. DEVICE DESCRIPTION OxyBox is a topical bandage system that incorporates a disposable oxygen It consists of three separate component parts: (1) the oxygen concentrator. concentrator, (2) the sterile cannula and (3) the bandage or wound dressing. Items (2) and (3) are medical grade commercial products that Oxyfast Corporation obtains from device manufacturers or distributors. The oxygen concentrator is a single patient, one time use, disposable, battery-operated device that is capable of delivering 98 to 100 percent oxygen (balance moisture) for seven days at a rate of ~3.0 ml/hour. The cannula conveys the oxygen from the oxygen concentrator to the area beneath the bandage overlying the wound. {2}------------------------------------------------ ## 5. Indications for Use The OxyBox System is intended to provide topical oxygen to treat 1) skin ulcerations due to diabetes, venous stasis, post surgical infections and gangrenous lesions, 2) decubitis ulcers, 3) amputations/infected stumps, 4) skin grafts, 5) burns, and 6) frostbite. ## TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6. The predicate devices all deliver oxygen to the wound site at very slightly enhanced pressure (1.04 atmosphere). OxyBox delivers oxygen to the wound at ambient pressure (typically 1 atmosphere). The device design is different in the sense that oxygen is generated in situ in the case of the OxyBox, and the predicate devices use an oxygen cylinder to deliver oxygen. The predicate devices as well as the OxyBox deliver near 100 percent pure oxygen to the wound site. Both the OxyBox and the predicate O2Boot™ are single use devices. #### 7. PERFORMANCE TESTING Testing was conducted to validate that the OxyBox performed according to its specifications. In addition, animal testing was conducted that demonstrated that topical oxygen delivered by the OxyBox provides increased epithelization compared to a control. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of three human profiles facing to the right, with flowing lines representing hair or movement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Oxyfast Corporation Melvyn Burk President University West Building, Suite 449 11000 Cedar Avenue Cleveland, Ohio 44106 - 3052 FEB 2 4 2003 Re: K023456 Trade/Device Name: OxyBox System Regulation Number: 878.5650 Regulation Name: Topical Oxygen Chamber Regulatory Class: Class III Product Code: KPJ Dated: October 14, 2002 Received: October 15, 2002 Dear Mr. Burk: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Aci include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in {4}------------------------------------------------ Page 2 - Mr. Melvyn Burk the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, -Miriaire C. Paroit (Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ KOZZ456 510(k) Number (if known): OxyBox System Device Name: Indications for Use: The OxyBox System is intended to provide topical oxygen to treat 1) skin ulcerations due to diabetes, venous stasis, post surgical infections and gangrenous lesions, 2) decubitis ulcers, 3) amputations/infected stumps, 4) skin grafts, 5) burns, and 6) frostbite. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) iriam C Provost Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number K023456 Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96) Oxyfast Corporation 510(k) OxyBox October 14, 2002
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...