PEDIATRIC ATTENUATED DEFIBRILLATION ELECTRODES, MODEL 9730, POWEDERHEART AED, MODEL 9200/9210

K022929 · Cardiac Science, Inc. · MKJ · Jan 27, 2003 · Cardiovascular

Device Facts

Record IDK022929
Device NamePEDIATRIC ATTENUATED DEFIBRILLATION ELECTRODES, MODEL 9730, POWEDERHEART AED, MODEL 9200/9210
ApplicantCardiac Science, Inc.
Product CodeMKJ · Cardiovascular
Decision DateJan 27, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5310
Device ClassClass 3
AttributesTherapeutic, Pediatric

Indications for Use

The Powerheart AED is intended for use by personnel who have been trained in its operation. The user should be qualified by training in basic life support or other physician-authorized emergency medical response. The device is indicated for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. Post-resuscitation, if the victim is breathing, the AED should be left attached to allow for acquisition and detection of the ECG rhythm. If a shockable ventricular tachyarrythmia recurs, the device will charge automatically and advise the operator to deliver therapy. When the patient is a child or infant up to 8 years of age, or up to 55 lbs (25kg), the Powerheart AED should be used with the Model 9730 Pediatric Attenuated Defibrillation Electrodes. The therapy should not be delayed to determine the patient's exact age or weight.

Device Story

Powerheart AED is a portable, battery-operated, semi-automatic, low-power DC defibrillator. Device inputs include patient ECG signals via electrodes. Internal circuitry diagnoses cardiac rhythm; advises operator if/when to deliver shock. For pediatric patients (up to 8 years/55 lbs), Model 9730 electrodes utilize an attenuating circuit to reduce biphasic waveform energy to pediatric-appropriate levels without impacting ECG monitoring. Device used in emergency settings by trained BLS personnel or physician-authorized responders. Output consists of visual/audible prompts for therapy delivery. Device monitors rhythm post-resuscitation; automatically recharges if shockable ventricular tachyarrhythmia recurs. Benefits include rapid, automated life-saving intervention for cardiac arrest.

Clinical Evidence

No clinical data. Performance verified via bench testing in accordance with industry standards ANSI AAMI DF-39. Biocompatibility assessments conducted on patient-contacting components.

Technological Characteristics

Portable, battery-operated, semi-automatic, low-power DC defibrillator. Uses biphasic waveform. Pediatric electrodes (Model 9730) incorporate an attenuating circuit to reduce energy delivery. Biocompatible materials used for patient-contacting components. Complies with ANSI AAMI DF-39 standards.

Indications for Use

Indicated for emergency treatment of unresponsive, non-breathing victims of sudden cardiac arrest. Pediatric use (children/infants up to 8 years or 55 lbs) requires Model 9730 Pediatric Attenuated Defibrillation Electrodes.

