IMMULITE/IMMULITE 1000 CALCITONIN, MODEL LKCL AND IMMULITE 2000 CALCITONIN, MODEL L2KCL
Applicant
Diagnostic Products Corp.
Product Code
JKR · Clinical Chemistry
Decision Date
Dec 9, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1140
Device Class
Class 2
Indications for Use
IMMULITE/IMMULITE 1000 Calcitonin For in vitro diagnostic use with the IMMULITE/IMMULITE 1000 Analyzer - for the quantitativ measurement of calcitonin (thyrocalcitonin) in human serum, as an aid in the diagnosis and treatment of diseases involving the thyroid and parathyroid glands, including carcinoma and hyperparathyroidism IMMULITE 2000 Calcitonin For in vitro diagnostic use with the IMMULITE 2000 Analyzer - for the quantitative measurement of calcitonin (thyrocalcitonin) in human serum, as an aid in the diagnosis and treatment of diseases involving the thyroid and parathyroid glands, including carcinoma and hyperparathyroidism.
Device Story
IMMULITE/IMMULITE 1000 and IMMULITE 2000 Calcitonin are automated, solid-phase, chemiluminescent enzyme immunoassays. Input: human serum sample. Process: sample incubated with alkaline phosphatase-conjugated polyclonal goat anti-calcitonin in a test unit/reaction tube containing a polystyrene bead coated with monoclonal murine anti-calcitonin; formation of antibody-sandwich complex; centrifugal wash removes unbound conjugate; addition of phosphate ester of adamantyl dioxetane substrate; hydrolysis by alkaline phosphatase produces unstable intermediate emitting light. Output: photon output measured by system luminometer, proportional to calcitonin concentration. Used in clinical laboratories; operated by trained laboratory personnel. Results used by physicians to aid diagnosis and management of thyroid/parathyroid conditions.
Clinical Evidence
Method comparison study performed on 53 patient samples for IMMULITE/IMMULITE 1000 and 67 patient samples for IMMULITE 2000, comparing results against the Nichols Advantage Calcitonin assay. Linear regression analysis for IMMULITE/IMMULITE 1000 yielded r=0.990; IMMULITE 2000 yielded r=0.982. No clinical prospective or retrospective diagnostic accuracy studies provided.
Technological Characteristics
Solid-phase, enzyme-labeled, two-site chemiluminescent immunometric assay. Solid-phase: polystyrene bead coated with monoclonal murine anti-calcitonin. Detection: alkaline phosphatase-conjugated polyclonal goat anti-calcitonin. Substrate: phosphate ester of adamantyl dioxetane. Energy source: automated luminometer. Connectivity: integrated with IMMULITE/IMMULITE 1000 or IMMULITE 2000 automated analyzers.
Indications for Use
Indicated for quantitative measurement of calcitonin in human serum to aid diagnosis and treatment of thyroid and parathyroid gland diseases, including carcinoma and hyperparathyroidism.
Regulatory Classification
Identification
A calcitonin test system is a device intended to measure the thyroid hormone calcitonin (thyrocalcitonin) levels in plasma and serum. Calcitonin measurements are used in the diagnosis and treatment of diseases involving the thyroid and parathyroid glands, including carcinoma and hyperparathyroidism (excessive activity of the parathyroid gland).
{1}------------------------------------------------
IC23304
# DEC 0 9 2002
## 510 (k) Summary Safety and Effectiveness
This summary of safety and effectiveness information has been prepared in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.
