INHEALTH SOFT SLEEVE COLONOSCOPE SPLINT, MODEL CS1200

K023259 · Helix Medical, Inc. · FDF · Dec 26, 2002 · Gastroenterology, Urology

Device Facts

Record IDK023259
Device NameINHEALTH SOFT SLEEVE COLONOSCOPE SPLINT, MODEL CS1200
ApplicantHelix Medical, Inc.
Product CodeFDF · Gastroenterology, Urology
Decision DateDec 26, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The InHealth Soft Sleeve Colonoscope Splint is indicated for use with a colonoscope to prevent the reformation of the sigmoid loop subsequent to the reduction of the curvature of the sigmoid colon by the colonoscope during the colonoscopy procedure. The InHealth Soft Sleeve Colonoscope Splint is intended to provide for easy advancement of the scope while avoiding damage to the colon.

Device Story

InHealth Soft Sleeve Colonoscope Splint is a disposable, non-sterile silicone overtube used during colonoscopy. Device features a soft, flexible, tapered distal tip and a proximal grip with an integrated injection port. During procedure, clinician inserts colonoscope through splint; water is injected into port to activate a lubricious, hydrophilic coating on the splint surface, facilitating smooth passage through the sigmoid colon. Distal tip design conforms to the colonoscope to minimize the gap between scope and splint. Proximal grip prevents the device from entering the rectum. Device prevents sigmoid loop reformation, aiding scope advancement and reducing potential for colon trauma. Intended for single-use by physicians in clinical settings.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Silicone tube with soft elastic distal tip and proximal grip. Features water-activated hydrophilic coating. Non-sterile, single-use disposable. Mechanical device; no energy source or software.

