← Product Code [LPH](/productcode/LPH) · K022463

# HA MODULAR REACH PROXIMAL POROUS AND HA MODULAR REACH COLLARED PROXIMAL POROUS FEMORAL STEMS (K022463)

_Biomet Orthopedics, Inc. · LPH · Aug 16, 2002 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K022463

## Device Facts

- **Applicant:** Biomet Orthopedics, Inc.
- **Product Code:** [LPH](/productcode/LPH.md)
- **Decision Date:** Aug 16, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The indications for the HA Modular Reach™ Proximal Porous and HA Modular Reach™ Collared Proximal Porous Femoral Stems are for 1) Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 2) Rheumatoid arthritis; 3) Correction of functional deformity; 4) Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and 5) Revision procedures where other treatments or devices have failed.

## Device Story

HA Modular Reach™ Proximal Porous and Collared Proximal Porous Femoral Stems are prosthetic hip joint components; intended for uncemented implantation. Device features modified profile on select sizes and addition of hydroxyapatite (HA) coating to promote osseointegration. Used by orthopedic surgeons in clinical settings for hip replacement procedures. Device replaces diseased or damaged femoral head/neck; restores joint function; corrects deformity. Substantial equivalence established via mechanical testing comparing modified design and coating to predicate.

## Clinical Evidence

No clinical testing was provided as a basis for substantial equivalence; bench testing only.

## Technological Characteristics

Prosthetic hip joint; metal/polymer/metal semi-constrained porous-coated uncemented femoral stem. Features hydroxyapatite (HA) coating. Design and sizing similar to predicate. Class II device.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- Modular Reach™ Hip ([K994038](/device/K994038.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

AUG 1 6 2002

K012463 pge 172

### SUMMARY OF SAFETY AND EFFECTIVENESS

| Applicant or Sponsor: | Biomet Orthopedics, Inc.<br>P.O. Box 587<br>Warsaw, IN 46581-0587 |
|-----------------------|-------------------------------------------------------------------|
|-----------------------|-------------------------------------------------------------------|

Contact Person: Dalene T. Binkley Telephone: (574) 267-6639

Proprietary Name: HA Modular Reach™ Proximal Porous and HA Modular Reach™ Collared Proximal Porous Femoral Stems

Common Name: Prosthetic Hip Joint - Hydroxyapatite (HA) Coated Porous Femoral Stem

Classification: hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis

Device Classification: Class II

Legally Marketed Device to which Substantial Equivalence is Claimed: Modular Reach™ Hip- 510(k) K994038.

Device Description: The HA Modular Reach™ Proximal Porous and HA Modular Reach™ Collared Proximal Porous Femoral Stems are the same as their predicate, except for a modified profile on two of the sizes and the addition of Hydroxyapatite Coating.

Indications for Use: The indications for the HA Modular Reach™ Proximal Porous and HA Modular Reach™ Collared Proximal Porous Femoral Stems are for ) Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 2) Rheumatoid arthritis; 3) Correction of functional deformity; 4) Treatment of nonunion, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and 5) Revision procedures where other treatments or devices have failed.

Summary of Technologies: The HA Modular Reach™ Proximal Porous and HA Modular Reach™ Collared Proximal Porous Femoral Stems- the materials, design, sizing, and indications are similar or identical to the predicate devices.

> MAILING ADDRESS P.O. Box 587 Warsaw, IN 46581-0587 - - - -

SHIPPING ADDRESS 56 E. Bell Drive Warsaw, IN 46582

트

# 000046

()FFICE 574.267.6639 ■ ■ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

E-MAIL biomet@biome1.com

{1}------------------------------------------------

Non-Clinical Testing: Mechanical Testing with an Engineering Justification determined that the HA Modular Reach™ Proximal Porous and HA Modular Reach™ Collared Proximal Porous Femoral Stems presented no new risks and were, therefore, substantially equivalent to the predicate device.

Clinical Testing: No clinical testing was provided as a basis for substantial equivalence.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of a human figure in profile, with three overlapping heads suggesting a sense of community or interconnectedness. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the graphic.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 6 2002

Ms. Dalene T. Binkley Regulatory Affairs Specialist Biomet Orthopedics, Inc. P.O. box 587 Warsaw, Indiana 46581-0587

Re: K022463

Trade/Device Name: HA Modular Reach™ Proximal Porous and HA Modular Reach™ Collared Proximal Porous Femoral Stems Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH Dated: July 15, 2002 Received: July 26, 2002

Dear Ms. Binkley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Dalene T. Binkley

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Sincerely yours,

Mark A. Millman

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page / of /

510 (k) NUMBER (IF KNOWN): __ K o 2 >46 -

DEVICE NAME: HA Modular Reach™ Proximal Porous and HA Modular Reach™ Collared Proximal Porous Femoral Stems

INDICATIONS FOR USE:

The indications for the HA Modular Reach™ Proximal Porous and HA Modular Reach™ Collared Proximal Porous Femoral Stems are for ) Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 2) Rheumatoid arthritis; 3) Correction of functional deformity; 4) Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and 5) Revision procedures where other treatments or devices have failed.

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109

OR

Over-The-Counter-Use No
(Optional Format 1-2-96)

Mark H. Millican

ivision Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K022463

0000008

---

**Source:** [https://fda.innolitics.com/device/K022463](https://fda.innolitics.com/device/K022463)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
