← Product Code [FTL](/productcode/FTL) · K023328

# REINFORCED ABSORBABLE SYNTHETIC SURGICAL MESH (K023328)

_Center For Biomaterials and Advanced Technologies · FTL · Dec 12, 2002 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K023328

## Device Facts

- **Applicant:** Center For Biomaterials and Advanced Technologies
- **Product Code:** [FTL](/productcode/FTL.md)
- **Decision Date:** Dec 12, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Reinforced Absorbable Synthetic Surgical Mesh is indicated for use in the surgical repair of damaged or ruptured soft tissue where reinforcing or bridging materials are needed. This device is intended for one-time use.

## Device Story

Reinforced Absorbable Synthetic Surgical Mesh consists of a woven polydioxanone mesh encased in a biocompatible, bioabsorbable elastomeric foam. The device is used by surgeons in clinical settings to provide wound support and facilitate tissue bridging during the healing process. The foam component provides porosity and wound support, while the polydioxanone mesh provides structural strength to prevent wound dehiscence. Both components are designed to be non-antigenic and non-pyrogenic, eliciting a typical tissue reaction during absorption. The device is intended for one-time use in soft tissue repair.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological comparison and existing biocompatibility data for the predicate devices. No additional biocompatibility or animal testing was performed.

## Technological Characteristics

Device is a composite of woven polydioxanone mesh and a biocompatible, bioabsorbable elastomeric foam. Materials are non-antigenic and non-pyrogenic. The device functions as a surgical implant for soft tissue reinforcement and bridging.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- PDS™ (polydioxanone) Mesh ([K844228](/device/K844228.md))
- AFB (Absorbable Foam Buttress) ([K014183](/device/K014183.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

I.

K023328

## SUMMARY OF SAFETY AND EFFECTIVENESS

### DEC 1 2 2002 GENERAL INFORMATION Center for Biomaterials and Advanced A. Submitted By: Technologies, Medical Devices Group, a division of Ethicon, Inc. Route 22 West, P.O. Box 151 Somerville, NJ 08876 (908) 218-2041 Tel: (908) 218-3679 Fax: Contact Person: David A. Dalessandro At address above Reinforced Absorbable Synthetic Surgical B. Device Trade Name: Mesh Surgical Mesh Common Name: Classification Name: Mesh, Surgical, Polymeric FTL, 21 CFR 878.3300 C. Predicate Device(s): Manufacturer Product Name 510(k) No. PDSTM (polydioxanone) Mesh Ethicon, Inc. K844228

J&J Corporate AFB (Absorbable Foam K014183 Biomaterials Center Buttress)

#### D. Device Description:

The reinforced absorbable synthetic surgical mesh is a foam encased mesh. The foam component is a biocompatible elastomeric polymer that is bioabsorbable. This elastomer exhibits a high percent elongation, while possessing good tensile strength and good recovery characteristics. It has been found to be non-antigenic, non-pyrogenic, and elicits a slight, but typical, tissue reaction during absorption. Therapeutically, the foam component provides wound support as well as increases the opportunity for bridging to occur due to its porosity. The mesh component is a woven polydioxanone mesh that is used to provide wound support during the healing process. It has been found to be non-antigenic, non-pyrogenic, and also elicits a slight, but typical, tissue reaction during absorption. Therapeutically, the mesh component retains strength for a period of time long enough to permit sufficient healing, preventing wound dehiscence.

{1}------------------------------------------------

Accordingly, no or minimal undesirable side effects occur during the healing process.

- Indications for Use: E.
Reinforced Absorbable Synthetic Surgical Mesh is indicated for use in the surgical repair of damaged or ruptured soft tissue where reinforcing or bridging materials are needed. This device is intended for one-time use.

- F. Technological Comparison:
The reinforced absorbable synthetic surgical mesh and PDSTM Mesh (K844228) have similar indications for use and overall function and perform in a similar manner with respect to reinforcement of soft tissue wound strength. The reinforced absorbable synthetic surgical mesh has similar absorption, degradation, and burst strength to the predicate device.

The foam component of the reinforced absorbable synthetic surgical mesh functions and performs in the same manner as AFB (Absorbable Foam Buttress) (K014183).

#### II. TESTING

The biocompatibility testing for AFB and PDS™ Mesh are sufficient and applicable to the reinforced absorbable synthetic surgical mesh. Accordingly, no additional biocompatibility or animal testing on the reinforced absorbable synthetic surgical mesh was performed.

#### III. CONCLUSIONS

In summary, Center for Biomaterials and Advanced Technologies, Medical Devices Group, a division of Ethicon, Inc. has demonstrated that the intended use for the reinforced absorbable synthetic surgical mesh is the same as the original devices. The technological characteristics have been described in sufficient detail to demonstrate that they are the same as the original devices. Therefore, this premarket notification has demonstrated substantial equivalence as defined and understood in the Federal Food, Drug, & Cosmetic Act and its amendments.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes extending from its head, representing the department's mission to promote health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is simple and recognizable, conveying the department's authority and purpose.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Center for Biomaterials and Advanced Technologies Melissa Mahall c/o Bio-Reg Associates 11800 Baltimore Avenue, Suite 105

Beltsville. Maryland 20705

## Re: K023328

Trade/Device Name: Reinforced Absorbable Synthetic Surgical Mesh Regulation Number: 878.3300 Regulation Name: Polymeric Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: October 2, 2002 Received: October 4, 2002

Dear Ms. Mahall:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the

{3}------------------------------------------------

Page 2 – Ms. Melissa Mahall

quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

Celia M. Witten, Ph.D., MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

KOZ3328

Device Name:

Reinforced Absorbable Synthetic Surgical Mesh

Sponsor Name:

Center for Biomaterials and Advanced Technologies, Medical Devices Group, a division of Ethicon, Inc.

Indications for Use:

Reinforced Absorbable Synthetic Surgical Mesh is indicated for use in the surgical repair of damaged or ruptured soft tissue where reinforcing or bridging materials are needed. This device is intended for one-time use.

Do Not Write Below This Line - Continue on Another Page if Needed

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use Over-The-Counter Use

风 

Miriam C. Provost

Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K023328

---

**Source:** [https://fda.innolitics.com/device/K023328](https://fda.innolitics.com/device/K023328)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com