Indications for Use

Indicated for use with a colonoscope in patients undergoing colonoscopy to prevent sigmoid loop reformation after curvature reduction and facilitate scope advancement while minimizing mucosal damage.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K023259 Pageing3 ## Helix Medical, Inc. 510(k) Summary InHealth Soft Sleeve Colonoscope Splint DEC 2 6 2002 #### 1. NAME OF SUBMITTER InHealth Technologies, a business unit of Helix Medical. Inc. 1110 Mark Ave. 93013 Carpinteria, Ca Establishment Registration Number: 2025182 #### DEVICE NAME AND CLASSIFICATION ll. Proprietary Name: InHealth Soft Sleeve Colonoscope Splint Common or Usual Name: Colonoscope Splint Class II: 21 CFR 876.1500 The Soft Sleeve Colonoscope Splint is classified as 'Endoscope and Accessories'. The classification reads "An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eve and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices." The product code for this device is FDF, Colonoscope, Gastro-Urology. #### III. PREDICATE DEVICES - Olympus Splinting Tube, Olympus America, Inc., Melville, NY; K954451 . - Pentax Splinting Tube, Pentax Precision Inst., Orangeburg, NY; K# Unknown . - ACMI Splinting Tube, AMCI Corporation, Houston, TX; K# Unknown ● #### DESCRIPTION IV. The InHealth Soft Sleeve Colonoscope Splint is a disposable splint or overtube for use in colonoscopy to prevent the reformation of the sigmoid loop subsequent to the reduction of the curvature of the sigmoid colon by the colonoscope during the colonoscopy procedure. This device is a silicone tube with a soft distal tip and a proximal grip. The distal tip provides a soft, conforming and flexible tapered shape that hugs the scope and facilitates insertion. The entire splint, with the exception of {2}------------------------------------------------ *023259 Proj c 24f 3 the outer surface of the proximal grip, is coated with a lubricious, hydrophilic coating. This lubricious coating, when wetted, enables a smooth passage of the device along the colonoscope and of the splint through the sigmoid colon. The proximal grip is fitted with an injection port for injecting water into the inside of the splint tube to facilitate lubrication. The tip is grooved to allow flow of water from the inside diameter of the splint to the outside diameter to provide a controlled lubrication of the splint, the colonoscope, and the surrounding mucosa. The proximal grip, which is larger than the shaft, ensures that the entire tube does not pass through the anus and enter the rectum. The Soft Sleeve Splint is a non-sterile disposable device, intended for one-time use only, not intended for reuse. #### V. INTENDED USE The InHealth Soft Sleeve Colonoscope Splint is indicated for use with a colonoscope to prevent the reformation of the sigmoid loop subsequent to the reduction of the curvature of the sigmoid colon by the colonoscope during the colonoscopy procedure. The InHealth Soft Sleeve Splint is intended to provide for easy advancement of the scope while avoiding damage to the colon. ### TECHNOLOGICAL CHARACTERISTICS VI. The Soft Sleeve Splint is intended for the same use as its predicate devices, Olympus Splinting Tube. Pentax Splinting Tube, and the ACMI Splinting Tube. All these devices are indicated for use with a colonoscope to prevent the reformation of the sigmoid loop subsequent to the reduction of the curvature of the sigmoid colon by the colonoscope during the colonoscopy procedure. The Olympus Splinting Tube is offered as a non-sterile product and is cleaned and disinfected between uses. As with the predicate devices, the InHealth Soft Sleeve Splint is intended for use in a non-sterile intact body orifice, and therefore is also provided non-sterile. The InHealth Soft Sleeve Splint is however, provided as a disposable device and therefore does not require any special cleaning. There is no reuse of this device. The design, form, and materials of the Soft Sleeve Splint and its predicate devices are equivalent, in that all are designed to be a coated shaft through which the colonoscope is advanced during a colonoscopy. Helix Medical has introduced no new technology, or change in the intended use with the Soft Sleeve Splint. The Soft Sleeve Splint differs slightly in that it has a soft elastic distal tip rather than a rigid shaft. The three predicate devices have a rubber or plastic type coating to provide a smooth splint surface. The Olympus requires an application of a lubricant to the splint prior to use. The InHealth Soft Sleeve Colonoscope Splint has a water {3}------------------------------------------------ K02 3259 Page 3 of 3 activated lubricious hydrophilic surface that provides a smooth splint surface as well as enabling easier passage of the colonoscope through the device and the device through the colon. The proximal grip is fitted with an injection port for injecting water into the inside of the splint tube to facilitate lubrication. All four devices are designed for the colonoscope to pass through the splint. The three predicate devices are designed such that a gap remains between the colonoscope and the splint. The InHealth Soft Sleeve Colonoscope Splint is designed to provide a soft conforming shape that hugs the scope, thus minimizing the gap. Helix Medical considers the use of the Soft Sleeve Splint to be substantially equivalent to its predicate devices, the Olympus Splint, Pentax Splint, and the ACMI Splint. {4}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is an image of an eagle with three lines extending from its head. ## Public Health Service od and Drug Administration 200 Corporate Boulevard Rockville MD 20850 Ms. Cynthia M. Anderson Vice President of Regulatory Affairs and Quality Assurance Helix Medical, Inc. InHealth Technologies 1110 Mark Avenue CARPINTERIA CA 93013 ## Re: K023259 Trade/Device Name: InHealth Soft Sleeve Colonoscope Splint (Model CS1200) Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FDF Dated: September 30, 2002 Received: September 30, 2002 Dear Ms. Anderson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours. Nancy C Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ K02-325.9 ## INDICATIONS FOR USE Helix Medical, Inc. Applicant: 510(k) Number (if known): N/A* Device Name: InHealth Soft Sleeve Colonoscope Splint Indications For Use: The InHealth Soft Sleeve Colonoscope Splint is indicated for use with a colonoscope to prevent the reformation of the sigmoid loop subsequent to the reduction of the curvature of the sigmoid colon by the colonoscope during the colonoscopy procedure. The InHealth Soft Sleeve Colonoscope Splint is intended to provide for easy advancement of the scope while avoiding damage to the colon. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH Office of Device Evaluation (ODE) | (Division Sign-Off) | Nancy C. Brogdon | |---------------------------------------------------------------|------------------| | Division of Reproductive, Abdominal, and Radiological Devices | | | 510(k) Number | K023259 | | Prescription Use<br>Per 21 CFR 801.109 | OR | Over-the-Counter | |----------------------------------------|----|------------------| |----------------------------------------|----|------------------|
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...