K022774 · Intl. Remote Imaging Systems · LKM · Oct 21, 2002 · Hematology
Device Facts
Record ID
K022774
Device Name
IQ 200 SYSTEM
Applicant
Intl. Remote Imaging Systems
Product Code
LKM · Hematology
Decision Date
Oct 21, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5200
Device Class
Class 2
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Microscopic formed elements concentration
Auto analyte recognition algorithm
—
Sensitivity exceeded 939UDx and Sysmex UF-100; specificity comparable to 939UDx and Sysmex UF-100.
—
—
Clinical performance study comparing iQ 200 System to 939UDx Urine Pathology System and Sysmex UF-100.
>1 (expert human identification)
Indications for Use
The iQ 200 System is an in-vitro diagnostic device used to automate the complete urinalysis profile, including urine test strip chemistry panel and microscopic sediment analysis. Optionally, the iQ 200 Analyzer can be used as a stand-alone unit, or the results from the iQ 200 Analyzer can be combined with other urine chemistry results received from an LIS. It produces quantitative or qualitative counts of all formed sediment elements present in urine, including cells, casts, crystals, and organisms. A competent human operator can set criteria for auto-reporting and flagging specimens for review. All instrument analyte image decisions may be reviewed and overridden by a trained technologist.
Device Story
Automated urinalysis system; processes urine samples via autosampler. Transforms samples into images for automated analyte recognition; classifies formed sediment elements (cells, casts, crystals, organisms). Operates in clinical laboratory settings; managed by trained technologists. System consolidates chemistry results from companion analyzer (ARKRAY AUTION MAX AX-4280) and microscopic findings on a dedicated Computer WorkStation. Technologists review, confirm, or modify automated classifications; results released off-line. Facilitates physician health assessment and differential diagnosis by providing quantitative/qualitative sediment counts.
Clinical Evidence
Comparative performance study against 939UDx Urine Pathology System and Sysmex UF-100. Results showed iQ 200 System had greater clinical sensitivity for abnormal specimens and individual microscopic formed elements. Specificity was comparable to predicates. With sensitivity fixed at 90%, iQ 200 specificity exceeded predicates for most elements. Quantitative reporting (R Square) for RBC, WBC, squamous epithelial cells, and casts matched expert human identification and predicate performance.
Technological Characteristics
Automated sample handling via autosampler; intelligent microscopy; automated analyte classification algorithm. Consolidates chemistry data from external analyzer via Computer WorkStation. Enables off-line review and manual override of classifications. Standalone or networked configuration.
Indications for Use
Indicated for routine urinalysis in patients undergoing screening for metabolic/systemic disease, renal/urinary tract dysfunction, infection, carcinoma, or injury; monitoring therapy effectiveness; and post-surgical/post-therapeutic recovery.
Regulatory Classification
Identification
An automated cell counter is a fully-automated or semi-automated device used to count red blood cells, white blood cells, or blood platelets using a sample of the patient's peripheral blood (blood circulating in one of the body's extremities, such as the arm). These devices may also measure hemoglobin or hematocrit and may also calculate or measure one or more of the red cell indices (the erythrocyte mean corpuscular volume, the mean corpuscular hemoglobin, or the mean corpuscular hemoglobin concentration). These devices may use either an electronic particle counting method or an optical counting method.
Predicate Devices
The Yellow IRIS® Urinalysis Workstation with CHEMSTRIP® Reader
IRIS Flow Microscope
900UDx Urine Pathology System
939UDx Urine Pathology System
Sysmex UF-100
Reference Devices
ARKRAY AUTION MAX AX-4280
Submission Summary (Full Text)
{1}
OCT 21 2002
International Remote Imaging Systems, Inc.
9162 Eton Avenue, Chatsworth, CA 91311
800-776-4747 818-709-1244 FAX 818-700-9661
# 510(k) SUMMARY
510(k) Submission: K022774
iQ™ 200 System
## General Information:
Date of Submission: August 20, 2002
Trade Name of Device: iQ™ 200 System
Common Name of Device: Automated urinalysis system
Classification Name: Automated urinalysis system, 21 CFR 862.2900, Class I
Microscope, 21 CFR 862.3600, Class I
Automated cell counter (Urine particle counter), 21 CFR 864.5200, Class II
Submitter's Name: Harvey L. Kasdan, Ph.D.
Chief Scientist
Iris Diagnostics, A Division of
International Remote Imaging Systems, Inc.
9162 Eton Avenue, Chatsworth, CA 91311
## Indications for Use (Brief):
The iQ 200 System is intended for analysis of urine chemistry, color, clarity, specific gravity, and formed sediment elements, which constitutes typical routine urinalysis.
