← Product Code [MOS](/productcode/MOS) · K023084

# MODEL 472GE-128: QUADRATURE LOWER EXTREMITY COIL (K023084)

_Medical Advances, Inc. · MOS · Dec 16, 2002 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K023084

## Device Facts

- **Applicant:** Medical Advances, Inc.
- **Product Code:** [MOS](/productcode/MOS.md)
- **Decision Date:** Dec 16, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Magnetic resonance imaging (MRI) and magnetic resonance angiography (MRA) of the musculoskeletal structures, soft tissue and vascular structures of the lower extremities.

## Device Story

Model 472GE-128 Quadrature Lower Extremity Coil is an accessory for GE Signa 3.0T MRI systems. It functions as a radiofrequency receiver coil to capture MR signals from lower extremity anatomy. The coil enables 2D/3D imaging, proton density, T1/T2 weighted imaging, and time-of-flight/phase-contrast angiography. Used in clinical radiology settings by trained MRI technologists/physicians. The device improves signal-to-noise ratio (SNR) by a factor of two compared to predicate coils, facilitating higher quality diagnostic images for clinical decision-making regarding musculoskeletal and vascular pathology.

## Clinical Evidence

Bench testing only. Performance verified through comparison of imaging parameters (SNR, uniformity, geometric distortion, slice thickness, spatial resolution) against predicate device. No clinical data provided.

## Technological Characteristics

Quadrature receiver coil for lower extremity MRI. Compatible with GE Signa 3.0T systems. Materials comply with UL 94 flammability standards. Electrical safety conforms to UL 2601-1 and IEC 601-1. Passive hardware component; no internal software or active processing algorithms.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KC23084

DEC 1 6 2002

# 510(k) Summary of Safety and Effectiveness

ﺮ

| Device Name                       | Model 472GE-128 Quadrature Lower Extremity<br>Coil                                                               |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------|
| Applicability                     | Compatible with GE Signa 3.0T MR Systems                                                                         |
| Reason for 510(k)                 | New device                                                                                                       |
| Classification Name               | Magnetic Resonance Diagnostic Device                                                                             |
| Device Classification Panel       | Radiology                                                                                                        |
| Device Classification Number      | 892.1000                                                                                                         |
| Product Code                      | 90MOS                                                                                                            |
| Common Name                       | Magnetic Resonance Imaging Coil                                                                                  |
| Proprietary Name                  | Model 472GE-128 Quadrature Lower Extremity<br>Coil                                                               |
| Establishment Registration Number | 2183683                                                                                                          |
| Address of MFG Facility           | IGC-Medical Advances Inc.<br>10437 Innovation Drive<br>Milwaukee, WI 53226                                       |
| Point of Contact                  | Anthony Dietzler<br>Quality Assurance Engineer<br>(414) 258-3808 Ext. 255                                        |
| Classification                    | Class II                                                                                                         |
| Intended Uses                     |                                                                                                                  |
| Diagnostic Uses                   | 2D, 3D imaging, proton density, T1 and T2<br>weighted imaging. 2D, 3D time of flight, phase<br>contrast imaging. |
| Anatomic Regions                  | Musculoskeletal structures, soft tissue and vascular<br>structures of the lower extremities                      |

{1}------------------------------------------------

#### Standards

| Performance Standards      | None Established under Section 514 |                                                                         |
|----------------------------|------------------------------------|-------------------------------------------------------------------------|
| Voluntary Safety Standards | UL 2601-1                          | Medical Electrical Equipment, Part I<br>General Requirements for Safety |
|                            | UL 94                              | Tests for Flammability of Plastic<br>Materials                          |
|                            | IEC 601-1                          | General Safety Requirements for<br>Medical Electrical Equipment         |

#### Overview

The Radiology Devices Panel considered potential concerns regarding the safe and effective operation of Magnetic Resonance Diagnostic Devices when they recommended reclassification to Class II on July 27, 1987. After reclassification, the FDA's Center for Devices and Radiological Health (CDRH) released a draft guidance document for the content and review of Magnetic Resonance Diagnostic Device premarket notification submissions that offered clarification of these concerns. Due to considerable technological advances in MRDDs, CDRH issued an updated guidance document on November 14, 1998. The following is a summary of the information contained within this premarket notification that addresses these concerns:

The GE Signa 3.0T MRI system, operated with the Medical Advances Quadrature Lower Extremity Coil, is substantially equivalent to the GE Signa 1.5T system operated with the legally marketed predicate device listed in section 4.0, within the Class II definition of Magnetic Resonance Diagnostic Device with respect to the safety parameter action levels:

#### Safety Parameters

| Maximum Static Magnetic Field:          | No change due to coil |
|-----------------------------------------|-----------------------|
| Rate of Magnetic Field Strength Change: | No change due to coil |
| RF Power Deposition:                    | No change             |
| Acoustic Noise Levels:                  | No change due to coil |
| Biocompatibility:                       | No change             |

{2}------------------------------------------------

## Imaging Performance Parameters

| Specification Volume:             | No change                                    |
|-----------------------------------|----------------------------------------------|
| Signal-to-Noise Ratio:            | 2 Times Improvement over Predicate<br>Device |
| Image Uniformity:                 | No change                                    |
| Geometric Distortion:             | No change                                    |
| Slice Thickness and Gap:          | No change                                    |
| High Contrast Spatial Resolution: | No change                                    |

### General Safety and Effectiveness Concerns

The device contains instructions for use. It includes indications for use, precautions, cautions, contraindications, warnings and quality assurance testing. This information assures safe and effective use of the device.

#### Substantial Equivalence Summary

The GE Signa 3.0T MRI system operated with the Medical Advances Quadrature Lower Extremity Coil addressed in this PMN, has the same intended use and technological characteristics as the GE Signa 1.5T system operated with the identified legally marketed predicate device. The use of this coil does not affect the GE Signa 3.0T system safety parameter specifications.

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the caduceus symbol. The text is in all capital letters and is written in a simple, sans-serif font.

#### Public Health Service

C 1 6 2002

d and Drug Administration 00 Corporate Boulevard Rockville MD 20850

Mr. Anthony Dietzler Ouality Assurance Engineering IGC-Medical Advances, Inc. 10437 Innovation Drive MILWAUKEE WI 53226

#### Re: K023084

Trade/Device Name: Model 472GE-128: Quadrature Lower Extremity Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 MOS Dated: September 16, 2002 Received: September 17, 2002

Dear Mr. Dietzler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Snigdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

- 510(k) Number (if known):

Model 472GE-128: Quadrature Lower Extremity Coil Device Name:

Indications for Use:

Magnetic resonance imaging (MRI) and magnetic resonance angiography (MRA) of the musculoskeletal structures, soft tissue and vascular structures of the lower extremities.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

David G. Ligman

of Reproductive, Abdom Offical Davi

---

**Source:** [https://fda.innolitics.com/device/K023084](https://fda.innolitics.com/device/K023084)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
