DATEX-OHMEDA S/5 COMPACT CRITICAL CARE MONITOR WITH L-CICU02 & L-CICU02A SOFTWARE
Applicant
Datex-Ohmeda
Product Code
MLD · Cardiovascular
Decision Date
Nov 14, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A is intended for multiparameter patient monitoring.
Device Story
S/5 Compact Critical Care Monitor is a modular patient monitor for hospital use. It accepts inputs from specialized plug-in modules measuring ECG, invasive/non-invasive blood pressure, pulse oximetry, temperature, CO2, and BIS. The monitor processes these signals to display numeric values, waveforms, and trends on a screen. It features audible/visual alarms and supports network connectivity to central stations. Operated by clinicians via keyboard and ComWheel interface. Software (L-CICU02/L-CICU02A) performs arrhythmia analysis, ST-segment calculation, heart rate, impedance/respiration rate, energy expenditure, EEG spectrum analysis, and evoked potential averaging. Output assists clinicians in real-time patient status assessment and clinical decision-making. Benefits include continuous monitoring of critical physiological parameters to improve patient care and safety.
Clinical Evidence
Bench testing only. Verification included electrical safety, electromagnetic compatibility, mechanical/environmental tolerance, and software validation. Compliance with IEC 60601-1, IEC 60601-2-27, IEC 60601-2-30, IEC 60601-2-34, IEC 60601-2-40, and other relevant ISO/EN standards was verified.
Technological Characteristics
Modular multiparameter patient monitor. Connectivity: Networked or standalone, wireless LAN support. Software: L-CICU02/L-CICU02A. Standards: IEC 60601-1, IEC 60601-2-27 (ECG), IEC 60601-2-30 (NIBP), IEC 60601-2-34 (InvBP), IEC 60601-2-40 (EEG). Interface: Keyboard and ComWheel. Supports plug-in modules for ECG, BP, SpO2, Temp, CO2, and BIS.
Indications for Use
Indicated for monitoring hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion, Bispectral index (BIS), and neurophysiological status of all hospital patients. BIS monitoring is for brain state via EEG signal processing as an aid in monitoring effects of certain anesthetic agents. For use by qualified medical personnel only.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Datex-Ohmeda S/5™ Compact Critical Care Monitor with S-00C03, S-00C04 software (K002158)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a sequence of handwritten characters, "K022740", followed by the text "Page 1 of 5". The characters are written in a dark ink, and the text indicating the page number is smaller and positioned below the sequence of characters. The image appears to be a scan or photograph of a document.
NOV 1 4 2002
# Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software
# GENERAL COMPANY INFORMATION as required by 807.92(a)(1)
# COMPANY NAME/ADDRESS/PHONE/FAX:
Datex-Ohmeda 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344
# NAME OF CONTACT:
Mr. Joel Kent
DATE: August 16, 2002
# DEVICE NAME as required by 807.92(a)(2)
# TRADE NAME:
Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software
# COMMON NAME:
Patient Monitor
# CLASSIFICATION NAME:
The following Class III classification appears applicable:
| DSI | Arrhythmia detector & alarm | 870.1025 |
|-----|-----------------------------|----------|
| MLD | Monitor ST-segment & alarm | 870.1025 |
{1}------------------------------------------------
# NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The Datex-Ohmeda S/5TM Compact Critical Care Monitor with L-CICU02 and L-CICU02A software (S/5™ CCCM) is substantially equivalent in safety and effectiveness to the legally marketed predicate Datex-Ohmeda S/5™ Compact Critical Care Monitor with S-00C03, S-00C04 software (K002158).
# DEVICE DESCRIPTION as required by 807.92(a)(4)
The S/5TM Compact Critical Care Monitor is a patient monitor, which displays the measurement of patient physiological parameters in the hospital setting. The measurement of patient physiological parameters is accomplished by specialized measurement modules which, when plugged into the frame, allow the modules to communicate with the monitor. The caregiver can select from a variety of available measurements (parameters) and apply those parameters that are best suited to patient care. Modules perform the functions of parameter measurement and minor data processing. The S/5TM Compact Critical Care Monitor displays parameters on screen, signals alarms and performs advanced data processing. There are two software options available for the S/5™ Compact Critical Care Monitor: L-CICU02 and L-CICU02A. L-CICU02A is equipped with extended arrhythmia analysis capability. Other than arrhythmia analysis capabilities, this software option is identical to L-CICU02.
The modifications to the device are:
- Technical alarms such as "SpO2 probe off", "Leads off" (ECG), and "Px No transducer" 1. (InvBP) transferred to the Central Station.
