The Summit Medical CellTrans™ Postoperative Autotransfusion System is intended for the collection for reinfusion of blood lost by a patient following surgery including but not limited to orthopaedic joint replacement. The device is indicated for autologous blood transfusion.
Device Story
Postoperative autotransfusion system; collects blood lost post-surgery via low-vacuum suction. Components: needle/drain assemblies, patient drain connection line, compressible bellows collection vessel, blood transfer bags. Operation: bellows compression creates vacuum; blood flows from wound through gross filter into transfer bags. Non-return valves prevent backflow. Once bag is full, it is removed for patient reinfusion; system supports second bag collection. Can function as standard wound drainage device after autotransfusion use. Used in clinical settings; operated by healthcare professionals. Benefits: enables autologous blood recovery, reducing need for homologous transfusion.
Clinical Evidence
Bench testing only. Performance validation included vacuum testing, pressure leak decay tests, functional testing using water and time-expired blood, and mechanical testing of bonded joints. Biocompatibility testing conducted per ISO 10993 and FDA Blue Book Memo G95-1 (cytotoxicity, sensitization, irritation, systemic toxicity, haemocompatibility). Haemocompatibility testing per ISO 10993-4 compared blood processed through the system against homologous (banked) blood, demonstrating performance consistent with design specifications.
Technological Characteristics
Sterile, single-use autotransfusion apparatus. Components include compressible bellows, gross filter, and transfer bags with non-return valves. Materials validated for biocompatibility per ISO 10993. Sterilization via Gamma radiation (ISO 11137, EN552). Mechanical operation via vacuum-assisted collection. No software or electronic components.
Indications for Use
Indicated for autologous blood transfusion in patients following surgery, including orthopaedic joint replacement, to collect and reinfuse lost blood.
Regulatory Classification
Identification
An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.
Predicate Devices
Betatrans Orthopaedic Autotransfusion System (K951592)
Submission Summary (Full Text)
{0}------------------------------------------------
CellTrans™ 510(k) Attachment 7
# 510(k) Summary
### Summit Medical CellTrans™ Postoperative Autotransfusion Device
### Manufacturer
Summit Medical Ltd Industrial Park Bourton on the Water Gloucestershire GL54 2HQ nk
#### Contact
James Bradbury Regulatory Affairs Manager
011 (44) 1451 821311 Phone: Fax: 011 (44) 1451 821092 james.Bradbury@summit-medical.co.uk e-mail:
#### Device Name
CellTrans 100 Postoperative Autotransfusion Transfusion System
### Classification Name
Autotransfusion Apparatus
#### Predicate Product
Betatrans Orthopaedic Autotransfusion System (K951592), Duxbury Scientific Inc.
#### Product Description
The CellTrans 10 device consists of a sterile, double wrapped package, containing two needle and drain assemblies, a blood collection assembly with patient drain connection line and a compressible bellows collection vessel with twin outlet ports, and two blood Transfer bags. The patient and transfer bag lines are fitted with pinch clamps, and in addition the transfer bag connection luers are fitted with sealing caps, for use after a
> Attachment 7 Page 3 of 5
{1}------------------------------------------------
filled bag is removed from the collection vessel. The collection bellows and transfer bags are fitted with non-return valves to prevent backflow of exudate to the patient.
The drains are placed in the patient, and connected to the collection system via the twin drain Y-connector. The transfer bags are fitted to the collection vessel outlet ports. Blood collection is initiated by compressing the bellows to apply a low vacuum to the wound, the collected blood then transferring into one of the transfer bags, via the gross filter. When the baq is full, or in accordance with the American Association of Blood Banks Guidelines for Blood Recovery and Reinfusion in Surgery (AABB), the bag can be removed from the collection system, and transferred for reinfusion to the patient. Blood collection can continue with the second transfer bag.
Following use of the second bag, the device can be used as a wound drainage device if this is clinically desirable, by draining into a standard wound drainage bag.
### Substantial Equivalence
The Summit Medical CellTrans™ Postoperative Autotransfusion System is substantially equivalent to a number of legally marketed Autotransfusion systems, specifically the Duxbury Scientific Inc. Betatrans Orthopaedic Autotransfusion system (K951592), in that they are designed with the same design principles, made of similar materials, and have the same indications and contraindications for use.
### Indications for Use
The Summit Medical CellTrans™ Postoperative Autotransfusion System is intended for the collection for reinfusion of blood lost by a patient following surgery including but not limited to orthopaedic joint replacement. The device is indicated for autologous blood transfusion.
### Safety and Effectiveness
Performance testing carried out includes vacuum testing, pressure leak decay tests, functional testing using both water and time-expired blood, mechanical testing of bonded joints, microbiological bioburden and endotoxin validation.
The CellTrans TM device is Gamma sterilised, validation of which was carried out in accordance with ISO 11137 and EN552.
Biocompatibility testing was carried out in accordance with ISO 10993 and FDA Blue Book Memo G95-1, including Cytotoxicity, Sensitization, Irritation or Intracutaneous Reactivity, Acute Systemic Toxicity, Subchronic Toxicity and Haemocompatibility.
{2}------------------------------------------------
Detailed Haemocompatibility testing was carried out in accordance with ISO 10993-4, with the focus of the testing on the evaluation of a number of blood cell parameters, comparing blood collected and passed through the CellTrans™ system with homologous (banked) blood.
The test results achieved demonstrate that the CellTrans™ device meets the applicable standards, is biocompatible, performs in accordance with design specifications.
No safety or effectiveness issues are raised when the Summit Medical CellTrans™ is compared with the predicate product, and therefore the Summit Medical CellTrans™ is substantially equivalent to the Duxbury Scientific Betatrans Orthopaedic Autotransfusion System (K951592).
Javes Brooks
James Bradbury Regulatory Affairs Manager Summit Medical Ltd.
25. July 2002
Date
Attachment 7 Page 5 of 5
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the text "Public Health Service". The text is in a simple, sans-serif font and is horizontally oriented. The text is black and the background is white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# 0 9 2002
Summit Medical Ltd. c/o Mr. Neil R. Armstrong MeddiQuest Ltd. Business and Technology Centre Bessemer Drive Stevenage SG1 2DX United Kingdom
Re: K022489
Cell Trans™ Postoperative Autotransfusion System Regulation Number: 868.5830 Regulation Name: Autotransfusion Apparatus Regulatory Class: II (two) Product Code: CAC Dated: July 26, 2002 Received: July 29, 2002
Dear Mr. Armstrong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Neil R. Armstrong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
l Qalath
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Summit Medical Ltd
# Indications for Use
510(k) Number:
KO22489
Device Name:
# Summit Medical CellTrans™ Postoperative Autotransfusion System
Indications for Use:
The Summit Medical CellTrans™ Postoperative Autotransfusion System is intended for the collection for reinfusion of blood lost by a patient following surgery including but not limited to orthopaedic joint replacement. The device is indicated for autologous blood transfusion.
(PLEASE DO NOT WRITE BELOW THIS LINE)
Concurrence of CDRH, Office of Device Evaluation(ODE)
Prescription Use
(Per 21 CFR 801.109)
ાર Over-the-counter Use _
Division of Cardiova 510(k) Number
Attachment 8 Page 3 of 3
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.