K023382 · Cooper Surgical · MQF · Dec 20, 2002 · Obstetrics/Gynecology
Device Facts
Record ID
K023382
Device Name
COOPERSURGICAL MALLEABLE SYTLET
Applicant
Cooper Surgical
Product Code
MQF · Obstetrics/Gynecology
Decision Date
Dec 20, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6110
Device Class
Class 2
Indications for Use
The sterile singe-use devices are to be used in conjunction with an Edwards Catheter for in-vitro fertilization procedures where passage through the cervix is impeded.
Device Story
Single-use sterile malleable stylet; used with assisted reproduction catheters to facilitate cervical passage during in-vitro fertilization. Device consists of stainless steel inner rod encapsulated in Teflon (FEP) and detachable rigid outer sheath; connected via male/female Luer Lock. Outer sheath features 1cm distal graduations. Used in clinical settings by physicians; provides mechanical support to catheter to navigate impeded cervix. Benefits patient by enabling successful catheter placement for IVF procedures.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-1 and mechanical testing for ISO 594-1 (6% luer taper) compliance.
Technological Characteristics
Materials: Stainless steel inner rod, Teflon (FEP) encapsulation, Polypropylene Luers, Polyethylene tip protector. Dimensions: 18cm or 23cm length. Energy: None (mechanical). Sterilization: Sterile, single-use. Standards: ISO 10993-1 (biocompatibility), ISO 594-1 (conical fittings).
Indications for Use
Indicated for patients undergoing in-vitro fertilization procedures where passage through the cervix is impeded, requiring use of an assisted reproduction catheter.
Regulatory Classification
Identification
Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
{1}------------------------------------------------
DEC 2 0 2002
FDA 510(k) Premarket Notification CooperSurgical Malleable Stylet
KO23382
# 510 (k) Summary
Date Prepared [21 CFR 807.92(a)(1)]
October 7, 2002
# Submitter's Information [21 CFR 807.92(a)(1)]
Joseph M. Azary C/o CooperSurgical, Inc. P.O. Box 2156 Huntington, CT. 06484
Azary Technologies has received authorization to submit this 510(k) on behalf of the sponsor CooperSurgical, Inc., 95 Corporate Drive, Trumbull, CT 06611.
# Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
Possible device trade names are: CooperSurgical Malleable Stylet or Edwards Malleable Stylet Common Name: Assisted Reproduction Catheters, Stylet
# Predicate Device [21 CFR 807.92(a)(3)]
Wallace Malleable Stylets - K990349
The subject devices have the same indications for use, material composition, sterilization method, and working dimensions as the predicate device. A minor difference lies in the luers of the subject device are slightly larger (both in diameter and length) than the predicate device. The subject device is packaged in a pouch composed of Tyvek and film, whereas the predicate device is packaged in a pouch composed of paper and film.
# Description of the Device [21 CFR 807.92(a)(4)]
Assisted Reproduction Catheters / Stylets are single-use sterile devices provided for use with CooperSurgical Assisted Reproduction Catheters, should passage through the cervix be impeded. The devices have an overall length of 23cm and 18cm and consist of a malleable inner stylet and a detachable rigid outer sheath, which is attached to the flexible inner stylet by a male/female Luer Lock adaptor. When the outer sheath Luer is attached to the hub of the stylet, the inner malleable stylet forms a smooth radius at the outer sheath. The outer sheath has a series of 1cm graduations at the distal end. The stainless steel stylet is encapsulated in Teflon (FEP).
The subject devices will be packaged in a flexible pouch composed of Tyvek heat sealed to polyethylene film. The pouch is designed to be peeled open. The pouches will be placed in a carton will contain 10 units.
The two versions are AR-ST18 (18cm length) and AR-ST23 (23cm length).
{2}------------------------------------------------
| Component | Material | Details |
|---------------------|-------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Stylet Outer Sheath | Teflon | Fluortek FEP-20 White (aka<br>Neoflon NP-20) |
| Stylet Inner Rod | Stainless Steel | No patient contact. The 316<br>Stainless Steel is encapsulated in<br>Teflon (FEP). |
| Outer Sheath | Teflon | Fluortek FEP-20 White (aka<br>Neoflon NP-20) |
| Luers | Polypropylene (pink colorant<br>used in outer luer and blue color-<br>ant used in stylet luer). | Montell Profax 6323 Compounded<br>by Chroma with PMS 240 U Pink.<br><br>Montell Profax 6323 Compounded<br>by Chroma with PMS 300C Blue. |
| Tip Protector | Polyethylene | No patient contact |
The subject devices are composed of the following materials:
# Intended Use (21 CFR 807.92(a)(5)]
The sterile singe-use devices are to be used in conjunction with an Edwards Catheter for in-vitro fertilization procedures where passage through the cervix is impeded.
#### Technological Characteristics [21 CFR 807.92(a)(6)]
CooperSurgical, Inc. believes that the subject device is substantially equivalent to the predicate device. The subject device is composed of the same materials, sterilized using the same method, complies with the same standards, has the same indications for use, and has the same working dimensions. The minor differences are with the packaging and the outer diameter and length of the Luer.
#### Performance Data [21 CFR 807.92(b)(1)]
The subject device has been subject to biocompatibility testing (for the materials that is equivalent to ISO 10993-1 Biocompatibility requirements. The subject device also complies with ISO 594-1 1986 Conical fittings with a 6% (luer) taper requirements.
#### Conclusion [21 CFR 807.92(b)(3)]
We believe the changes are minor and conclude that the subject devices are as safe and effective as the predicate devices.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three lines extending from its head, representing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service .
nd Drug Administration 0 Corporate Boulevard ckville MD 20850
CooperSurgical, Inc. % Mr. Joseph M. Azary Azary Technologies, LLC P.O. Box 2156 HUNTINGTON CT 06484
### Re: K023382
Trade/Device Name: CooperSurgical, Inc. Malleable Stylet Regulation Number: 21 CFR 884.6110 Regulation Name: Assisted reproduction catheters Regulatory Class: II Product Code: 85 MQF Dated: December 3, 2002 Received: December 4, 2002
# Dear Mr. Azary:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx. | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Henry C. Snogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
FDA 510(k) Premarket Notification CooperSurgical Malleable Stylet
5.10(k) Number (if known): K023382
Device Name: CooperSurgical, Inc. Malleable Stylet
Indications For Use: The sterile singe-use devices are to be used in conjunction with CooperSurgical Assisted Reproduction Catheters for in-vitro fertilization procedures where passage through the cervix is impeded. : · : . :
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
17
(Division Sign-Off) Division of Reproductive, Al and Radiological Device 510(k) Number _
Page 4 510(k) Malleable Stylet
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.