K023200 · Sb Lucius, Inc. · EJT · Nov 4, 2002 · Dental
Device Facts
Record ID
K023200
Device Name
CERANUM 76
Applicant
Sb Lucius, Inc.
Product Code
EJT · Dental
Decision Date
Nov 4, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3060
Device Class
Class 2
Intended Use
This is a yellow ceramic alloy for single units, implants prosthetics, single & 3/4 crowns, and short span bridges.
Device Story
CERANUM 76 is a high noble, micro-fine, yellow, gold-based dental casting alloy. It is used by dental professionals for the fabrication of inlays, onlays, single crowns, bridges, implant superstructures, and as a substrate for medium expansion with lower fusing porcelains and indirect restorative composites. The alloy provides strength, durability, and the aesthetic color of gold to dental restorations. It is processed via standard dental casting techniques. The device is intended for clinical use in dental restorative procedures.
Clinical Evidence
Bench testing only. Physical and mechanical properties (melting point, Vickers hardness, yield strength, elongation, and coefficient of thermal expansion) were evaluated and compared to the predicate device. Testing followed ANSI/ADA 5 and ISO 9693 standards.
Technological Characteristics
High noble, gold-based alloy. Composition: 76% Au, 4% Pt, 10% Pd, 2.97% Ag, 2.97% Cu, 3.97% In. Tested per ANSI/ADA 5 and ISO 9693. Form factor: dental casting alloy for indirect restorations.
Indications for Use
Indicated for patients requiring dental restorations including single units, implant prosthetics, single and 3/4 crowns, and short span bridges.
Regulatory Classification
Identification
A noble metal alloy is a device composed primarily of noble metals, such as gold, palladium, platinum, or silver, that is intended for use in the fabrication of cast or porcelain-fused-to-metal crown and bridge restorations.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is FDA's “Class II Special Controls Guidance Document: Dental Noble Metal Alloys.” The devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. See § 872.1(e) for availability of guidance information.
{0}------------------------------------------------
K023200
#### 4 2002 NOV
## 510(K) SUMMARY
SB LUCIUS, INC. Submitter of 510(k): 9778 Katella Ave. Ste, 205, Anaheim, CA 92804 Phone: (714) 530-2814, Fax : (714) 530-3448,
| Contact person: | Dae-Kyu Chang |
|----------------------|-----------------------------------------------------------------|
| Phone: | (714) 530-2814 |
| Fax: | (714) 530-3448 |
| E-mail: | Gatorkr@yahoo.co.kr |
| Date of Summary: | Sept. 16, 2002 |
| Trade name: | CERANUM 76 |
| Common: | Dental casting alloy |
| Classification name: | Gold based alloys and precious metal alloys for<br>clinical use |
| Product code: | EJT |
| Classification: | Class II |
Legally marketed device: Metalor's V-Supra Plus 510(k) number: K993508
## SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS
Test methods applied: as in ANSI/ADA 5 and ISO 9693
## Comparison of composition:
## COMPOSITION (WEIGHT, %)
| Device | Au (%) | Pt (%) | Pd (%) | Aq (%) | CU(%) | IN(%) |
|--------------|--------|--------|--------|--------|-------|-------|
| V-Supra Plus | 76.00 | 7.80 | 9.80 | 2.50 | 0.0 | 3.70 |
| CERANUM 76 | 76.00 | 4.00 | 10.00 | 2.97 | 2.97 | 3.97 |
# Comparison of physical and mechanical properties:
| Alloy | Melting Point<br>Range (°F) | Hardness<br>(Vickers) | Yield<br>strength<br>(MPa) | Elongation<br>(%) | CTE<br>(x10-6/<br>°C) |
|--------------|-----------------------------|-----------------------|----------------------------|-------------------|-----------------------|
| V-Supra Plus | 2048-2,246 | 230 | 540 | 10.0 | 14.4 |
| CERANUM 76 | 2,066-2,192 | 170 | 350 | 13.0 | 15.1 |
{1}------------------------------------------------
## Discussion:
Since the composition of the leqally marketed alloy and the new device is very similar, it may be assumed that the biological compatibility of the alloys is also very similar.
## Conclusion:
The main elements and their concentration are almost identical. CERANUM 76 is a High Noble, Micro-fine, Yellow, Gold based alloy to be used for inlays, onlays, single crowns, bridge, implant superstructures and substrate for medium expansion with lower fusing porcelains and indirect restorative composites. CERANUM 76 is a high gold ceramic alloy, which heightens the porcelain esthetics of the restoration and provides strength, durability and color of gold. Despite minor differences in the materials, we believe that CERAUM 76 is a substantially equivalent to Metalor's V-Supra Plus. These changes do not affect safety or effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three heads facing to the right. The words "DEPARTMENT OF HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
ood and Drug Administratio 9200 Corporate Boulevard Rockville MD 20850
NOV 4:2002
Mr. Dae-Kyu Chang SB Lucius, Incorporated 9778 Katella Avenue, Suite 205 Anaheim, California 92804
Re: K023200
Trade/Device Name: Ceranum 76 Regulation Number: 21 CFR 872.3060 Regulation Name: Gold-Based Alloys and Precious Metal Alloys for Clinical Use Regulatory Class: II Product Code: EJT Dated: September 16, 2002 Received: September 25, 2002
Dear Mr. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
## Page 2 - Mr. Chang
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus. permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy A. Ulatowski
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### LUCIUS. INC
9778 Katella Ave
Ste-205, Anaheim, CA 92804
x : (14) 5
# INDICATIONS FOR USE
11023200 510(K) Number :
Device Name(s) : ceranum 76
This is a yellow ceramic alloy for single units, implants prosthetics, single & 3/4 crowns, and short span bridges.
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF CHRD,OFFICE OF DEVICE EVALUATION(OED)
Susan Quinn
esthesiology, General Hospital,
510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.