← Product Code [GEI](/productcode/GEI) · K022532

# TISSUELINK BIPOLAR FLOATING BALL DEVICE (K022532)

_Tissuelink Medical, Inc. · GEI · Feb 12, 2003 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K022532

## Device Facts

- **Applicant:** Tissuelink Medical, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Feb 12, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The TissueLink Bipolar Floating Ball device is a sterile, single use electrosurgery device intended to be used in conjunction with an electrosurgical generator for delivery of radiofrequency current and saline for hemostatic sealing and coagulation of soft tissue and bone at the operative site. It is intended for, but not limited to, endoscopic and open abdominal, orthopaedic, spine and thoracic surgery. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).

## Device Story

The TissueLink Bipolar Floating Ball device is a sterile, single-use electrosurgical instrument designed for use with an external electrosurgical generator. It delivers radiofrequency (RF) current and saline to the operative site to achieve hemostatic sealing and coagulation of soft tissue and bone. The device is utilized by surgeons in endoscopic and open surgical procedures, including abdominal, orthopaedic, spine, and thoracic surgeries. By combining RF energy with saline irrigation, the device facilitates controlled tissue coagulation, aiding in hemorrhage control during surgery. It is intended for professional use in clinical settings.

## Technological Characteristics

Sterile, single-use electrosurgical device. Operates via bipolar radiofrequency current delivery combined with saline irrigation for tissue coagulation. Designed for use with an external electrosurgical generator.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- TissueLink Bipolar Floating Ball ([K020574](/device/K020574.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) Premarket Submission – TissueLink Medical, Inc. –

Expanded indications Bipolar Floating Ball device

Image /page/0/Picture/2 description: The image shows a sequence of handwritten characters and numbers. The sequence reads 'KO22532', with a horizontal line drawn underneath the entire sequence. The characters are written in a cursive style, and the numbers are slightly stylized.

# 510(k) Summary of Safety and Effectiveness

FEB 1 2 2003

This summary of 510(k)-safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

### Application Information:

| Date Prepared:       | July 31, 2002                                                           |
|----------------------|-------------------------------------------------------------------------|
| Submitter:           | TissueLink Medical Inc.                                                 |
| Address:             | One Washington Center Suite 400<br>Dover, NH 03820                      |
| Contacts:            | Vicki S. Anastasi<br>Directory Regulatory Affairs                       |
| Telephone Number:    | (508) 922-1622                                                          |
| FAX Number:          | (508) 497-9925                                                          |
|                      | Roberta L. Thompson<br>Vice President, Clinical, Regulatory and Quality |
| Telephone Number:    | (603) 742-1515 ext. 106                                                 |
| Fax Number:          | (603) 742-1488                                                          |
| Device Information:  |                                                                         |
| Trade Name:          | TissueLink Bipolar Floating Ball device                                 |
| Common Name:         | Electrosurgery Cauterizing Pen                                          |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories. 21CFR.  |

#### Predicate Devices:

Claim of Substantial Equivalence of the TissueLink Bipolar Floating Ball device is made to:

Identical device, TissueLink Bipolar Floating Ball, K#020574

878.4400

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure embracing a sphere. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 2 2003

Ms. Vicki S. Anastasi Director, Regulatory Affairs TissueLink Medical, Inc. One Washington Center, Suite 400 Dover, New Hampshire 03820

Re: K022532

Trade/Device Name: TissueLink Bipolar Floating Ball Device Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: November 13, 2002 Received: November 14, 2002

Dear Ms. Anastasi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Vicki S. Anastasi

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Expanded indications Bipolar Floating Ball device

13

## Indications for use Statement

Page _____ of ________________________________________________________________________________________________________________________________________________________________

K022532 510(k) Number (if known):

Device Name:

TissueLink Bipolar Floating Ball device

Indications for Use:

The TissueLink Bipolar Floating Ball device is a sterile, single use electrosurgery device intended to be used in conjunction with an electrosurgical generator for delivery of radiofrequency current and saline for hemostatic sealing and coagulation of soft tissue and bone at the operative site. It is intended for, but not limited to, endoscopic and open abdominal, orthopaedic, spine and thoracic surgery. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

|  | Concurrence of CDRH, Office of Device Evaluation (ODE) |  |
|--|--------------------------------------------------------|--|
|--|--------------------------------------------------------|--|

| Prescription Use     |  |
|----------------------|--|
| (Per 21 CFR 801.109) |  |

|  | OR | Over-The-Counter Use |
|--|----|----------------------|
|--|----|----------------------|

|  | Optional Format 1 - |
|--|---------------------|
|--|---------------------|

|  | Miriam C. Provost<br>(Division Sign-Off)<br>Division of General, Restorative<br>and Neurological Devices |
|--|----------------------------------------------------------------------------------------------------------|
|--|----------------------------------------------------------------------------------------------------------|

| TissueLink Medical, Inc. | 510(k) Number | KO22532 |
|--------------------------|---------------|---------|
|--------------------------|---------------|---------|

---

**Source:** [https://fda.innolitics.com/device/K022532](https://fda.innolitics.com/device/K022532)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