Regulatory Classification

Identification

An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ IC022929 ## SECTION 17. 510K SUMMARY OF SAFETY AND EFFECTIVENESS JAN 2 7 2003 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92. ## 17.1 Adminstrative Information | Name and Address<br>Submitted by: | Cardiac Science Incorporated<br>16931 Millikan Ave<br>Irvine, CA 92606 | |---------------------------------------------------------------|----------------------------------------------------------------------------| | Contact Person:<br>Telephone No.:<br>Facsimile No.:<br>Email: | Kenneth Olson<br>952-939-4181<br>952-939-4191<br>kolson@cardiacscience.com | | Date Prepared: | January 15, 2003 | #### Device Information 17.2 | Common or Usual Name: | Automated External Defibrillator | |-----------------------|-----------------------------------------------------------------------------------------------------------| | Trade Name: | Powerheart® AED Model 9200/9210<br>Use with Models 9730 Pediatric Attenuated<br>Defibrillation Electrode. | #### Device Classification 17.3 Automatic External Defibrillator Classification Name: 21 CFR 870.1025 MKJ Device Class: III > Multifunctional Electrocardiograph Electrode 21 CFR 870.2360 MLN Device Class: III #### 17.4 Device Description The Powerheart AED is a portable, battery-operated, semi-automatic, low power DC defibrillator. It will diagnose the patient's cardiac rhythm and advise the operator if and {1}------------------------------------------------ when to deliver the shock energy. The Model 9730 Pediatric Attenuated Defibrillation Electrodes are specially designed electrodes for operation with only the Powerheart AED biphasic waveform. The Model 9730 pediatric electrode has been designed to reduce the energy delivered to the patient from a standard biphasic defibrillation waveforms generated by the Powerheart AED to levels applicable for use on children and infants up to 8 yrs of age or weighing less than 55 lbs (25 kg). This is performed though the use of an attenuating circuit that absorbs the excess energy during defibrillation without adversely affecting the electrode's performance during ECG monitoring. The Model 9730 pediatric electrode is designed as a single use device and is intended for use to defibrillate and monitor of children. ### 17.5 Indication for use The Powerheart AED is intended for used by personnel who have been trained in its operation. The user should be qualified by training in basic life support or other physician-authorized emergency medical response. The device is indicated for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. Post-resuscitation, if the victim is breathing, the AED should be left attached to allow for acquisition and detection of the ECG rhythm. If a shockable ventricular tachyarrythmia recurs, the device will charge automatically and advise the operator to deliver therapy. When the patient is a child or infant up to 8 years of age, or up to 55 lbs (25kg), the Powerheart AED should be used with the Model 9730 Pediatric Attenuated Defibrillation Electrodes. The therapy should not be delayed to determine the patient's exact age or weight. {2}------------------------------------------------ ## 17.6 Identification of Predicate Device | Company | Device | 510(k) # | Date Cleared | |-------------------------------------------------------------------|------------------------------------------------------------------------------|----------|--------------| | Agilent Technologies<br>(Philips Medical<br>System) | FR2 AED with Model M3870A Pediatric<br>Attenuated Defibrillation Electrodes. | K003819 | May 2, 2001 | | Survivalink<br>(Wholly Owned<br>Subsidiary of Cardiac<br>Science) | Powerheart AED | K011901 | Feb 1, 2002 | #### Substantial Equivalence 17.7 The Powerheart AED with the Model 9730 Pediatric Attenuated Defibrillation Electrodes covered by this submission is substantially equivalent to other legally marketed AED and pediatric attenuated defibrillation electrodes. Specifically, the Powerheart AED with the Model 9730 electrode is substantially equivalent to Agilent Technologies (acquired by Philips Medical System) FR2 AED with model M3870A pediatric attenuated defibrillation cleared under premarket 510(k) notification, K0038198 on May 2, 2002. ## 17.8 Performance Data Performance testing was conducted to verify that the Powerheart AED with the Model 9730 Pediatric Attenuated Defibrillation Electrodes has been designed in accordance with industry standards ANSI AAMI DF-39. In addition, biocompatibility assessments were conducted on the components that come into contact with the patient's skin. The Powerheart AED with the Model 9730 electrodes were found to perform as intended. ## 17.9 Conclusions Cardiac Science has demonstrated through its evaluation and testing of the Powerheart AED with the model 9730 Pediatric Attenuated Defibrillation Electrodes that the device is equivalent to the predicate device with respect to intended use, technological characteristics, materials, function and safety and effectiveness. This notification contains all information required by 21 CFR 807.87. A completed copy of the Premarket Notification 510(k) Reviewer's Checklist is provided in this submission. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all capital letters and is evenly spaced around the symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 2 7 2003 Cardiac Science Incorporated Mr. Kenneth Olson Chief technology Officer 16931 Millikan Avenue Irvine, CA 92606 Re: K022929 > Trade/Device Name: Powerheart AED Model 9200/9210 with Model 9730 Pediatric Attenuated Defibrillation Electrodes Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm. Regulatory Class: Class III Product Code: MKJ, MLN Dated: December 13, 2002 Received: December 16, 2002 Dear Mr. Olson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {4}------------------------------------------------ Page 2 -- Mr. Kenneth Olson forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely vours. K.G. Tell Bram D. uckerman. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE K022929 510(k) Number: Device Name: Powerheart AED Model 9200/9210 Us Ha Model 9730 Pediating Cetternated Efectrodes Indications for Use: The Powerheart AED is intended for use by personnel who have been trained in its operation. The user should be qualified by training in basic life support or other physician-authorized emergency medical response. The device is indicated for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. Post-resuscitation, if the victim is breathing, the AED should be left attached to allow for acquisition and detection of the ECG rhythm. If a shockable ventricular tachyarrythmia recurs, the device will charge automatically and advise the operator to deliver therapy. When the patient is a child or infant is up to 8 years of age, or up to 55 lbs (25kg), the Powerheart AED should be used with the Model 9730 Pediatric Attenuated Defibrillation Electrodes. The therapy should not be delayed to determine the patient's exact age or weight. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | | |------------------------------------|--| | Division of Cardiovascular Devices | | | 510(k) Number | K022929 | |---------------|---------| |---------------|---------| | Prescription Use | | OR | Over-The-Counter Use | |----------------------|--|----|----------------------| | (Per 21 CFR 801.109) | | | |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...