- 2. Changed to include support for alarm silencing and alarm limits adjustment from Central Station.
- The alarm priority of "X module removed" message has been increased from note level 3. to yellow level.
- Software can additionally send request for recording to Central Station. 4.
- Snapshot printing speed 25 mm/s has been added. ર :
- Rotating of the "ComWheel" now scrolls through different trend pages. 6.
- Standby information is additionally sent to Central Station. 7.
- Mean Arterial Pressure value replaces O2 FI value in vital parameters numerical trend. 8.
- The NMT cycle time selections have been changed to be: 9. 20 s / 1 min / 5 min / 15 min / 30 min / 60 min / 120 min.
- 10. Lifetime of an old NIBP value has been changed from 60 min to 245 min.
- Support for M-BIS (Bispectral Index) module has been added. 11. M-BIS has a separate 510(k)
- 12. Support for the M-MINIC (CO2 module) module has been added. The Mini CO2 module M-miniC is the subject of a separate 510(k) premarket notification.
- Support for the N-DIS external device interfacing modules has been added. Support is 13. for interfacing the following device categories: ventilators/anesthesia machines, standalone monitors, blood gas analyzers and heart-lung machines.
- The N-DIS modules have a separate 510(k) premarket notification. Support for wireless LAN communication between the monitor and Datex-Ohmeda S/5 14.
- Central station.
The S/5TM CCCM uses several types of plug-in measurement modules.
Modules are the subject of separate 510(k)'s and are not part of this notification.
The S/5™ CCCM is typically furnished with a module that measures ECG, invasive and noninvasive blood pressures, pulse oximetry and temperature.
Modules are placed in the S/5™ Compact Monitor frame and are automatically recognized by the monitor. The patient cables are connected to the module plug in jacks and then monitoring can
{2}------------------------------------------------
begin. The S/5™ CCCM can display measurements in the form of numeric values, traces and trends. Audible and visual alarms are used to indicate patient status. The priority profile of an alarm depends on the parameter. The S/5™ CCCM is operated by a keyboard. Typically pressing a key results in a pop up menu appearing on the screen. Selections can then be made easily from the menu using a unique ergonomically designed pointing device on the keyboard called a ComWheel™. The software L-CICU02 and L-CICU02A perform some module-related tasks like arrhythmia analysis. ST-value calculation, heart rate calculation, impedance and respiration rate calculation, energy expenditure calculation, EEG spectrum analysis and evoked potential response averaging. All the module communication is also handled in the main software. There are various optional types of keyboards, some are like standard keyboards and another is a hand-held Remote controller (REMCO) which is still directly connected to the S/5™ Compact Critical Care Monitor via a long cord but provides more flexibility in controlling the monitor while the doctor or nurse is handling other patient care needs. The S/5™ Compact Critical Care Monitor can be in a stand-alone or networked configuration. If networked, measurement data is sent to the network for central station or monitor viewing. Trends can be sent via a network to a central computer for archiving.
# INTENDED USE as required by 807.92(a)(5)
## Intended use:
The S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A is intended for multiparameter patient monitoring.
# Indications for use:
The S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software is indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion, Bispectral index (BIS), and neurophysiological status of all hospital patients. The S/57M Compact Critical Care Monitor with L-CICU02 and L-CICU02A software when using BIS is for monitoring the state of the brain by data acquisition and processing of electroencephalograph signals and may be used as an aid in monitoring the effects of certain anesthetic agents*. (*Gan TJ, Glass P, Windsor A, Payne F, Rosow C, Sebel P, Manberg P. Bispectral Index Monitoring Allows Faster Emergence and Improved Recovery from Propofol, Alfentanil and Nitrous Oxide Anesthesia. Anesthesiology, October 1997; (4) 87:808-15.) The S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software is indicated for use by qualified medical personnel only.
# SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software (S/5™ CCCM) is substantially equivalent to the predicate Datex-Ohmeda S/5TM Compact Critical Care Monitor with S-00C03, S-00C04 software (K002158). The S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software is indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion, Bispectral index (BIS), and neurophysiological status of all hospital patients. The S/5TM Compact Critical Care Monitor with L-CICU02 and L-CICU02A software when using BIS is for monitoring the state of the brain by data acquisition and processing of electroencephalograph signals and may be used as an aid in monitoring the effects of certain anesthetic agents *.
(*Gan TJ, Glass P, Windsor A, Payne F, Rosow C, Sebel P, Manberg P. Bispectral Index Monitoring Allows Faster Emergence and Improved Recovery from Propofol, Alfentanil and Nitrous Oxide Anesthesia. Anesthesiology, October 1997; (4) 87:808-15.)
