← Product Code [LLZ](/productcode/LLZ) · K022877

# OBSERVER ULTRASOUND REPORTING AND IMAGING SYSTEM (K022877)

_Vision Chips, Inc. · LLZ · Feb 5, 2003 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K022877

## Device Facts

- **Applicant:** Vision Chips, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Feb 5, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

OBserver Ultrasound Reporting and Imaging System is intended to automate and enhance the process by which ultrasound examination is entered and reported and by which ultrasound images are stored and reviewed by physicians in the field of maternal-fetal medicine. The system also provides significant research data for academic institutions.

## Device Story

OBserver is an OB/GYN ultrasound reporting and imaging system. It ingests ultrasound data via serial interface (ASCII text) or network (DICOM); captures images via frame grabber from video output. System stores still images as JPEG/DICOM and cine loops as MPEG/AVI/multi-frame DICOM. Used in clinical settings by perinatologists, obstetricians, gynecologists, radiologists, and technicians to generate electronic/printed reports, perform OB/GYN calculations (e.g., fetal measurements to gestational age), and conduct research. Physicians review captured images and data to support clinical decision-making in maternal-fetal medicine and antepartum surveillance.

## Clinical Evidence

No clinical data provided. Bench testing only.

## Technological Characteristics

System utilizes standard off-the-shelf hardware (computers, storage, monitors, frame grabbers). Connectivity via serial interface or network (DICOM). Software supports OB/GYN reporting and calculations. Runs on Microsoft operating systems (text reporting also supports Macintosh).

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- ALI Ultras PACS and comPACS ([K963610](/device/K963610.md))
- R4's Remote Fetal Medicine Ultrasound System ([K000443](/device/K000443.md))
- Digisonics OB-View ([K970402](/device/K970402.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) Summary K022877

## Vision Chips, Inc. OBserver™ OB/GYN Ultrasound Reporting and Imaging System

#### Name of Submitter:

Vision Chips, Inc. 27111 Aliso Creek Road, Suite 120 Aliso Viejo, CA 92656 (949) 362-0565 (949) 362-0591 Fax

#### Contact Person:

Ronald "Jock" Philip, President

## Date of Summary:

August 29, 2002/Revised 11/25/02

## Device and Common Names:

Name of Device: OBserver OB/GYN Ultrasound Reporting, Information & Imaging System Common Name: Image Acquisition, Review and Reporting System Classification Name: Picture Archiving and Communications System

## Device Description:

#### General Description

The Vision Chips ultrasound reporting and imaging system consists of comprehensive reporting and image archiving software for use in the field of obstetrics and gynecology. It enables physicians to generate reports documenting the results of ultrasound exams and to view the images that were captured during those exams. The system also enables physicians to perform research on all data that have been entered into the program.

Ultrasound machines may transfer ASCII text based information to the system by means of a serial interface. Ultrasound images are received either by means of a standard frame grabber connected to the ultrasound video output or directly across the network using DICOM protocols. Still images are stored as standard jpegs or in DICOM format, cine loops are stored as mpegs, AVIs or in multi frame DICOM format.

All hardware required for the system, including computers, storage devices, monitors and frame grabbers, consists of standard, off the shelf components.

#### Indications for Use

OBserver Ultrasound Reporting and Imaging System is intended to automate and enhance the process by which ultrasound examination information is entered and by which ultrasound images are stored and reviewed by physicians in the field of maternal-fetal medicine. The system also provides significant research data for academic institutions.

{1}------------------------------------------------

## User Characteristics

OBserver is a comprehensive Ob/Gyn Ultrasound and Antepartum surveillance(NST, CST, BPP & AFI) information reporting system. It allows perinatologists, obstetricians, gynecologists, radiologists and ultrasound technicians to capture examination data and images from ultrasound machines and to generate electronic and printed medical reports from them.

## Substantial Equivalence:

OBserver Ultrasound Reporting and Imaging System is substantially equivalent to the following legally-marketed devices:

## 1) ALI Ultras PACS and comPACS (K963610)

2) R4's Remote Fetal Medicine Ultrasound System. (K000443)

3) Digisonics OB-View (K970402)

OBserver Ultrasound Reporting & Imaging System, like R4's Remote Fetal

Medicine Ultrasound System, Digisonics' OB-View, and ALL's includes review and reporting capabilities, ob/gyn calculations, and digital imaging. Vision Chips' OBserver, like R4's Fetal Ultrasound Reporting and ALI's comPACS, performs ob/gyn reporting and includes calculations to convert fetal measurements to gestational ages. They all have the ability to review ultrasound images captured during the course of the examination. The ultrasound images are captured and converted to digital format with the use of a frame grabber. They are saved as JPEG files on standard computer storage media. The Vision Chips and ALI systems will also accept images and other data that have been transmitted across a network in DICOM format. All of the systems run on Microsoft operating systems although the text reporting section of Vision Chips' system will also run on Macintosh operating systems.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird with three parallel lines forming its body and wings, suggesting movement or flight.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# FEB 0 5 2003

Mr. Ronald Philip President Vision Chips, Inc. 27111 Aliso Creek Road #120 VIEJO CA 92656

Re: K022877

Trade/Device Name: Observer Ultrasound Reporting and Imaging System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: November 26, 2002 Received: November 26, 2002

Dear Mr. Philip:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050,

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx. 1xxx                        | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

510(k) Number (if known): K 0 2 2 8 77 Device Name: ORS Ultrasound Indications For Use:

OBserver Ultrasound Reporting and Imaging System is intended to automate and enhance the process by which ultrasound examination is entered and reported and by which ultrasound images are stored and reviewed by physicians in the field of maternal-fetal medicine. The system also provides significant research data for academic institutions.

David M. Seaman

*Prescription Use*

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

---

**Source:** [https://fda.innolitics.com/device/K022877](https://fda.innolitics.com/device/K022877)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
