K023075 · Bd Vacutainer Systems Preanalyical Solutions · JKA · Nov 25, 2002 · Clinical Chemistry
Device Facts
Record ID
K023075
Device Name
BD VACUTAINER PLUS SST SERUM SEPARATOR TUBE
Applicant
Bd Vacutainer Systems Preanalyical Solutions
Product Code
JKA · Clinical Chemistry
Decision Date
Nov 25, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
The Vacutainer™ PLUS SST™ Tube is a plastic evacuated blood collection tube with silica clot activator and gel that provides a means of collecting, transporting, separating, and processing blood in a closed tube. Blood collected in a Vacutainer™ PLUS SST™ Tube is primarily used for clinical laboratory testing in chemistry using patient serum, but may be used for other assays requiring serum specimens as determined by the laboratory. In addition, the Vacutainer™ PLUS SST™ Tube is compatible with many commonly used therapeutic drugs therefore suitable for therapeutic drug monitoring (TDM). Blood can be collected, processed and stored in a Vacutainer™ PLUS SST™ tube for at least 24 hours for therapeutic drug monitoring without large losses in recovery.
Device Story
Sterile, plastic, evacuated blood collection tube; contains closure assembly, inert polyester gel barrier, silica clot activator, and silicone surfactant coating. Used in clinical laboratory settings for blood collection and processing. Centrifugation causes gel barrier to form at serum/blood clot interface, mechanically separating serum from cells. Enables collection, transport, and storage of blood for at least 24 hours for therapeutic drug monitoring (TDM) without significant recovery loss. Supports clinical decision-making by maintaining specimen integrity for TDM and chemistry assays.
Clinical Evidence
Clinical evaluation performed to validate the 24-hour stability claim for therapeutic drug monitoring (TDM). Results demonstrated that blood collected, processed, and stored in the tube for at least 24 hours maintains specimen integrity without large losses in recovery for TDM assays.
Technological Characteristics
Plastic evacuated tube; components include closure assembly, inert polyester gel barrier, silica clot activator, and silicone surfactant coating. Mechanical separation principle via centrifugation. Non-powered, standalone device.
Indications for Use
Indicated for the collection, transport, separation, and processing of blood for clinical laboratory testing, specifically chemistry assays and therapeutic drug monitoring (TDM) using patient serum.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
Vacutainer™ PLUS SST™ Serum Separator Tube (K960250)
Submission Summary (Full Text)
{0}------------------------------------------------
16023075
# NOV 2 5 2002
## J. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
## 510(k) Summary Of Safety and Effectiveness
- I. General Information
This Summary of Safety and Effectiveness information is being submitted in accordance with the requirements of the SMDA of 1990 and 21 § 807.92
#### Establishment:
| • | Address: | BD Vacutainer Systems, Preanalytical<br>Solutions<br>1 Becton Drive, MC 300<br>Franklin Lakes, NJ 07417-1885 |
|---|------------------------|--------------------------------------------------------------------------------------------------------------|
| • | Registration Number: | 2243072 |
| • | Contact Person: | Jing Zhang<br>Manager Regulatory Affairs<br>Telephone No.:(201) 847-4717<br>Fax No. (201) 847-4858 |
| • | Date of Summary: | Nov. 21, 2002 |
| | Device | |
| • | Trade Name: | BD Vacutainer™ PLUS SST™ Tube |
| • | Classification Name: | Tubes, Vials, Systems, Serum Separators<br>Blood Collection |
| • | Classification: | Class II |
| • | Performance Standards: | None Established under 514 of the Food,<br>Drug and Cosmetic Act |
{1}------------------------------------------------
- II. Safety and Effectiveness Information Supporting the Substantial Equivalence Determination
#### > Device Description:
The Vacutainer™ PLUS SST™ Tubes are sterile, plastic, evacuated blood collection tubes. The Vacutainer™ PLUS SST™ Tube consists of: (1) a closure assembly, (2) an inert polyester gel barrier, (3) silica clot activator, and (4) a silicone surfactant coated plastic tube. The specimen is centrifuged and the barrier material forms at the serum/blood clot interface, mechanically separating the serum from cells. The serum portion is used for clinical laboratory assays involving the use of patient serum.
