REFLECTION MODIFIED ACETABULAR SHELLS

K022556 · Smith & Nephew, Inc. · LPH · Aug 28, 2002 · Orthopedic

Device Facts

Record IDK022556
Device NameREFLECTION MODIFIED ACETABULAR SHELLS
ApplicantSmith & Nephew, Inc.
Product CodeLPH · Orthopedic
Decision DateAug 28, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion; femoral neck fracture and trochanteric fractures of the proximal femur with heads involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. The Reflection Modified Acetabular Shells are designed for uncemented applications and are single use only.

Device Story

Reflection Modified Acetabular Shells are titanium acetabular components featuring peripheral fixation holes and a porous titanium bead coating; optional hydroxyapatite coating available. Designed for uncemented application in total hip arthroplasty; intended for use with existing Reflection Hip System UHMWPE acetabular liners. Implanted by orthopedic surgeons in clinical settings during primary or revision hip surgery. Device provides structural support for hip joint reconstruction; mechanical testing confirms ability to withstand in vivo loading. Benefits include restoration of hip function in patients with degenerative or traumatic joint damage.

Clinical Evidence

Bench testing only; no clinical data provided. Mechanical test data indicates the device is equivalent to currently marketed components and capable of withstanding expected in vivo loading.

Technological Characteristics

Titanium acetabular shell; porous titanium bead coating (-45/+60); optional hydroxyapatite coating. Designed for uncemented application. Compatible with existing Reflection Hip System UHMWPE liners. Single-use.

Indications for Use

Indicated for individuals undergoing primary and revision hip surgery due to trauma, inflammatory joint disease (e.g., rheumatoid arthritis), or noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, pelvic fracture, diastrophic variant, remote osteomyelitis, nonunion, proximal femur fractures, femoral osteotomy, Girdlestone resection, fracture dislocation, or deformity).

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Submission Summary (Full Text)

{0}------------------------------------------------ # AUG 2 8 2002 K022556 Page 141 # Summary of Safety and Effectiveness Smith & Nephew, Inc. Reflection Modified Acetabular Shells # Contact Person and Address Kim Kelly Project Manager, Clinical and Regulatory Affairs Smith & Nephew, Inc., Orthopaedic Division 1450 East Brooks Road Memphis, TN 38116 (901) 399-6566 ## Device Description The Reflection Modified Acetabular Shells are titanium acetabular components with peripheral fixation holes in the spline of the device. The shells are coated with a -45/+60 porous titanium bead coating. The shells are also offered in a hydroxyapatite coating applied to the porous surface of the implant. The components are designed for use with existing UHMWPE acetabular liners of the Reflection Hip System. #### Device Classification Name 21 CFR 888.3358 Hip joint metal/polymer/metal, semi-constrained porous coated uncemented prosthesis: Class II #### Indications for Use Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion; femoral neck fracture and trochanteric fractures of the proximal femur with heads involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. The Reflection Modified Acetabular Shells are designed for uncemented applications and are single use only. ## Mechanical and Clinical Data A review of the mechanical test data indicated that the Reflection Modified Acetabular Shells are equivalent to devices currently on the market and capable of withstanding expected in vivo loading without failure. ### Substantial Equivalence Information The Reflection Modified Acetabular Shells shares similarities to currently marketed Smith & Nephew acetabular components in regards to material composition, metallic bead coating, hydroxyapatite coating, shape, locking mechanism, and accessories. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight, composed of three curved lines. AUG 2 8 2002 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Kim P. Kelly Project Manager, Clinical and Regulatory Affairs Smith & Nephew, Inc. Orthopaedic Division 1450 East Brooks Road Memphis, TN 38116 Re: K022556 Trade/Device Name: Reflection Modified Acetabular Shells Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: II Product Code: LPH, MEH Dated: July 31, 2002 Received: August 2, 2002 Dear Ms. Kelly: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 – Ms. Kim P. Kelly This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Sincerely yours, Mark N Mulkerson Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ K022556 # Reflection Modified Acetabular Shells Indications Statement Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion; femoral neck fracture and trochanteric fractures of the proximal femur with heads involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Concurrence of CDRH, Office of Device Evaluation Prescription Use OR (Per 21 CFR 801.109) Over-The Counter Use Na Mark N. Milliken (Division Sign-Off) Division of General, Restorative and Neurological Devic 510(k) Number 202253
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...