K023272 · The Olympus Optical Co. · FCG · Dec 23, 2002 · Gastroenterology, Urology
Device Facts
Record ID
K023272
Device Name
DISPOSABLE ASPIRATION NEEDLE NA-200H
Applicant
The Olympus Optical Co.
Product Code
FCG · Gastroenterology, Urology
Decision Date
Dec 23, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
This instrument has been designed to be used with the ultrasound endoscope for ultrasonically quided fine needle aspiration (FNA) of submucosal and extramural lesions within the gastrointestinal tract (i.e. pancreatic masses, mediastinal masses, perirectal masses and lymph nodes).
Device Story
Disposable aspiration needle NA-200H used with ultrasound endoscope for fine needle aspiration (FNA) of GI tract lesions; includes pre-assembled sterile needle, VACLOK syringe, and stopcock. Device facilitates tissue sampling of submucosal and extramural masses (pancreatic, mediastinal, perirectal, lymph nodes). Operated by clinicians during endoscopic procedures. Syringe/stopcock assembly allows for controlled aspiration. Pre-assembled, single-use design improves procedural convenience and sterility compared to previous reusable or non-pre-assembled models. Output is tissue sample for diagnostic evaluation.
Clinical Evidence
No clinical data provided; substantial equivalence is based on non-clinical testing and design comparison to predicate devices.
Technological Characteristics
Disposable aspiration needle; includes VACLOK syringe and stopcock. Sterile, single-use, pre-assembled. Mechanical biopsy instrument. No software or electronic components.
Indications for Use
Indicated for ultrasonically guided fine needle aspiration (FNA) of submucosal and extramural lesions within the gastrointestinal tract, including pancreatic, mediastinal, and perirectal masses, and lymph nodes.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
DEC 2 3 2002
K023272
Ogz1og2
# SMDA 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR Section 807.92.
# A. Submitter's Name, Address, Phone and Fax Number
| 1. Applicant: | Olympus Optical Co., Ltd.<br>2-3-1 Shinjuku Monolis Nishi-shinjuku<br>Shinjuku-ku, Tokyo, Japan, 163-0914 |
|-------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration No. | 8010047 |
| Address, Phone and Fax<br>Numbers of R&D<br>Department Endoscope<br>Division: | 2951 Ishikawa-Cho,<br>Hachioji-shi, Tokyo 192-8507 Japan<br>Tel. 426-42-2891<br>Fax 426-42-2291 |
| 2. Initial Importer | Olympus America Inc.<br>Two Corporate Center Drive<br>Melville, NY 11747-3157 |
| Registration No. | 2429304 |
| 3. Contact Person | Laura Storms-Tyler<br>Director, Regulatory Affairs<br>Olympus America Inc.<br>Two Corporate Center Drive<br>Melville, NY 11747-3157<br>Tel (631)-844-5688<br>Fax (631)-844-5416 |
#### B. Device Name, Common Name
- Trade/Proprietary Name and Common Name 1. Trade Name: DISPOSABLE ASPIRATION NEEDLE NA-200H ASPIRATION NEEDLE Common Name:
- 2. Class, Classification Number and Classification Name
| Classification Number | Classification Name | Class |
|-----------------------|--------------------------------------------|-------|
| 21 CFR 876.1075 | Gastroenterology-urology biopsy instrument | II |
# 3. Identification of Legally Marketed Devices Which We Claim Substantial Equivalence
| Description | Model No | Manufacturer | #K |
|-----------------------------|----------|---------------------------|---------|
| ASPIRATION NEEDLE | NA-10J-1 | Olympus Optical Co., Ltd. | K973128 |
| VACLOK Syringe | - | Merit Medical System Inc. | K994253 |
| STOPCOCK(One-port manifold) | - | Merit Medical System Inc. | K934123 |
{1}------------------------------------------------
Kor 3272
pg 2 of r
## C. Description of the Device(s)
#### NA-200H
DISPOSABLE ASPIRATION NEEDLE NA-200H is a modified version of NA-10J-1 Aspiration Needle, which was cleared #K973128. When compared with the predicate device, the following modifications have been made to the NA-10J-1 needle:
- 1. The NA-200H is provided pre-assembled, sterile single use disposable.
- 2. The NA-200H includes a sterile single use disposable syringe for aspiration.
The comparison table provided shows all other similarities and differences to the legally marketed device. Please refer to Attachment 5 for details.
#### VACLOK Syringe/ STOPCOCK
The K12-MS2666 Syringe (VACLOK Syringe and Stopcock) is attached to the aspiration port on the handle section of the NA-200H. These devices will be provided sterile single use. We are offering these devices together to improve convenience to the customer. The syringe to be included in the package has been cleared by FDA as a cardiologist or radiologist during angiographic or radiologic, it is intended to be used for aspiration in this submission.
## D. Intended Use of the Device(s)
This instrument has been designed to be used with the ultrasound endoscope for ultrasonically quided fine needle aspiration (FNA) of submucosal and extramural lesions within the gastrointestinal tract (i.e. pancreatic masses, mediastinal masses, perirectal masses and lymph nodes).
## E. Summary including Conclusions drawn from Non-clinical Tests
When compared to the predicate device, Disposable Aspiration Needle NA-200H, VackLock Syringe and Stopcock does not incorporate any significant changes in the intended use, method of operation, material, or design that could affect the safety or effectiveness.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the caduceus.
#### Public Health Service
Food and Drug Administratio 200 Corporate Boulevard Rockville MD 20850
Ms. Laura Storms-Tyler Director, Regulatory Affairs Olympus America, Inc. Two Corporate Center Drive MELVILLE NY 11747-3157
#### Re: K023272
Trade/Device Name: Disposable Aspiration Needle NA-200H Regulation Number: 21 CFR §876.1076 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: II Product Code: 78 FCG Dated: September 30, 2002 Received: October 1, 2002
Dear Ms. Storms-Tyler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy Chogden
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{4}------------------------------------------------
510(k) Number(if known):
02.327.2
Device Name: DISPOSABLE ASPIRATION NEEDLE NA-200H
Indications for Use:
This instrument has been designed to be used with the ultrasound endoscope for ultrasonically guided fine needle aspiration (FNA) of submucosal and extramural lesions within the gastrointestinal tract (i.e. pancreatic masses, mediastinal masses, perirectal masses and lymph nodes).
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
David G. Ingram
(Division Sign-Off) Division of Reproductive, and Radiological Devices 510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.