{3}------------------------------------------------
Page 4 of 5
There are two software options available for S/5™ Compact Critical Care Monitor: L-CICU02 and L-CICU02A (collectively referred to as L-CICU02(A) ). (Note: L- refers to software license). Only one software can be used at any given time in the monitor. The software is preloaded in the factory and can also be later loaded in the customer site. The new device with different software options S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A, is compared to predicates as outlined below.
The basic model of the monitor is S/5™ Compact Critical Care Monitor with L-CICU02, which is a new revision of the predicate devices, S/5™ Compact Critical Care Monitor with S-00C03 software (K002158). The S/5™ Compact Critical Care Monitor with L-CICU02 may be equipped with extended bedside arrhythmia analysis capability and in this case the monitor is called S/5TM Compact Critical Care Monitor with L-CICU02A. The arrhythmia analysis functionality of S/5™ Compact Critical Care Monitor with L-CICU02A is substantially equivalent to the functionality of the predicate device
S/5TM Compact Critical Care Monitor with L-00C04. The S/5™ CCCM is a modular multiparameter patient monitor providing connections to measurement modules. The general construction, indications for use and intended use of the S/5TM CCCM are the same as for the predicate S/5TM Compact Critical Care Monitor with S-00C03, S-00C04 software (K002158).Based on the above and a detailed analysis in Tab 4 Comparison and other documentation included in this 510(k) notification and attachments it is evident that the main features and indications for use of S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software is substantially equivalent to the predicate S/5™ Compact Critical Care with S-00C03, S-00C04 software (K002158).
# SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
The Datex-Ohmeda S/5™ Compact Critical Care monitor with L-CICU02 and L-CICU02A software complies with the safety standards below and is therefore safe and effective for the intended use. The device has been thoroughly tested including electrical safety, electromagnetic compatibility, mechanical and environmental tolerance, software validation and verification of specifications. Verification of compliance with the following mandatory and voluntary standards has been made:
- IEC 60601-1:1988+ Amdt.:1:1991 + Amdt. 2:1995 ●
- EN 60601-1: 1990 + A1:1993+A2:1995+A13:1996 .
- CAN/CSA-C22.2 No.601.1-MS0 +S1:1994+Amdt. 2:1998 .
- IEC 60601-2-27:1994/EN 60601-2-27:1994
- IEC 60601-2-30:1995/EN 60601-2-30:1995 ●
- IEC 60601-2-34:1994/EN 60601-2-34:1994 .
- IEC 60601-2-40:1998 .
- IEC 60601-1-2(1993)/EN 60601-1-2
- . IEC 60601-1-4: 1996+Amdt. 1:1999/EN 60601-1-4
- ISO 9918:1993/EN 864:1996 .
- ISO 9919:1992/EN865:1997 ●
- ISO 7767:1997/EN12598:1999 .
- ISO 11196:1995 + Corr. 1:1997/EN ISO11196:1997 ●
- IEC 601-2-10:1987/HD 395.2.10:1988 + Am.1:2000 .
- IEC 60601-2-26:1994/EN60601-2-26 .
- IEC 60068-2
- . UL 2601-1:1997
- ANSI/AAMI ES-1:1993 .
- ANSI/AAMI EC57:1998 .
- . FDA 21 CFR 898.12
{4}------------------------------------------------
Conclusion:
The summary above shows that there are no new questions of safety and effectiveness for the Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software
as compared to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a human figure with three lines representing the head, body, and legs.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 4 2002
Datex-Ohmeda c/o Mr. Joel C. Kent Manager, Quality and Regulatory Affairs 86 Pilgrim Rd. Needham, MA 02492
Re: K022740
Trade Name: Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software Regulation Name: Arrhythmia Detector and Alarm Regulation Number: 21 CFR 870.1025 Regulatory Class: Class III (three) Product Code: MLD Dated: August 17, 2002 Received: August 19, 2002
Dear Mr. Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 – Mr. Joel Kent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must. comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Elias Mallis
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### 510(k) Number (if known):
# Device Name: Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU02 and I CICU02A software
K022740
#### Indications For Use:
The S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software is indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion, Bispectral index (BIS), and neurophysiological status of all hospital patients.
The S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software when using BIS is for monitoring the state of the brain by data acquisition and processing of electroencephalograph signals and may be used as an aid in monitoring the effects of certain anesthetic agents.
The S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software is indicated for use by qualified medical personnel only.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Per 21 CFR 801.109)
OR
Over-The-Counter Use _
(Optional Format 1-2-96)
Elisa Mallen
Division of Cardiovascular & Respiratory Devices
510(k) Number K022740
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.