#### > Intended Use:
The Vacutainer™ PLUS SST™ Tube is a plastic evacuated blood collection tube with silica clot activator and gel that provides a means of collecting, transporting, separating, and processing blood in a closed tube. Blood collected in a Vacutainer™ PLUS SST™ Tube is primarily used for clinical laboratory testing in chemistry using patient serum, but may be used for other assays requiring serum specimens as determined by the laboratory. In addition, the Vacutainer™ PLUS SST™ Tube is compatible with many commonly used therapeutic drugs therefore suitable for therapeutic drug monitoring (TDM). Blood can be collected, processed and stored in a Vacutainer™ PLUS SST™ tube for at least 24 hours for therapeutic drug monitoring without large losses in recovery.
### > Claims:
Blood can be collected, processed and stored in a Vacutainer™ PLUS SST™ tube for at least 24 hours for therapeutic drug monitoring without large losses in recovery. The storage conditions and time until analysis should be considered when collecting blood in Vacutainer™ PLUS SST™ tubes for TDM. This will ensure specimen integrity is maintained.
#### Synopsis of Test Methods and Results: A
Clinical evaluations were performed to determine the safety and efficacy of the Vacutainer™ PLUS SST™ Tube. The results of the clinical evaluation demonstrated that blood can be collected, processed and stored in a Vacutainer™ PLUS SST™ tube for at least 24 hours for therapeutic drug monitoring without large losses in recovery. The storage conditions and time until analysis should be considered when collecting blood in Vacutainer™ PLUS SST™ tubes for TDM. This will ensure specimen integrity is maintained.
{2}------------------------------------------------
- > Substantial Equivalence:
Based on a comparison of the device features, materials, and intended use, the Vacutainer™ PLUS SST™ Tubes are substantially equivalent to the commercially available predicate device. The only difference between the predicate and the Vacutainer™ PLUS SSTTM Tubes is the performance claim for therapeutic drug monitoring (TDM). The predicate device, K number, and clearance date are identified below:
| Manufacturer | Predicate Device | K-Number | Clearance Date |
|---------------------------------------------------|-------------------------------------------------------|----------|----------------|
| BD Vacutainer Systems,<br>Preanalytical Solutions | VacutainerTM<br>PLUS SSTTM<br>Serum Separator<br>Tube | K960250 | March 29, 1996 |
3M
Jing Zhang Manager Regulatory Affairs
11/21/02
Date
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads, representing the department's mission to protect the health of all Americans. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around it.
Food and Drug Administratio 2098 Gaither Road Rockville MD 20850
Mr. Jing Zhang Manager, Regulatory Affairs · · BD Vacutainer Systems, Preanalytical Solutions 1 Becton Drive, MCC 300 Franklin Lakes, NJ 07417
Re: k023075
Trade/Device Name: BD Vacutainer™ PLUS SSTTM Tube Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code: JKA Dated: August 29, 2002 Received: September 16, 2002
Dear Mr. Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your deviceto proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## B. INDICATIONS FOR USE
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: BD Vacutainer™ PLUS SST™ Tube
Indications for Use:
The Vacutainer™ PLUS SST™ Tube is a plastic evacuated blood collection tube with silica clot activator and gel that provides a means of collecting, transporting, separating, and processing blood in a closed tube. Blood collected in a Vacutainer™ PLUS SST™ Tube is primarily used for clinical laboratory testing in chemistry using patient serum, but may be used for other assays requiring serum specimens as determined by the laboratory. In addition, the Vacutainer™ PLUS SST™ Tube is compatible with many commonly used therapeutic drugs therefore suitable for therapeutic drug monitoring (TDM). Blood can be collected, processed and stored in a Vacutainer™ PLUS SST™ tube for at least 24 hours for therapeutic drug monitoring without large losses in recovery.
(Please do not write below this line-continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR § 801.109)
(Optional format 1-2-96)
Or Over-the-Counter Use***_***
Sean Cooper
(Division Sign-Off)
Division of Clinical Labora. Jvices
510(k) Number K023075
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.