← Product Code [CBK](/productcode/CBK) · K023350

# ESPRIT VENTILATOR WITH RESPIRATORY MECHANICS (K023350)

_Respironics California, Inc. · CBK · Nov 6, 2002 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K023350

## Device Facts

- **Applicant:** Respironics California, Inc.
- **Product Code:** [CBK](/productcode/CBK.md)
- **Decision Date:** Nov 6, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5895
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Esprit ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support to adult and pediatric patients as prescribed by a physician. The Esprit Ventilator is intended for use in either invasive or non-invasive applications. Esprit is not intended for use in the presence of flammable anesthetics. Esprit is a prescription use device that is intended for sale by or on the order of a physician.

## Device Story

Esprit Ventilator with Respiratory Mechanics is a microprocessor-controlled, electrically powered mechanical ventilator. It provides ventilatory support and diagnostic monitoring of lung parameters. Clinicians use the device in clinical settings to perform maneuvers including Vital Capacity, Maximum Inspiratory Pressure, P0.1 (pressure change after 100ms of inspiratory effort), and static compliance/resistance measurements. The device calculates additional metrics such as Ti/Ttot, peak lung flow, and dynamic compliance/resistance. Data is displayed on the user interface in graphical and text formats. These outputs assist clinicians in monitoring respiratory status and informing clinical decision-making regarding patient ventilation.

## Clinical Evidence

No clinical data. Software verification and validation testing were performed in accordance with FDA's 1998 Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Results confirmed that all design and system requirements were met.

## Technological Characteristics

Microprocessor-controlled, electrically powered mechanical ventilator. Features diagnostic spirometry capabilities. No new technological characteristics introduced with the Respiratory Mechanics feature. Software-based monitoring and calculation of respiratory parameters.

## Regulatory Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

## Predicate Devices

- Puritan-Bennett 7200a Ventilator ([K902506](/device/K902506.md))
- Puritan-Bennett 840 Ventilator ([K970460](/device/K970460.md))
- Infrasonics Adult Star Ventilator ([K905272](/device/K905272.md))
- Dräger Evita 4 Ventilator ([K961687](/device/K961687.md))
- Respironics BiPAP Vision ([K982454](/device/K982454.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K023350

# 6 510(k) SUMMARY

| Company<br>Information: | Respironics, California Inc.<br>2271 Cosmos Court<br>Carlsbad, CA. 92009                               |         |
|-------------------------|--------------------------------------------------------------------------------------------------------|---------|
| Contact:                | Mary Funk<br>Regulatory Affairs Project Manager                                                        |         |
|                         | Phone Number: (760) 918-7328<br>Fax Number: (760) 918-0169                                             |         |
| Date Prepared:          | November 5, 2002                                                                                       |         |
| Product Name:           | Esprit Ventilator with Respiratory Mechanics                                                           |         |
| Classification:         | Class II<br>Continuous Ventilator (per 21 CFR 868.5895)<br>Diagnostic Spirometer (per 21 CFR 868.1840) |         |
| Predicate<br>Devices:   | Puritan-Bennett 7200a Ventilator                                                                       | K902506 |
|                         | Puritan-Bennett 840 Ventilator                                                                         | K970460 |
|                         | Infrasonics Adult Star Ventilator                                                                      | K905272 |
|                         | Dräger Evita 4 Ventilator                                                                              | K961687 |
|                         | Respironics BiPAP Vision                                                                               | K982454 |

### Description of Device Modification:

Esprit Respiratory Mechanics is a diagnostic tool that enables a clinician to take dynamic and static measurements of a patient's lung parameters. The following maneuvers can be performed using Esprit Respiratory Mechanics:

- Vital Capacity .
- . Maximum Inspiratory Pressure
- PO.1 (Pressure change after the first 100 milliseconds of inspiratory effort) .
- Static Compliance and Resistance .

These tools are used by clinicians to monitor a patient's respiratory status and are displayed on the Esprit User Interface both graphically and in text format. Respiratory Mechanics also calculates the following:

- . Ti/Ttot
- Peak Lung Flow
- Dynamic Compliance and Resistance .

{1}------------------------------------------------

#### Intended Use:

The Esprit ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support to adult and pediatric patients as prescribed by a physician. The Esprit Ventilator is intended for use in either invasive or non-invasive applications.

Esprit is not intended for use in the presence of flammable anesthetics. Esprit is a prescription use device that is intended for sale by or on the order of a physician.

#### Technological Characteristics:

The Esprit ventilator does not incorporate any new technological characteristics with the addition of Respiratory Mechanics.

#### Determination of Substantial Equivalence:

The Esprit Ventilator with Respiratory Mechanics has similar performance characteristics, the same intended use, the same environment of use and patient populations as the currently marketed predicate devices. The labeling and instructional information, including warning and caution statements, is similar to that of the predicate devices. The addition of this new feature does not raise new questions of safety or effectiveness for the Esprit.

#### Summary of Performance Testing:

Software validation testing was performed in accordance with FDA's Guidance for the Content of Premarket Submissions for Software contained in Medical Devices (1998). The results of all verification and validation testing demonstrate that all design and system requirements for the Esprit ventilator with Respiratory Mechanics have been met.

#### Conclusion:

The technological characteristics of the Esprit ventilator with Respiratory Mechanics and the results of the performance testing do not raise new questions of safety and effectiveness when compared to the legally marketed predicate devices.

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Mary Funk Regulatory Affairs Project Manager Respironics California, Incorporated Hospital Division 2271 Cosmos Court Carlsbad, California 92009

Re: K023350

Trade/Device Name: Esprit Ventilator with Respiratory Mechanics Regulation Number: 21 CFR 868.5895 and 21 CFR 868.1840 Regulation Name: Continuous Ventilator and Diagnostic Spirometer Regulatory Class: II Product Code: CBK and BZG Dated: October 4, 2002 Received: October 7, 2002

Dear Ms. Funk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{3}------------------------------------------------

Page 2 - Ms. Funk

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice > requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646 . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Timothy A. Ulatowski

Timot Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Special 510(k): Device Modification Esprit Ventilator with Respiratory Mechanics

.

11 - 11 - 11

.

.

.

#### STATEMENT OF INDICATIONS FOR USE 8

| Applicant:           | Respironics California, Inc.                                                                                                                                                                                                                                                                                                                                                   |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number:       | K023350                                                                                                                                                                                                                                                                                                                                                                        |
| Device Name:         | Esprit Ventilator with Respiratory Mechanics                                                                                                                                                                                                                                                                                                                                   |
| Indications for use: | The Esprit ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult and pediatric patients as prescribed by a physician. The Esprit Ventilator is intended for use in either invasive or non-invasive applications. |
| Prescription Use:    | Yes (Per 21 CFR 801.109)                                                                                                                                                                                                                                                                                                                                                       |

## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE) (Per 21 CFR 801.109) (Optional Format 1-2-96)

(

v. General Hospital,

510(k) Number

Prescription Use √

or

OTC Use

---

**Source:** [https://fda.innolitics.com/device/K023350](https://fda.innolitics.com/device/K023350)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