## Intended Use:
The iQ 200 System is an in-vitro diagnostic device used to automate the complete urinalysis profile, including urine test strip chemistry panel and microscopic sediment analysis. Optionally, the iQ 200 Analyzer can be used as a stand-alone unit, or the results from the iQ 200 Analyzer can be combined with other urine chemistry results received from an LIS. It produces quantitative or qualitative counts of all formed sediment elements present in urine, including cells, casts, crystals, and organisms. A competent human operator can set criteria for auto-reporting and flagging specimens for review. All instrument analyte image decisions may be reviewed and overridden by a trained technologist.
510(k) Summary, iQ™ 200 System
{2}
510(k) Summary, iQ™ 200 System
Page 2
## Substantial Equivalence to Predicate Devices:
The iQ 200 System is substantially equivalent under Section 510(k) of the Food, Drug, and Cosmetic Act to The Yellow IRIS® Urinalysis Workstation with CHEMSTRIP® Reader, the IRIS Flow Microscope, the 900UDx Urine Pathology System, the 939UDx Urine Pathology System, and the Sysmex UF-100.
## Summary of Technological Characteristics:
The iQ 200 System provides automatic sample handling for automated intelligent microscopy and automatic analyte classification for improved data reporting, presentation and management. Specimens are aspirated by an autosampler rather than poured manually. Each image is analyzed and assigned a classification by an auto analyte recognition algorithm. Using these classifications and the known observation volume, microscopic analyte concentrations may be automatically reported. If results from a specimen are not autoreported, microscopic examination results are displayed on an independent Computer WorkStation. Operators can then confirm or modify analyte classifications and release reports off-line for enhanced convenience, obviating the need to process a second aliquot for review. Chemistry results from the companion ARKRAY AUTION MAX AX-4280 are automatically consolidated by the Computer WorkStation for display and reporting.
## Performance Studies:
Microscopic analysis performance of the iQ 200 System were compared with that of 939UDx Urine Pathology System and the Sysmex UF-100.
## Conclusions Drawn From Clinical Tests:
Clinical sensitivity (fraction of abnormal specimens detected) of the iQ 200 System exceeded that of the 939UDx Urine Pathology System and Sysmex UF-100, while clinical specificity (fraction of normal specimens detected) was comparable. The iQ 200 System generally had greater sensitivity for detection of individual microscopic formed elements than either the 939UDx Urine Pathology System or the Sysmex UF-100. With sensitivity for detection of each microscopic formed element fixed at 90%, iQ 200 System specificity was greater than that of either the 939UDx Urine Pathology System or Sysmex UF-100 for most formed elements. iQ 200 System automatically reported concentrations of RBC, WBC, squamous epithelial cells and casts matched those determined from expert human identification as well as the 939UDx Urine Pathology System predicate automatically reported concentrations matched those of the human expert when measured by the square of the regression correlation coefficient, R Square. The iQ 200 System has been shown to be suitable for its intended use.
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
OCT 21 2002
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
Harvey L. Kasdan, Ph.D.
Chief Scientist
International Remote Imaging Systems, Inc.
9172 Eton Avenue
Chatsworth, California 91311-5874
Re: k022774
Trade/Device Name: iQ™ 200 System
Regulation Number: 21 CFR § 864.5200
Regulation Name: Automated Cell Counter
Regulatory Class: II
Product Code: LKM, KQO
Dated: August 20, 2002
Received: August 21, 2002
Dear Dr. Kasdan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
# PREMARKET NOTIFICATION
510(k) Number (if known): K022774
Device Name: iQ 200 System
## Indications for Use
The iQ 200 System is intended for analysis of urine chemistry, specific gravity, and formed sediment elements, which constitutes typical routine urinalysis.
Urinalysis is ordered by physicians as a screening procedure for detection of possible abnormal metabolic or systemic disease, indicated by chemical composition of urine, and of potential renal or urinary tract disease or dysfunction, indicated by urine concentration and by the nature and distribution of urinary formed elements. Urine profile testing is commonly employed in the initial clinical evaluation of patients admitted for hospital care or undergoing physical examinations.
Routine urinalysis is also indicated in diagnosis of patients with possible renal or urinary tract infection, carcinoma, or other injury, as well as for monitoring the status and effectiveness of drug, radiation, or dietary therapy, and of post-surgical or post-therapeutic recovery.
The information produced by the iQ 200 System concerning the composition of patient urines is ordered at the discretion of the physician, and is part of a larger body of laboratory and other test results available to assist the physician in health assessments or differential diagnoses. Routine urinalysis findings are always subject to judgment and interpretation by physicians relative to the patient's overall clinical presentation and history.

IQ™ 200 System
August 20, 2002
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.