| Name: | Diagnostic Products Corporation |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 5700 West 96th Street<br>Los Angeles, CA 90045 |
| Telephone Number: | (310) 645-8200 |
| Facsimile Number: | (310) 645-9999 |
| Contact Person: | Edward M. Levine, Ph.D.<br>Director of Clinical Affairs |
| Date of Preparation: | October 1, 2002 |
| Device Name: | |
| Trade: | IMMULITE®/IMMULITE® 1000 Calcitonin<br>IMMULITE® 2000 Calcitonin |
| Catalog Number: | LKCL<br>L2KCL |
| CFR: | A calcitonin test system is a device intended to measure the<br>thyroid hormone calcitonin (thyrocalcitonin) levels in<br>plasma and serum. Calcitonin measurements are used in<br>the diagnosis and treatment of diseases involving the<br>thyroid and parathyroid glands, including carcinoma and<br>hyperparathyroidism (excessive activity of the parathyroid<br>gland). |
| Common: | Reagent system for the determination of calcitonin in<br>serum. |
| Classification: | Class II device, JKR (21 CFR 862.1140) |
| Panel: | Clinical Chemistry |
| CLIA Complexity Category: | We believe the category to be moderate, based on previous<br>classification of analogous tests. |
| Manufacturer: | Diagnostic Products Corporation (DPC)<br>5700 West 96th Street<br>Los Angeles, CA 90045-5597 |
,
{2}------------------------------------------------
| Establishment<br>Registration #: | DPC's establishment Registration No. is 2017183 |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantially Equivalent<br>Predicate Device: | Nichols Advantage™ Calcitonin (K964635) |
| Description of Device: | IMMULITE/IMMULITE 1000 Calcitonin and IMMULITE<br>2000 Calcitonin are solid-phase, chemiluminescent enzyme<br>immunoassays for use with their respective<br>IMMULITE/IMMULITE 1000 and IMMULITE 2000<br>Automated Analyzers. |
| Intended Use of the<br>Device: | IMMULITE/IMMULITE 1000 Calcitonin and IMMULITE<br>2000 Calcitonin are for <i>in vitro</i> diagnostic use for the<br>quantitative measurement of calcitonin in human serum, as<br>an aid in the diagnosis and treatment of diseases involving<br>the thyroid and parathyroid glands, including carcinoma<br>and hyperparathyroidism. |
### Technology:
This section does not contain any new information for a reviewer who is familiar with the DPC IMMULITE System based upon the review of previous IMMULITE and IMMULITE 2000 assay submissions.
IMMULITE/IMMULITE 1000 Calcitonin is a solid-phase, enzyme-labeled, two-site chemiluminescent immunometric assay. The solid-phase, a polystyrene bead enclosed within an IMMULITE/IMMULITE 1000 Test Unit, is coated with a monoclonal murine antibody specific for calcitonin.
While the patient serum sample and alkaline phosphatase-conjugated to polyclonal goat anti-calcitonin are incubated for approximately 30 minutes at 37℃ in the Test Unit with intermittent agitation, calcitonin in the sample is bound to form an antibody sandwich complex. Unbound enzyme conjugate is then removed by a centrifygal wash, after which substrate is added and the Test Unit is incubated for a further 10 minutes.
The chemiluminescent substrate, a phosphate ester of adamantyl dioxetane, undergoes hydrolysis in the presence of alkaline phosphatase to vield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light, thus improving precision by providing a window for multiple readings. The bound complex and thus also the photon output, as measured by the luminometer - is proportional to the concentration of calcitonin in the sample.
{3}------------------------------------------------
IMMULITE 2000 Calcitonin is a solid-phase, enzyme-labeled, two-site chemiluminescent immunometric assay. The solid-phase, a polystyrene bead, is coated with a monoclonal murine antibody specific for calcitonin.
While the patient serum sample and alkaline phosphatase-conjugated to polyclonal goat anti-calcitonin are incubated for approximately 30 minutes at 37℃ in the Reaction Tube with intermittent agitation, calcitonin in the sample is bound to form an antibody sandwich complex. Unbound enzyme conjugate is then removed by a centrifugal wash, after which substrate is added and the Reaction Tube is incubated for a further 5 minutes.
The chemiluminescent substrate, a phosphate ester of adamantyl dioxetane, undergoes hydrolysis in the presence of alkaline phosphatase to yield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light, thus improving precision by providing a window for multiple readings. The bound complex and thus also the photon output, as measured by the luminometer - is proportional to the concentration of calcitonin in the sample.
Nichols Advantage Calcitonin assay is a two-site chemiluminescence immunoassay for the measurement of calcitonin in human serum. It utilizes two mouse monoclonal antibodies to human calcitonin. One of the mouse monoclonal antibodies is labeled with acridinium ester for detection, while a second mouse monoclonal antibody is coupled to biotin. Calcitonin is "sandwiched" between these antibodies.
The sample containing calcitonin is incubated simultaneously with both antibodies. The formation of a soluble sandwich complex occurs only in the presence of calcitonin molecules, which bridge the two antibodies. Therefore, only peptides that bridge these two antibodies can be quantitated.
