MODEL OBC-63 BREAST ARRAY COIL
K022488 · Mri Devices Corp. · MOS · Aug 23, 2002 · Radiology
Device Facts
| Record ID | K022488 |
| Device Name | MODEL OBC-63 BREAST ARRAY COIL |
| Applicant | Mri Devices Corp. |
| Product Code | MOS · Radiology |
| Decision Date | Aug 23, 2002 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 892.1000 |
| Device Class | Class 2 |
Intended Use
To be used in conjunction with a Magnetic Resonance Scanner to produce diagnostic images of the female breast, chest wall, and axillary tissues that can be interpreted by a trained physician.
Device Story
Model OBC-63 Breast Array Coil is a radiofrequency (RF) coil accessory for Magnetic Resonance (MR) scanners. It functions as a receiver to detect MR signals from female breast, chest wall, and axillary tissues. The device is used in clinical settings by trained medical professionals. It transforms detected electromagnetic signals into raw data for the MR scanner's image reconstruction system. The resulting images are interpreted by physicians to aid in clinical diagnosis. The device enhances signal-to-noise ratio and spatial resolution compared to body coils, facilitating detailed anatomical visualization.
Clinical Evidence
Bench testing only.
Technological Characteristics
RF receiver coil array; designed for integration with MR scanners; non-invasive; patient-contact materials compliant with medical device standards; no internal software or active processing circuitry.
Indications for Use
Indicated for female patients requiring diagnostic imaging of the breast, chest wall, and axillary tissues via Magnetic Resonance Scanner.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Related Devices
- K961525 — ARRAY OPEN BREAST COIL (MODEL OBC-63 PA) · Mri Devices Corp. · Sep 5, 1996
- K023164 — BBC-63-GE BREAST ARRAY COIL · Mri Devices Corp. · Oct 11, 2002
- K024208 — OBC-250-PAN BREAST ARRAY COIL · Mri Devices Corp. · Jan 7, 2003
- K993776 — OBC-300 BREAST ARRAY COIL · Mri Devices Corp. · Dec 3, 1999
- K993233 — OBC-085 BREAST ARRAY COIL · Mri Devices Corp. · Oct 15, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## AUG 2 3 2002
Mr. Thomas Schubert President MRI Devices Corporation 1515 Paramount Drive WAUKESHA WI 53186
Re: K022488
Trade/Device Name: Model OBC-63 Breast Array Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device
Regulatory Class: II Product Code: 90 MOS Dated: July 26, 2002 Received: July 29, 2002
Dear Mr. Schubert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Snogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Section C – Statement of Indications for Use:
Applicant: MRI Devices Corporation 510(k) number (if known):_ K 022488 Device Name: Model OBC-63 Breast Array Coil
Indications for use:
To be used in conjunction with a Magnetic Resonance Scanner to produce diagnostic images of the female breast, chest wall, and axillary tissues that can be interpreted by a trained physician.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Brogdon
`(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices`
K022488
```
510(k) Number
Prescription Use ***_*** or Over-The-Counter Use **_**
(Per 21 CFR 801.109)
(Optional Format 1-2-96)