After an initial incubation period, steptavidin coated magnetic particles are added to the reaction mixture and a second incubation follows. This allows for a highly specific and efficient means of binding the sandwich complex to the solid phase via the high affinity interaction between biotin and streptavidin. Free labeled antibody is separated from the labeled antibody bound to the magnetic particles by aspiration of the reaction mixture and subsequent washing. The wells containing the washed magnetic particles are transported into the system luminometer, which automatically injects Trigger 2, initiating the chemiluminescence reaction. The light is quantitated by the luminometer and expressed as RLU. The amount of bound labeled antibody is directly proportional to the concentration of calcitonin in the sample.
{4}------------------------------------------------
### Performance Equivalence:
Diagnostic Products Corporation asserts that the IMMULITE/IMMULITE 1000 Calcitonin and IMMULITE 2000 Calcitonin produce substantially equivalent results to other commercially marketed Calcitonin assays, such as Nichols Advantage™ Calcitonin used with the Nichols Advantage™ Specialty System. Each product is designed for the quantitative measurement of calcitonin in human serum. Each product is intended strictly for in vitro diagnostic use as an aid in the diagnosis and treatment of diseases involving the thyroid and parathyroid glands, including carcinoma and hyperparathyroidism.
### Method Comparison
The IMMULITE/IMMULITE 1000 Calcitonin procedure was compared to a commercially available assay (Nichols Advantage) on 53 patient samples, with calcitonin concentrations ranging up to approximately 1,200 pg/mL. Linear regression analysis vielded the following statistics.
| | (IMMULITE/IMMULITE 1000) = 0.71 (Nichols) + 4.2 pg/mL | r = 0.990 |
|--|-------------------------------------------------------|-----------|
| | Mean = 155 pg/mL (IMMULITE/IMMULITE 1000) | |
155 pg/mL (IMMULITE/IMMULITE 1000) Means: 212 pg/mL (Nichols)
The IMMULITE 2000 Calcitonin procedure was compared to a commercially available assay (Nichols Advantage) on 67 patient samples, with calcitonin concentrations ranging up to approximately 1,500 pg/mL. Linear regression analysis yielded the following statistics.
(IMMULITE 2000) = 0.81 (Nichols) - 0.4 pg/mL r = 0.982
Means: 193 pg/mL (IMMULITE 2000) 239 pg/mL (Nichols)
#### Conclusion:
The data presented in this summary of safety and effectiveness is the data that the Food and Drug Administration used in granting DPC substantial equivalence for IMMULITE® IMMULITE® 1000 Calcitonin and IMMULITE® 2000 Calcitonin.
{5}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The seal is black and white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# DEC 0 9 2007
Edward M. Levine, Ph.D. Director of Clinical Affairs Diagnostics Products Corporation 5700 West 96th Street Los Angeles, CA 90045-5597
Re: k023304
.
Trade/Device Name: IMMULITE®/ IMMULITE® 1000 Calcitonin IMMULITE® 2000 Calcitonin Regulation Number: 21 CFR 862.1140 Regulation Name: Calcitonin test system Regulatory Class: Class II Product Code: JKR; JJY Dated: October 1, 2002 Received: October 3, 2002
Dear Dr. Levine
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{6}------------------------------------------------
Page 2 –
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# 510(k) Number (if known): Device Name: IMMULITE®/IMMULITE® 1000 Calcitonin IMMULITE® 2000 Calcitonin
Indications For Use:
### IMMULITE/IMMULITE 1000 Calcitonin
For in vitro diagnostic use with the IMMULITE/IMMULITE 1000 Analyzer - for the quantitativ measurement of calcitonin (thyrocalcitonin) in human serum, as an aid in the diagnosis and treatment of diseases involving the thyroid and parathyroid glands, including carcinoma and hyperparathyroidism
#### IMMULITE 2000 Calcitonin
For in vitro diagnostic use with the IMMULITE 2000 Analyzer - for the quantitative measurement of calcitonin (thyrocalcitonin) in human serum, as an aid in the diagnosis and treatment of diseases involving the thyroid and parathyroid glands, including carcinoma and hyperparathyroidism.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
22 n 002
(Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.